Commission Regulation (EU) 2017/771 of 3 May 2017 amending Regulation (EC) No 152/2009 as regards the methods for the determination of the levels of dioxins and polychlorinated biphenyls (Text with EEA relevance. )
Congener | TEF value | Congener | TEF value |
---|---|---|---|
Dibenzo-p-dioxins ("PCDDs") and Dibenzo-p-furans '("PCDFs") | |||
2,3,7,8-TCDD | 1 | ||
1,2,3,7,8-PeCDD | 1 | Non-ortho PCBs | |
1,2,3,4,7,8-HxCDD | 0,1 | PCB 77 | 0,0001 |
1,2,3,6,7,8-HxCDD | 0,1 | PCB 81 | 0,0003 |
1,2,3,7,8,9-HxCDD | 0,1 | PCB 126 | 0,1 |
1,2,3,4,6,7,8-HpCDD | 0,01 | PCB 169 | 0,03 |
OCDD | 0,0003 | Mono-ortho PCBs | |
2,3,7,8-TCDF | 0,1 | PCB 105 | 0,00003 |
1,2,3,7,8-PeCDF | 0,03 | PCB 114 | 0,00003 |
2,3,4,7,8-PeCDF | 0,3 | PCB 118 | 0,00003 |
1,2,3,4,7,8-HxCDF | 0,1 | PCB 123 | 0,00003 |
1,2,3,6,7,8-HxCDF | 0,1 | PCB 156 | 0,00003 |
1,2,3,7,8,9-HxCDF | 0,1 | PCB 157 | 0,00003 |
2,3,4,6,7,8-HxCDF | 0,1 | PCB 167 | 0,00003 |
1,2,3,4,6,7,8-HpCDF | 0,01 | PCB 189 | 0,00003 |
1,2,3,4,7,8,9-HpCDF | 0,01 | ||
OCDF | 0,0003 |
"Screening methods" means methods used for selection of those samples with levels of PCDD/Fs and dioxin-like PCBs that exceed the maximum levels or the action thresholds. They shall allow a cost-effective high sample-throughput, thus increasing the chance to discover new incidents with high exposure and health risks to consumers. Screening methods shall be based on bioanalytical or GC-MS methods. Results from samples exceeding the cut-off value used to check compliance with the maximum level shall be verified by a full re-analysis from the original sample using a confirmatory method. "Confirmatory methods" means methods that provide full or complementary information enabling the PCDD/Fs and dioxin-like PCBs to be identified and quantified unequivocally at the maximum or in case of need at the action threshold. Such methods utilize gas chromatography/high resolution mass spectrometry (GC-HRMS) or gas chromatography/tandem mass spectrometry (GC-MS/MS).
performed by a screening method with a false-compliant rate below 5 %, indicates that the level does not exceed the respective maximum level of PCDD/Fs and the sum of PCDD/Fs and dioxin-like PCBs laid down by Directive 2002/32/EC, performed by a confirmatory method, does not exceed the respective maximum level of PCDD/Fs and the sum of PCDD/Fs and dioxin-like PCBs laid down by Directive 2002/32/EC, taking into account the expanded measurement uncertainty.
(a) Screening methods The goal of screening methods is to select those samples with levels of PCDD/Fs and dioxin-like PCBs that exceed the maximum levels or the action thresholds. Screening methods shall ensure cost-effective high sample-throughput, thus increasing the chance to discover new incidents with high exposure and health risks of consumers. Their application shall aim to avoid false-compliant results. They may comprise bioanalytical and GC-MS methods. Screening methods compare the analytical result with a cut-off value, providing a yes/no-decision over the possible exceedance of the maximum level or action threshold. The concentration of PCDD/Fs and the sum of PCDD/Fs and dioxin-like PCBs in samples suspected to be non-compliant with the maximum level shall be determined or confirmed by a confirmatory method. In addition, screening methods may give an indication of the levels of PCDD/Fs and dioxin-like PCBs present in the sample. In case of application of bioanalytical screening methods the result is expressed as Bioanalytical Equivalents (BEQ), whereas in case of application of physico-chemical GC-MS methods it is expressed as Toxic Equivalents (TEQ). The numerically indicated results of screening methods are suitable for demonstrating compliance or suspected noncompliance or exceedance of action thresholds and give an indication of the range of levels in case of follow-up by confirmatory methods. They are not suitable for purposes such as evaluation of background levels, estimation of intake, following of time trends in levels or re-evaluation of action thresholds and maximum levels. (b) Confirmatory methods Confirmatory methods allow the unequivocal identification and quantification of PCDD/Fs and dioxin-like PCBs present in a sample and provide full information on congener level. Therefore, those methods allow the control of maximum levels and action thresholds, including the confirmation of results obtained by screening methods. Furthermore, results may be used for other purposes such as determination of low background levels in feed monitoring, following of time trends, exposure assessment and building of a database for possible re-evaluation of action thresholds and maximum levels. They are also important for establishing congener patterns in order to identify the source of a possible contamination. Such methods utilise GC-HRMS. For confirming compliance or non-compliance with the maximum level, also GC-MS/MS can be used.
(a) the concentration of an analyte in the extract of a sample which produces an instrumental response at two different ions to be monitored with a S/N (signal/noise) ratio of 3:1 for the less intensive raw data signal; or (b) if for technical reasons the signal-to-noise calculation does not provide reliable results, the lowest concentration point on a calibration curve that gives an acceptable (≤ 30 %) and consistent (measured at least at the start and at the end of an analytical series of samples) deviation to the average relative response factor calculated for all points on the calibration curve in each series of samples. The limit of quantification (LOQ) is calculated from the lowest concentration point taking into account the recovery of internal standards and sample intake.
Screening with bioanalytical or physico-chemical methods | Confirmatory methods | |
---|---|---|
False-compliant rate | < 5 % | |
Trueness | – 20 % to + 20 % | |
Repeatability (RSD | < 20 % | |
Intermediate precision (RSD | < 25 % | < 15 % |
Separation of PCDD/Fs from interfering chlorinated compounds such as non-dioxin-like PCBs and chlorinated diphenyl ethers shall be carried out by suitable chromatographic techniques (preferably with a florisil, alumina and/or carbon column). Gas-chromatographic separation of isomers shall be < 25 % peak to peak between 1,2,3,4,7,8-HxCDF and 1,2,3,6,7,8-HxCDF.
For GC-HRMS: In HRMS, the resolution shall typically be greater than or equal to 10000 for the entire mass range at 10 % valley.Fulfilment of further identification and confirmation criteria as described in internationally recognised standards, for example, in standard EN 16215:2012 (Animal feed — Determination of dioxins and dioxin-like PCBs by GC-HRMS and of indicator PCBs by GC-HRMS) and/or in EPA methods 1613 and 1668 as revised.
For GC-MS/MS: Monitoring of at least 2 specific precursor ions, each with one specific corresponding transition product ion for all labelled and unlabelled analytes in the scope of analysis. Maximum permitted tolerance of relative ion intensities of ± 15 % for selected transition product ions in comparison to calculated or measured values (average from calibration standards), applying identical MS/MS conditions, in particular collision energy and collision gas pressure, for each transition of an analyte. Resolution for each quadrupole to be set equal to or better than unit mass resolution (unit mass resolution: sufficient resolution to separate two peaks one mass unit apart) in order to minimise possible interferences on the analytes of interest. Fulfilment of the further criteria as described in internationally recognised standards, for example, in standard EN 16215:2012 (Animal feed — Determination of dioxins and dioxin-like PCBs by GC-HRMS and of indicator PCBs by GC-HRMS) and/or in EPA methods 1613 and 1668 as revised, except the obligation to use GC-HRMS.
When calculating the concentrations from a TCDD calibration curve, values at the higher end of the curve will show a high variation (high coefficient of variation (CV)). The working range is the area where this CV is smaller than 15 %. The lower end of the working range (reporting limit) shall be set at least by a factor of three above the procedure blanks. The upper end of the working range is usually represented by the EC 70 value (70 % of maximal effective concentration), but lower if the CV is higher than 15 % in this range. The working range shall be established during validation. Cut-off values (see point 7.3) shall be well within the working range.Standard solutions and sample extracts shall be tested in triplicate or at least in duplicate. When using duplicates, a standard solution or a control extract tested in four to six wells divided over the plate shall produce a response or concentration (only possible in the working range) based on a CV < 15 %.
Levels in samples shall be estimated by comparison of the test response with a calibration curve of TCDD (or PCB 126 or a PCDD/PCDF/dioxin-like PCB standard mixture) to calculate the BEQ level in the extract and subsequently in the sample. Calibration curves shall contain 8 to 12 concentrations (at least in duplicates), with enough concentrations in the lower part of the curve (working range). Special attention shall be paid to the quality of the curve-fit in the working range. As such, the R 2 value is of little or no value in estimating the goodness of fit in non-linear regression. A better fit shall be achieved by minimising the difference between calculated and observed levels in the working range of the curve, for example by minimising the sum of squared residuals.The estimated level in the sample extract shall be subsequently corrected for the BEQ level calculated for a matrix or solvent blank sample (to account for impurities from solvents and chemicals used), and the apparent recovery (calculated from the BEQ level of suitable reference samples with representative congener patterns around the maximum level or action threshold). To perform a recovery correction, the apparent recovery shall be within the required range (see point 7.1.4). Reference samples used for recovery correction shall comply with the requirements laid down in point 7.2.
(1) from the lower band of the 95 % prediction interval at the decision limit of the confirmatory method.(2) from multiple analysis of samples (n ≥ 6) contaminated at the decision limit of the confirmatory method as the lower end-point of the data distribution (represented in the figure by a bell-shaped curve) at the corresponding mean BEQ value.
(a) two specific ions for HRMS; (b) three specific ions for LRMS; (c) two specific precursor ions, each with one specific corresponding transition product ion for for MS-MS.
(a) results shall be corrected for recoveries of internal standards; (b) recoveries of isotope-labelled internal standards shall be between 60 and 120 %; (c) lower or higher recoveries for individual congeners with a contribution to the sum of non-dioxin-like PCBs below 10 % are acceptable.
(a) recovery of internal standard(s) shall be controlled for every sample; (b) recoveries of internal standard(s) shall be between 60 and 120 %; (c) results shall be corrected for recoveries of internal standards.
Isotope dilution mass spectrometry | Other techniques | |
---|---|---|
Trueness | – 20 to + 20 % | – 30 to + 30 % |
Intermediate precision (RSD %) | ≤ 15 % | ≤ 20 % |
Difference between upper and lower-bound calculation | ≤ 20 % | ≤ 20 % |
----------------------
Congener | TEF value | Congener | TEF value |
---|---|---|---|
Dibenzo-p-dioxins ("PCDDs") and Dibenzo-p-furans '("PCDFs") | |||
2,3,7,8-TCDD | 1 | ||
1,2,3,7,8-PeCDD | 1 | Non-ortho PCBs | |
1,2,3,4,7,8-HxCDD | 0,1 | PCB 77 | 0,0001 |
1,2,3,6,7,8-HxCDD | 0,1 | PCB 81 | 0,0003 |
1,2,3,7,8,9-HxCDD | 0,1 | PCB 126 | 0,1 |
1,2,3,4,6,7,8-HpCDD | 0,01 | PCB 169 | 0,03 |
OCDD | 0,0003 | Mono-ortho PCBs | |
2,3,7,8-TCDF | 0,1 | PCB 105 | 0,00003 |
1,2,3,7,8-PeCDF | 0,03 | PCB 114 | 0,00003 |
2,3,4,7,8-PeCDF | 0,3 | PCB 118 | 0,00003 |
1,2,3,4,7,8-HxCDF | 0,1 | PCB 123 | 0,00003 |
1,2,3,6,7,8-HxCDF | 0,1 | PCB 156 | 0,00003 |
1,2,3,7,8,9-HxCDF | 0,1 | PCB 157 | 0,00003 |
2,3,4,6,7,8-HxCDF | 0,1 | PCB 167 | 0,00003 |
1,2,3,4,6,7,8-HpCDF | 0,01 | PCB 189 | 0,00003 |
1,2,3,4,7,8,9-HpCDF | 0,01 | ||
OCDF | 0,0003 |