Commission Implementing Regulation (EU) 2022/2346 of 1 December 2022 laying down common specifications for the groups of products without an intended medical purpose listed in Annex XVI to Regulation (EU) 2017/745 of the European Parliament and of the Council on medical devices (Text with EEA relevance)
Modified by
- Commission Implementing Regulation (EU) 2023/1194of 20 June 2023amending Implementing Regulation (EU) 2022/2346 as regards the transitional provisions for certain products without an intended medical purpose listed in Annex XVI to Regulation (EU) 2017/745 of the European Parliament and of the Council(Text with EEA relevance), 32023R1194, June 21, 2023
(a) the product was already lawfully marketed in the Union before 22 June 2023 and continues to comply with the requirements of Union and national law that were applicable to it before22 June 2023 ;(b) there are no significant changes in the design and intended purpose of the product.
(a) the product was already lawfully marketed in the Union before 22 June 2023 and continues to comply with the requirements of Union and national law that were applicable to it before22 June 2023 ;(b) there are no significant changes in the design and intended purpose of the product.
(a) risk management planning; (b) identification of hazards and risk analysis; (c) risk evaluation; (d) risk control and evaluation of residual risks; (e) risk management review; (f) production and post-production activities.
(a) references and description of the device, including its parts and components; (b) a list of the activities to be performed in each step of the risk management process, their scope and the actions for the verification of completion and effectiveness of the risk control measures; (c) a specification of the life cycle phases of the device covered by each activity included in the plan; (d) a specification of responsibilities and authorities for the execution of the activities, for the approvals of the results and for the risk management review; (e) a specification of criteria for risk acceptability based on the policy referred to in Section 2.2; (f) a specification of criteria for the collection of relevant information from the production and from the post production phases and for the use of such information to review and update, if necessary, the risk management results.
(a) include a description of the device, its intended use and the reasonably foreseeable misuse; (b) list the qualitative and quantitative characteristics that could affect the safety of the device; (c) list the known and the foreseeable hazards associated with the device, its intended use, its characteristics and its reasonably foreseeable misuse, when used in both normal and fault conditions; (d) list the hazardous situations resulting from the consideration of the foreseeable events for each identified hazard; (e) include the qualitative or quantitative terms and descriptions, or the categorization, for the estimation of the severity and the probability of occurrence of harms. (f) for each hazardous situation, list the estimated severity and probability of occurrence of harms and the resulting estimation of risks.
(a) a list of the implemented risk control measures and the evaluation of their effectiveness; (b) a list of the residual risks after completed implementation of the risk control measures; (c) the evaluation of acceptability for residual risks and for the overall residual risk, in accordance with the criteria referred to in Section 3.1, point (e); (d) the verification of the effects of the risk control measures.
(a) inherent safety ensured by design; (b) inherent safety ensured by manufacturing; (c) protective measures in the device or in the manufacturing process; (d) information for safety and, where appropriate, user training.
(a) the risk management process has been carried out in accordance with the risk management planning documents referred to in Section 3.1; (b) the overall residual risk is acceptable and the risks have been eliminated or reduced as far as possible; (c) the system to collect and review information on the device from the production and the post-production phases is implemented.
(a) specify the system to collect and review information on the device from the production and the post-production phases; (b) list the sources of publicly available information on the device, on equivalent devices without an intended medical purpose or on analogous devices with a medical purpose; (c) specify the criteria to evaluate the impact of the information collected on the results of previous risk management activities and the consequent actions on the device.
(a) hazards or hazardous situations that have not been identified previously; (b) hazardous situations for which the risk is no longer acceptable; (c) whether the overall residual risk is no longer acceptable.
(a) include new hazards or hazardous situations and evaluate the related risks; (b) re-evaluate hazardous situations, residual risks and the overall residual risk no longer acceptable; (c) establish the need for actions in relation to the devices already made available on the market.
(a) the different degree of understanding of users and consumers, with particular emphasis on devices intended to be used by lay persons; (b) the work environment where the device is intended to be used, especially in case of use outside a medical or otherwise professionally controlled work environment.
(a) the information regarding the categories of users and consumers referred to in Section 2.5; (b) the expected performance of the device; (c) the risks arising from the use of the device.
(a) the information regarding the categories of users and consumers referred to in Section 2.5; (b) a description of the expected performance of the device, in such a way so that the user and the consumer understands which non-medical effect can be expected from the use of the device; (c) a description of the residual risks of the device, including their control measures, presented in a clear and easily understandable way so that the consumer can make an informed decision on whether to be treated with it, have it implanted or otherwise use it; (d) the expected lifetime or the expected resorption period of the device and any necessary follow-up; (e) reference to any harmonised standard and common specifications applied.
Design and manufacturing (a) the shape of the device, in particular in view of avoiding irritation by edges or sharps, disconnection or dislocation from the cornea, wrinkling or folding, unequal pressure on the cornea related to positioning; (b) the selection of raw materials for lens, for surface treatments and, if relevant, for lens storage solutions in view of biological safety, biocompatibility, chemical and biological contaminants as well as permeability of oxygen and compatibility with lens storage solutions; (c) biological safety and biocompatibility of the final product, with its packaging and storage solution, including consideration of at least the aspects of cytotoxicity, sensitization, irritation, acute systemic toxicity, subacute toxicity, implantation, sterilization residues and degradation products, extractable and leachable substances. Where the cumulative contact duration is expected to exceed 30 days, aspects of subchronic toxicity, chronic toxicity and genotoxicity shall also be considered; (d) microbiological properties, including bioburden, microbiological contamination of the final device, residual bacterial endotoxins, sterility, contact lens disinfection and preservation; (e) appropriateness of the primary packaging in terms of keeping the lens sterile, permanently covered by storage fluid and avoiding degradation of the product, for example by leaching of container or cover materials, by intrusion of microbial contaminations; (f) the effect of long-term storage and the conditions of storage on the stability and properties of the lens;
Distribution chain (a) lack of pre-use testing of suitability of lens wearing performed by ophthalmologist, optometrist, specialised optician or qualified contact lens specialist; (b) lack of expertise of distributors outside the classic optician distribution chain with regard to both the selection of appropriate lenses and their use, storage and safe transport; (c) lack of expertise of distributors outside the classic optician distribution chain with regard to safety or handling advice to the users;
User-related hazards/risks (a) lack of experience with and training on the use of contact lenses of certain uses; (b) identification of contra-indications under which contact lenses shall not be used; (c) possible reduced availability to the cornea of tear film and oxygen; (d) lack of hygiene, such as failure to wash and dry hands prior to users placing, using and removing lenses resulting in possible infection, severe inflammation or other diseases of the eye; (e) possible vision hindrance and reduced transmission of light; (f) any possible factors that could cause deterioration of eye sight such as coloration, lack of precise fitting to the eye’s surface and lack of correction; (g) identification of any non-medical conditions under which contact lenses are not be used. Conditions to be considered shall include driving, piloting or operating heavy machinery and water-based activities such as showering, bathing and swimming; (h) increased risk of eye damage if the lenses are worn extensively (for example for long periods, consecutive multiple use); (i) increased risk of eye damage if lenses are still worn when eye redness and irritation occur; (j) the effect of duration of use on any of the risks mentioned above; (k) possible misuse of the primary packaging as containment for storage between several uses; (l) for multiple use contact lenses, risks linked to re-use and irregular re-use by the same consumer; (m) lack of familiarity of consumers with emergency measures in case of any undesirable side-effects.
(a) The field of vision shall not be reduced by the lens, including in case of reasonably foreseeable dislocation or imprecise placing. The lens shall permit transmission of sufficient light for adequate visibility under any condition of use. (b) All materials of the lens and the inner side of its primary packaging, including its storage solution, shall be biocompatible, non-irritating and non-toxic. In addition, the substances used for colouring of or printing on the contact lenses shall not leach under the intended conditions of use. (c) Lenses and the inner side of their primary packaging, including its storage solution shall be sterile and non-pyrogenic. If in contact with the eye, the storage liquid shall not injure the cornea, eye and the surrounding tissue. (d) Lenses shall be designed so as not to compromise the health of the cornea, eye and surrounding tissue. Lens features such as low oxygen permeability, imprecise placing, dislocation, sharp edges, abrasion, unequal mechanical pressure distribution shall be considered. (e) As regards lenses for multiple use, the manufacturer shall either provide effective maintenance liquids and means for cleaning and disinfecting together with the lens sufficient for the entire lifetime of the lens, or indicate the required maintenance liquids and means for cleaning and disinfecting. The manufacturer shall also either provide or indicate any other equipment or tools for the maintenance and cleaning of the lenses for multiple use. (f) As regards lenses for multiple use, the manufacturer shall validate the maximum number of re-uses and maximum duration of use (for example in hours per day and/or number of days). (g) Manufacturers shall consider whether eye drops need to be used to compensate for dryness. Where such eye-drops are needed, manufacturers shall define criteria to demonstrate their suitability; (h) Manufacturers shall establish a procedure for the identification of any undesirable side-effects by the user and how to deal with them, including reporting to the manufacturer of such undesirable side-effects; (i) The instructions for use and the label shall be designed and written in a manner so that they can be understandable by a lay person and that enables a lay person to use the device safely.
(a) where devices are intended for single use, in addition to the internationally recognized symbol, in bold font of largest used size on the label the text "Do not re-use"; (b) indication of the dimensions of the lens (outer diameter of the lens and base curve radius); (c) the recommendation to read the instructions for use.
(a) in bold font of largest used size in the instructions the text: "Do not re-use", in addition to the internationally recognized symbol, where devices are intended for single use; (b) a warning "Used lenses shall not be used by other persons"; (c) the indication of the dimensions of the lens (outer diameter of the lens and base curve radius); (d) the indication of the materials of the lens, including its surface and colouring pigments; (e) the indication of water content and oxygen permeability; (f) an indication of the possible effect of incorrect storage conditions on the quality of the product and maximum storage time; (g) instructions on what to do in case of displacement; (h) hygiene measures prior to use (for example washing and drying hands), during use and after use; (i) a warning "Do not contaminate lenses with make-up or aerosols."; (j) a warning "Do not clean lenses with tap water."; (k) for multiple use lenses, a detailed description of the cleaning and disinfection procedure, including the description of the necessary equipment, tools and solutions, which shall be named in detail; a description of required storage conditions; (l) for multiple use lenses, the maximum number of re-uses and maximum duration of use(s) (for example in hours per day and/or number of days); (m) where the use of eye drops is recommended, a description of suitable eye drops and the description on how to use them; (n) listing of contraindications under which contact lenses are not to be used. Such a list shall include: dry eyes (inadequate tear fluid), use of eye medication, allergies, inflammation or redness in or around the eye, poor health affecting the eye such as cold and flu, previous medical intervention which may adversely affect the use of the device, any other systemic illness affecting the eye; (o) a warning: "Do not use whilst participating in traffic-related situations (for example driving, riding a bike), operating machinery or whilst undertaking water-related activities such as showering, bathing and swimming."; (p) a warning: "Avoid activities where possible vision hindrance and reduced transmission of light create a risk."; (q) a statement regarding increased risk of eye damage in case of continuous wear when eye redness and irritation occur; (r) a warning "Do not use after date of expiry."; (s) a clear indication of the maximum wearing time; (t) a warning "Do not use lenses beyond maximum wearing time."; (u) a warning "Do not use lenses during sleeping periods"; (v) a statement on the increased risk of eye damage if the lenses are worn extensively (for example multiple re-uses); (w) a warning "Do not use in excessively dry or dusty environments."; (x) a warning "Do not re-use the primary packaging as containment for storage between uses.", where primary packaging is not intended by the manufacturer for such use; (y) a warning: "Do not re-use the storage solution."; (z) a list of risks linked to ocular health associated with lens wear, as identified by risk analysis, including, if applicable, reduced availability to the cornea of water and oxygen (oxygen transmissibility); (aa) a list of possible undesirable side-effects, their probability of occurrence and their indicators; (bb) instructions on how to deal with complications, including emergency measures; (cc) an instruction "Remove the lens immediately in case of: irritation or eye pain such as stinging, burning, itching, foreign body sensation; reduced comfort when compared with previous wearing of an identical lens; unusual secretions or excessive tear-flow, redness of the eye, severe or persistent dryness, reduced or blurred vision linked to the use of the lens.
If any of these symptoms continue after removal of the lens, contact a qualified healthcare professional, such as an ophthalmologist, or an optometrist, authorised by national law to treat such symptoms. The continuation of these symptoms might indicate a more serious condition."; (dd) information on when and how to report undesirable side-effects to the manufacturer.
(a) physical and chemical characteristics and full composition of the implant; (b) the selection of raw materials in view of biological safety, biocompatibility and chemical and biological additives or contaminants; (c) for resorbable devices, resorption and life-time in the body, indicating the half-life and the end of the resorption; (d) biological safety and biocompatibility of the final product, including consideration of at least the aspects of cytotoxicity, sensitisation, irritation, material mediated pyrogenicity, acute systemic toxicity, subacute toxicity, subchronic toxicity, chronic toxicity, genotoxicity, carcinogenicity, implantation, sterilisation residues and degradation products, extractable and leachable substances; (e) microbiological properties, including bioburden, microbiological contamination of the final device, residual bacterial endotoxins and sterility; (f) the specific anatomical location for which clinical and other data support the use of the device; (g) consumer specific factors (for example previous accidents, special conditions, age restrictions); (h) potential interactions with magnetic field, (for example heating related to magnetic resonance imaging); (i) use of accessories (for example delivery instruments designed to be specifically used with the device for the implantation procedure) and their compatibility with the implant; (j) time interval between implantations, where applicable.
(a) microbiological contamination; (b) presence of manufacturing debris; (c) implantation procedure related aspects (including use errors); (d) implant failure (for example rupture, unintended degradation); (e) implant dislodgement and migration; (f) asymmetry; (g) implant visibility through the skin; (h) implant deflation and wrinkling; (i) gel bleeding and leakage; (j) sweating and silicone migration; (k) local inflammation and swelling; (l) regional swelling or lymphadenopathy; (m) capsule formation and contracture; (n) discomfort or pain; (o) hematoma; (p) infection and inflammation; (q) superficial wound; (r) wound dehiscence; (s) extrusion of implant and interruption of wound healing; (t) scarring and scar hyperpigmentation and hypertrophy; (u) nerve injury; (v) seroma; (w) compartment pressure problems and compartment syndrome; (x) limitation in cancer diagnosis; (y) over-sized implants; (z) vascular damage; (aa) breast implant associated anaplastic large cell lymphoma (BIA-ALCL); (bb) granuloma, including siliconoma where applicable; (cc) necrosis.
(a) Devices shall be sterile and non-pyrogenic. Where implants are supplied non-sterile with the intention to be sterilised before use, adequate instruction for sterilisation shall be provided. (b) The safe use of the device shall be supported by clinical and other data considering the anatomical location. (c) Long-term data shall be collected to evaluate the presence of non-degradable substances originating from the devices. (d) Presence of substances referred to in Section 10.4.1, points (a) and (b), of Annex I to Regulation (EU) 2017/745 shall be evaluated independently of their concentration. (e) Manufacturers shall provide training on the implantation and safe use of the device. That training shall be accessible to users.
(a) in bold font of largest used size on the label the text: "Only to be implanted in an appropriate medical environment by appropriately trained medical doctors who are qualified or accredited in accordance with national law."; (b) a clear indication that devices are not to be used in persons who are less than 18 years old; (c) the overall qualitative composition of the product.
(a) on top in bold fonts of largest used size in the instructions of use the text: "Only to be implanted in an appropriate medical environment by appropriately trained medical doctors who are qualified or accredited in accordance with national law."; (b) a clear indication that devices are not to be used in persons who are less than 18 years old; (c) the recommendation for the user to consider any previous procedures, accidents, conditions, medications or other simultaneous treatments of the consumer that may affect the procedure (for example skin diseases, traumas and auto-immune diseases); (d) the instruction for the user to consider any specific risks that may be applicable to activities of the consumer (for example profession, sports or other activities regularly performed by the consumer); (e) a comprehensive list of contra-indications. This list shall include keloid scars; (f) the overall qualitative and quantitative composition of the product; (g) the recommendation for the user on a post-implantation monitoring time in order to identify any potential undesirable side-effects; (h) an indication of the appropriate time interval between treatments, where applicable; (i) a requirement for the user to provide the consumer with a copy of the annex provided for in Section 6.2 before the consumer is treated with the device.
(a) information listed in Section 12.1, points (a) to (e), of Annex I; (b) a list of all residual risks and potential side-effects, including those commonly related to surgery such as bleeding, potential drug interactions and the risks associated with anaesthesia, in a clear way; (c) information on when and how to report undesirable side-effects to the manufacturer, information on device removal, information on when to contact a healthcare professional; (d) details on volume and size of the device; (e) the statement "The users received appropriate training on how to safely use the device.", where relevant.
(a) physical and chemical characteristics of the device; (b) the selection of raw materials in view of biological safety, biocompatibility and chemical and biological additives or contaminants; (c) biological safety and biocompatibility of the final product, including consideration of at least the aspects of cytotoxicity, sensitisation, irritation, material mediated pyrogenicity, acute systemic toxicity, subacute toxicity, subchronic toxicity, chronic toxicity, genotoxicity, carcinogenicity, implantation, sterilisation residues and degradation products, extractable and leachable substances; (d) resorption and life-time in the body, indicating the half-life and the end of the resorption, including the possibility of metabolisation (for example enzymatic degradation of the filler material such as hyaluronidase for hyaluronic acid fillers); (e) microbiological properties, bioburden, microbiological contamination of the final device, residual bacterial endotoxins and sterility; (f) the specific anatomical location of injection or introduction; (g) consumer specific factors (for example previous and current treatments (medical and surgical), age restrictions, pregnancy, breast-feeding); (h) if applicable, risks related to the use of local anaesthetic, either as part of the product or stand-alone; (i) for non-resorbable devices, the risk associated with the removal of the device; (j) aspects associated with the use of the device, including: injection technique; means of injection (for example rollers, catheters or needles); maximum quantity injected depending on location and applied technique; possible repeated injections; force required to administer the product; product temperature; transfer of the product (for example from a vial to a syringe).
(a) microbiological contamination; (b) presence of manufacturing debris; (c) hazards associated with the procedure to inject or otherwise introduce the device (including use errors); (d) migration of the device; (e) device visibility through the skin; (f) unintended local inflammation and swelling; (g) regional swelling or lymphadenopathy; (h) capsule formation and contracture; (i) discomfort or pain; (j) hematoma; (k) infection and inflammation; (l) superficial wound; (m) interruption of wound healing; (n) scarring and scar hyperpigmentation and hypertrophy; (o) nerve injury; (p) seroma; (q) compartment pressure problems and compartment syndrome; (r) granuloma, including siliconoma where applicable; (s) edema; (t) vascular damage; (u) severe allergic reactions; (v) blindness; (w) necrosis.
(a) Devices shall be sterile non-pyrogenic and intended for single use. (b) The safe use of the device shall be supported by clinical and other data considering the anatomical location. (c) Long-term data shall be collected to evaluate the presence of non-degradable substances originating from the devices. (d) Manufacturers shall provide training on the administration and safe use of the device. That training shall be accessible to users. (e) Presence of substances referred to in Section 10.4.1, points (a) and (b), of Annex I to Regulation (EU) 2017/745 shall be evaluated independently of their concentration.
(a) in bold fonts of largest used size on the label the text: "Only to be administered by appropriately trained healthcare professionals who are qualified or accredited in accordance with national law". (b) a clear indication that devices are not to be used in persons who are less than 18 years old.
(a) on top in bold font of largest used size in the instructions for use the text: "Only to be administered by appropriately trained healthcare professionals who are qualified or accredited in accordance with national law."; (b) a clear indication that devices are not to be used in persons who are less than 18 years old; (c) precise and detailed technical information for a good administering practice; (d) description of treatment of the most common side-effects, such as overdosing, swelling, hardening, nodules and immune responses, with the instruction to consult a medical professional if needed; (e) instructions for users as to how and when new injections can be placed at previously injected locations; (f) a list of constituents, which specifies: all constituents responsible for the intended action with specification of their concentration and, where applicable, their molecular weight range, their particle size and their degree of cross-linking, together with the method used for its determination; other constituents such as cross-linking agents, solvents, anaesthetics and preservatives, with specification of their concentration;
(g) the recommendation for the user to consider any previous procedures, accidents, conditions, medications or other simultaneous treatments of the consumer that may affect the procedure (for example skin diseases, traumas and auto-immune diseases); (h) the recommendation for the user of a post-administration monitoring time in order to identify any potential undesirable side-effects; (i) a requirement for the user to provide the consumer with a copy of the annex provided for in Section 6.2 before the consumer is treated with the device.
(a) information listed in Section 12.1, points (a) to (e), of Annex I; (b) all residual risks and potential undesirable side-effects listed in a clear way and described in a language commonly understood by lay persons. This includes a clear declaration on the presence of any substances referred to in Section 10.4.1 of Annex I to Regulation (EU) 2017/745, heavy metals or other contaminants; (c) information on when and how to report undesirable side-effects to the manufacturer; (d) information on when to contact healthcare professional; (e) any contra-indications to the procedure; (f) the statement "The users received appropriate training on the conditions to safely use the device.", where relevant.
(1) "liposuction" means the surgical removal of localised subcutaneous fat deposits by aspiration; (2) "liposuction devices" means devices intended by the manufacturer to be used for the purpose of liposuction; (3) "lipolysis" means the localised destruction of fat deposit; (4) "lipolysis devices" means devices intended by the manufacturer to be used for the purpose of lipolysis; (5) "lipoplasty" means the modification of body contours by removal of excess fat; (6) "lipoplasty devices" means devices intended by the manufacturer to be used for the purpose of lipoplasty.
(a) the volume of adipose tissue which may be removed or, in the case of lipolysis, destroyed and the expected metabolic effect, including the metabolisation of released tissue components, taking into account the probable different characteristics of the person undergoing treatment; (b) the minimum time lapse between subsequent procedures; (c) the anatomical location of the use of the device; (d) the cannula type, for example the diameter and nature of the tip of the cannula; (e) the amount of suction which will be applied; (f) the use and subsequent metabolisation of infiltrative fluid with a justification for the choice of fluid and its composition; (g) the type of liposuction which the device is intended to provide, for example dry or wet, and the type of anaesthetic; (h) whether the device is a simple liposuction device, i.e. blunt cannula suction, or whether it incorporates any other mechanism of action, for example the use of laser energy or ultrasound; (i) the age distribution, gender and body-mass-index of the population to which the clinical data or other sources of data relate; (j) the way in which energy is emitted.
(a) post-operative seroma; (b) tissue injury, organ perforation and bleeding; (c) post-operative ecchymosis and edema; (d) interference with active implantable or active body-worn medical devices and with metallic passive medical devices or other metallic objects present on or inside the body; (e) thermal injury; (f) mechanical injuries, including those caused by unintended cavitation, and corresponding side-effects; (g) inflammation.
(a) haemorrhage; (b) perforation of abdominal viscera, thorax or peritoneum; (c) pulmonary embolism; (d) bacterial infections such as necrotizing fasciitis, gas gangrene and sepsis; (e) hypovolemic shock; (f) thrombophlebitis; (g) seizures; (h) risks related to local anaesthetic use: consideration should be given to lidocaine-induced cardiotoxicity or lidocaine-related drug interactions for tumescent liposuction.
(a) burns to incision sites and overlying tissue; (b) other harmful effects of the internal or external local discharge of energy; (c) over-exposure; (d) neurovascular and local tissue injury, including reduction in cutaneous sensory nerve function; (e) remodelling of collagen that may lead to neoformations; (f) reorganisation of the dermis, with reference to reticular dermis; (g) body deformity or similar poor aesthetic outcome causing the need for medical intervention; (h) for lipolysis devices that are surgically invasive, the hazards linked to the types and sizes of incision.
(a) coagulant disorders, being treated with anticoagulant medications; (b) uncontrolled hypertension; (c) diabetes mellitus; (d) phlebitis and vasculitis; (e) cancer or tumours; (f) extreme obesity (body mass index above 40); (g) pregnancy; (h) vascular fragility; (i) recent surgery (6 weeks); (j) skin infections and open lesions; (k) varicose veins in the area of treatment; (l) medical conditions, such as heart, lung, or circulatory system disease; (m) age less than 18; (n) incapability to understand the consequences, implications and risks of the medical procedures (for example liposuction, lipolysis, lipoplasty) where the devices are used; (o) elevated body temperature (pyrexia).
(a) any metallic passive medical device or other metallic object present on or inside the body; (b) any active implantable or active body-worn medical device.
(a) hyper- or hypovolaemia; (b) bradycardia; (c) venous thromboembolism; (d) fat embolism; (e) infection; (f) fluid accumulation; (g) skin erythema or panniculitis; (h) contour irregularities.
(a) "Careful consideration shall be given to consumer suitability with respect to medication which has the potential to cause bradycardia or hypotension as this has been reported as the cause of death in a number of consumers undergoing tumescent liposuction. Consumers taking drugs such as beta-adrenergic antagonists, non-dihydropyridine calcium-channel blockers, cardiac glycosides, and centrally acting alpha-adrenergic agonists shall be subject to very careful consideration as deaths have been reported due to bradycardia and hypotension. The procedure has to be preceded by a medical consultation which has to be documented and during which chronic disease and drugs taken by patient need to be considered."; (b) "Consumers shall be warned that they may experience extended post-operative analgesia (for example for 24 hours or more) which may result in reduced sensation in the areas infiltrated and therefore consumers shall be warned to protect themselves from injury.".
(a) information listed in Section 12.1, points (a), (b) and (c), of Annex I; (b) the statement "The users received appropriate training on the conditions to safely use the device.", where relevant; (c) information on when and how to report undesirable side-effects to the manufacturer; (d) a recommendation to undergo a medical consultation, including a diagnostic examination, of the areas intended for the treatment.
(1) "device for professional use" means a device that is intended to be used in a healthcare environment or otherwise controlled professional environment by professionals having proven qualification in the safe and effective use of the device; (2) "device for home use" means a device that is intended to be used in private environments, not in a controlled professional environment, by lay persons.
(a) various skin types and the degree of tanning of the skin; (b) presence of any skin abnormality (for example relief, texture or colour) or disease affecting the skin; (c) age of the consumers; (d) possibility of concurrent medical treatments or drug misuse; (e) use of photosensitising medicines or cosmetics; (f) reduced reaction to harm caused by local or systemic anaesthesia; (g) exposure to other light sources.
(a) burns; (b) formation of scars and keloids; (c) hypopigmentation and hyperpigmentation; (d) accelerated aging of skin; (e) allergic/chemical skin reaction (for example to colour pigments of tattoos or make-up); (f) formations of skin cancers; (g) alteration of skin cancers, skin diseases, nevi, herpes, possible delay of disease diagnoses (for example melanoma, endocrine diseases); (h) reactions in case of possible drug intake or use of cosmetics; (i) possible reactions to sun or sunbed exposure; (j) possible photosensitive dermatosis; (k) vitiligo; (l) erythema, mostly temporary and occasionally persistent; (m) purpura resulting from bleeding from small blood vessels; (n) crusting; (o) edema; (p) blistering; (q) inflammation, folliculitis, skin infection; (r) eye damage, including damage to retina and cornea; (s) prickling or feeling of heat; (t) dry skin and itching due to shaving or combination of shaving and light treatment; (u) excessive pain; (v) paradoxical hypertrichosis (increased growth of hair after treatment); (w) overexposure; (x) unintended release of radiation; (y) ignition, explosion or production of fumes.
(a) avoidance of unauthorized access to or unintended use of the devices (for example by means of key switch or code or dual control of energy emission); (b) display of the characteristics of the emitted optical radiation for the purpose of permanent surveillance and recording of the emission through the device in addition to the requirements of Section 16.2 of Annex I to Regulation (EU) 2017/745; (c) continuous contact controls and an interlock system ensuring that the device works only in case of suitable skin contact with the emitting area of the device; (d) avoidance of overexposure for each session of the treatment by particular measures; (e) where the wavelength of the radiation emitted is less than 1200 nm, instruments or methods to assess the skin pigmentation to ensure proper settings for the treatment;(f) measures to avoid overexposure by repeated treatment sessions or repeated treatments; (g) low energy preset; (h) optimised limitation of pulse energy and pulse duration (exposure time on tissue) and a combination of these two parameters with the wavelength range; (i) optimized limitation of treatment areas (spot sizes) also taking into account the parameters referred to in point (h); (j) minimisation of scattered radiation; (k) minimisation of the risk of accidental emission; (l) emergency stop function (for example emergency stop switch); (m) for devices for hair removal: minimisation of ultraviolet radiation (to be achieved for example by using appropriate high quality band edge filter); (n) devices intended to deliver a permanent change of the appearance shall not be used on persons who are less than 18 years old; (o) information for the user on the correct functioning of the device and the actual mode of operation by means of acoustic or optical means in standby mode, in operating mode and in case of loss of skin contact during the procedure; (p) instruction of the user to protect nevi or lesions during the procedure.
(a) set limits for the duration of exposure and include automatic deactivation to avoid risk of overexposure; (b) include continuous contact controls and an interlock system ensuring that the device works only in case of full skin contact with the emitting area of the device, instead of applying the requirements laid down in point (c) of Section 5.1; (c) include an integrated skin tone sensor assessing the skin patch of or near to the area to be treated and allowing emission output only if skin pigmentation is suited for treatment and if there is continuous full skin contact after skin tone analysis, instead of applying the requirements laid down in point (e) of Section 5.1.
(a) the minimum radiation intensity, duration and frequency of use necessary to trigger the desired effect; (b) the maximum and the recommended radiation intensity, duration and frequency of use; (c) the minimum interval between several applications at the same location; (d) the risks arising from excessive use; (e) the radiation intensity, duration and frequency which triggers a sharp increase of risks, if any; (f) the radiation intensity, duration and frequency beyond which there is no more additional performance; (g) the pulse energy, fluence, wavelength range [nm], pulse duration [ms], pulse profile(s); (h) the maximum admissible treatment spot size [cm 2 ];(i) description of the minimum homogeneity of the treatment spot; (j) description of requirements for the spatial distribution of the treatments spots, taking into account that overlapping treated areas shall not lead to overexposure; (k) safety features of the device; (l) the expected lifetime of the device; (m) the expected stability of performance; (n) cosmetics and drugs interacting or expected to interact with the treatment and their description; (o) other sources of radiation, such as prolonged exposure to sun light or sunbeds, that might increase the risks; (p) for devices for professional use, a requirement for the user to provide the consumer with a copy of the annex provided for in Section 6.11 before the consumer is treated with the device.
(a) the description or a listing of appropriate accessories or conditions of other products used in the procedure; (b) the safety precautions to be taken, which include the use of non-reflective instruments (no mirrors shall be used), the use of absorbing or diffusing surfaces of tools as well as the avoidance of inflammable products and substances and, where applicable, the need to provide adequate room ventilation; (c) an adequate warning notice outside the procedure room.
(a) to avoid at all times exposure of eyes to emitted light; (b) for users, consumers and any other person likely to be exposed to the radiation due to reflection, misuse or mishandling of the emitting device to wear appropriate eye protection during treatments with intense pulsed light or laser devices, especially when these devices are to be used close to the face.
(a) information listed in Section 12.1, points (a), (b) and (c), of Annex I; (b) the statement "The users received appropriate training on the conditions to safely use the device.", where relevant; (c) information on when and how to report undesirable side-effects to the manufacturer; (d) a recommendation to undergo a medical consultation, including a diagnostic examination, of the skin areas intended for the treatment.
(a) the incorrect placement of electrodes and coils may result in failed performance, enhanced electrical currents in tissue or unintended neural responses; (b) brain stimulation may have very different neural responses and thus unintended effects on different groups of persons. Some groups may be particularly vulnerable: persons who are less than 18 years old, young adults, pregnant women, psychiatric patients, persons with psychological disorders or medical conditions affecting the central nervous system, alcohol addicts, users of addictive substances and other substances that modify a person’s natural perception; (c) the presence of active implantable or body-worn medical devices and/or metallic passive medical devices or other metallic objects present on or inside the body may give rise to specific risks arising from the application of electrical energy and magnetic fields; (d) excessive, frequent and cumulative long-term use may result in unforeseen neural effects which in some cases might result in structural changes in the brain.
(a) psychological risks; (b) neural and neuro-toxicity risks; (c) short-term, medium-term and long-term cognitive side-effects, such as compensatory trade-offs (for example the decline or sub-serving of brain regions which are not stimulated); (d) transient auditory threshold shift or tinnitus; (e) long-term side-effect changes of the brain functioning; (f) hazards linked to the long-term effects of repeated stimulation; (g) hazards linked to the use of the device in certain environments highly stimulating or attention demanding; (h) atypical or other idiosyncratic effects; (i) specific hazards arising at the interface between electrodes and skin; (j) electromagnetic interference or injury caused by interaction with active implants (for example pacemakers, implanted cardioverter-defibrillators, cochlear implants, neural implants), active devices (for example neural stimulation devices, medication infusion devices), non-active metallic implants (for example metallic dental implants) or body-worn devices (for example biosensors); (k) hazards associated with device usage after intake of alcohol and/or soft-drugs and/or central nervous system stimulating substances/pharmaceuticals; (l) hazards associated with possible potentiating effects of combined use (usage of few/several devices in the same time targeting same person or different person) and reasonable foreseeable misuse.
(a) persons with a history of epilepsy; (b) persons undergoing pharmaceutical treatment for conditions related to the central nervous system; (c) persons undergoing therapeutic treatment which change the excitability of the central nervous system; (d) users of illicit substances or other substances that modify a person’s natural perception regardless of whether those are commonly understood as therapeutic drugs; (e) persons who have a tumour in the central nervous system; (f) persons who have vascular, traumatic, infectious or metabolic lesions or diseases of the brain; (g) persons who suffer from sleep disorders, drug dependency or alcoholism; (h) persons who are less than 18 years old; (i) pregnant women.
(a) avoidance of unauthorised access to the device (for example by means of key switch or code) and unintended use of the device (for example by means of dual control of energy emission); (b) minimisation of stray magnetic fields; (c) minimisation of the risk of accidental emission; (d) emergency stop function (for example emergency stop switch); (e) automatic deactivation where maximum acceptable output is reached; (f) automatic deactivation where maximum acceptable duration of exposure is reached; (g) automatic deactivation in case of overexposure due to a combination of output and duration; (h) videos with instructions on how to safely use the device made available on the internet; (i) provision of appropriate training accessible to users on safe and effective use of the device; (j) information for the user on the correct functioning of the device and the actual mode of operation through acoustic or optical means in standby mode, in operating mode and in case of loss of full skin contact during the procedure.
(a) specific risks for persons listed in Section 4.1; (b) risks for persons with active implantable or active body-worn medical devices; (c) risks for persons with metallic passive medical devices or other metallic objects present on or inside the body; (d) information about how to deal with over-exposure to energy; (e) information on how to deal with psychological disturbances.
(a) the duration, intensity and frequency of stimulation and all risks arising from use, including from excessive use; (b) the energy delivered, brain area targeted, wave forms and pulse characteristics.
(a) the information listed in Section 12.1, points (a), (b) and (c), of Annex I; (b) the statement "The users received appropriate training on the conditions to safely use the device.", where relevant; (c) information on when and how to report undesirable side-effects to the manufacturer.