(a) products for general laboratory use or research-use only products, unless such products, in view of their characteristics, are specifically intended by their manufacturer to be used for in vitro diagnostic examination;(b) invasive sampling products or products which are directly applied to the human body for the purpose of obtaining a specimen; (c) internationally certified reference materials; (d) materials used for external quality assessment schemes.
Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU (Text with EEA relevance. )
Modified by
- Regulation (EU) 2022/112 of the European Parliament and of the Councilof 25 January 2022amending Regulation (EU) 2017/746 as regards transitional provisions for certain in vitro diagnostic medical devices and the deferred application of conditions for in-house devices(Text with EEA relevance), 32022R0112, January 28, 2022
- Commission Delegated Regulation (EU) 2023/503of 1 December 2022amending Regulation (EU) 2017/746 of the European Parliament and of the Council as regards the frequency of complete re-assessments of notified bodies(Text with EEA relevance), 32023R0503, March 8, 2023
- Regulation (EU) 2023/607 of the European Parliament and of the Councilof 15 March 2023amending Regulations (EU) 2017/745 and (EU) 2017/746 as regards the transitional provisions for certain medical devices and in vitro diagnostic medical devices(Text with EEA relevance), 32023R0607, March 20, 2023
- Regulation (EU) 2024/1860 of the European Parliament and of the Councilof 13 June 2024amending Regulations (EU) 2017/745 and (EU) 2017/746 as regards a gradual roll-out of Eudamed, the obligation to inform in case of interruption or discontinuation of supply, and transitional provisions for certain in vitro diagnostic medical devices(Text with EEA relevance), 32024R1860, July 9, 2024
Corrected by
- Corrigendum to Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU, 32017R0746R(02), May 3, 2019
- Corrigendum to Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU, 32017R0746R(03), December 27, 2019
(1) "medical device" means "medical device" as defined in point (1) of Article 2 of Regulation (EU) 2017/745; (2) " in vitro diagnostic medical device" means any medical device which is a reagent, reagent product, calibrator, control material, kit, instrument, apparatus, piece of equipment, software or system, whether used alone or in combination, intended by the manufacturer to be usedin vitro for the examination of specimens, including blood and tissue donations, derived from the human body, solely or principally for the purpose of providing information on one or more of the following:(a) concerning a physiological or pathological process or state; (b) concerning congenital physical or mental impairments; (c) concerning the predisposition to a medical condition or a disease; (d) to determine the safety and compatibility with potential recipients; (e) to predict treatment response or reactions; (f) to define or monitoring therapeutic measures.
Specimen receptacles shall also be deemed to be in vitro diagnostic medical devices;(3) "specimen receptacle" means a device, whether of a vacuum-type or not, specifically intended by its manufacturer for the primary containment and preservation of specimens derived from the human body for the purpose of in vitro diagnostic examination;(4) "accessory for an in vitro diagnostic medical device" means an article which, whilst not being itself anin vitro diagnostic medical device, is intended by its manufacturer to be used together with one or several particularin vitro diagnostic medical device(s) to specifically enable thein vitro diagnostic medical device(s) to be used in accordance with its/their intended purpose(s) or to specifically and directly assist the medical functionality of thein vitro diagnostic medical device(s) in terms of its/their intended purpose(s);(5) "device for self-testing" means any device intended by the manufacturer to be used by lay persons, including devices used for testing services offered to lay persons by means of information society services; (6) "device for near-patient testing" means any device that is not intended for self-testing but is intended to perform testing outside a laboratory environment, generally near to, or at the side of, the patient by a health professional; (7) "companion diagnostic" means a device which is essential for the safe and effective use of a corresponding medicinal product to: (a) identify, before and/or during treatment, patients who are most likely to benefit from the corresponding medicinal product; or (b) identify, before and/or during treatment, patients likely to be at increased risk of serious adverse reactions as a result of treatment with the corresponding medicinal product;
(8) "generic device group" means a set of devices having the same or similar intended purposes or a commonality of technology allowing them to be classified in a generic manner not reflecting specific characteristics; (9) "single-use device" means a device that is intended to be used during a single procedure; (10) "falsified device" means any device with a false presentation of its identity and/or of its source and/or its CE marking certificates or documents relating to CE marking procedures. This definition does not include unintentional non-compliance and is without prejudice to infringements of intellectual property rights; (11) "kit" means a set of components that are packaged together and intended to be used to perform a specific in vitro diagnostic examination, or a part thereof;(12) "intended purpose" means the use for which a device is intended according to the data supplied by the manufacturer on the label, in the instructions for use or in promotional or sales materials or statements or as specified by the manufacturer in the performance evaluation; (13) "label" means the written, printed or graphic information appearing either on the device itself, or on the packaging of each unit or on the packaging of multiple devices; (14) "instructions for use" means the information provided by the manufacturer to inform the user of a device's intended purpose and proper use and of any precautions to be taken; (15) "Unique Device Identifier" ("UDI") means a series of numeric or alphanumeric characters that is created through internationally accepted device identification and coding standards and that allows unambiguous identification of specific devices on the market; (16) "risk" means the combination of the probability of occurrence of harm and the severity of that harm; (17) "benefit-risk determination" means the analysis of all assessments of benefit and risk of possible relevance for the use of the device for the intended purpose, when used in accordance with the intended purpose given by the manufacturer; (18) "compatibility" is the ability of a device, including software, when used together with one or more other devices in accordance with its intended purpose, to: (a) perform without losing or compromising the ability to perform as intended, and/or (b) integrate and/or operate without the need for modification or adaption of any part of the combined devices, and/or (c) be used together without conflict/interference or adverse reaction;
(19) "interoperability" is the ability of two or more devices, including software, from the same manufacturer or from different manufacturers, to: (a) exchange information and use the information that has been exchanged for the correct execution of a specified function without changing the content of the data, and/or (b) communicate with each other, and/or (c) work together as intended;
(20) "making available on the market" means any supply of a device, other than a device for performance study, for distribution, consumption or use on the Union market in the course of a commercial activity, whether in return for payment or free of charge; (21) "placing on the market" means the first making available of a device, other than a device for performance study, on the Union market; (22) "putting into service" means the stage at which a device, other than a device for performance study, has been made available to the final user as being ready for use on the Union market for the first time for its intended purpose; (23) "manufacturer" means a natural or legal person who manufactures or fully refurbishes a device or has a device designed, manufactured or fully refurbished, and markets that device under its name or trade mark; (24) "fully refurbishing", for the purposes of the definition of manufacturer, means the complete rebuilding of a device already placed on the market or put into service, or the making of a new device from used devices, to bring it into conformity with this Regulation, combined with the assignment of a new lifetime to the refurbished device; (25) "authorised representative" means any natural or legal person established within the Union who has received and accepted a written mandate from a manufacturer, located outside the Union, to act on the manufacturer's behalf in relation to specified tasks with regard to the latter's obligations under this Regulation; (26) "importer" means any natural or legal person established within the Union that places a device from a third country on the Union market; (27) "distributor" means any natural or legal person in the supply chain, other than the manufacturer or the importer, that makes a device available on the market, up until the point of putting into service; (28) "economic operator" means a manufacturer, an authorised representative, an importer or a distributor; (29) "health institution" means an organisation the primary purpose of which is the care or treatment of patients or the promotion of public health; (30) "user" means any healthcare professional or lay person who uses a device; (31) "lay person" means an individual who does not have formal education in a relevant field of healthcare or medical discipline; (32) "conformity assessment" means the process demonstrating whether the requirements of this Regulation relating to a device have been fulfilled; (33) "conformity assessment body" means a body that performs third-party conformity assessment activities including calibration, testing, certification and inspection; (34) "notified body" means a conformity assessment body designated in accordance with this Regulation; (35) "CE marking of conformity" or "CE marking" means a marking by which a manufacturer indicates that a device is in conformity with the applicable requirements set out in this Regulation and other applicable Union harmonisation legislation providing for its affixing; (36) "clinical evidence" means clinical data and performance evaluation results, pertaining to a device of a sufficient amount and quality to allow a qualified assessment of whether the device is safe and achieves the intended clinical benefit(s), when used as intended by the manufacturer; (37) "clinical benefit" means the positive impact of a device related to its function, such as that of screening, monitoring, diagnosis or aid to diagnosis of patients, or a positive impact on patient management or public health; (38) "scientific validity of an analyte" means the association of an analyte with a clinical condition or a physiological state; (39) "performance of a device" means the ability of a device to achieve its intended purpose as claimed by the manufacturer. It consists of the analytical and, where applicable, the clinical performance supporting that intended purpose; (40) "analytical performance" means the ability of a device to correctly detect or measure a particular analyte; (41) "clinical performance" means the ability of a device to yield results that are correlated with a particular clinical condition or a physiological or pathological process or state in accordance with the target population and intended user; (42) "performance study" means a study undertaken to establish or confirm the analytical or clinical performance of a device; (43) "performance study plan" means a document that describes the rationale, objectives, design methodology, monitoring, statistical considerations, organisation and conduct of a performance study; (44) "performance evaluation" means an assessment and analysis of data to establish or verify the scientific validity, the analytical and, where applicable, the clinical performance of a device; (45) "device for performance study" means a device intended by the manufacturer to be used in a performance study. A device intended to be used for research purposes, without any medical objective, shall not be deemed to be a device for performance study; (46) "interventional clinical performance study" means a clinical performance study where the test results may influence patient management decisions and/or may be used to guide treatment; (47) "subject" means an individual who participates in a performance study and whose specimen(s) undergo in vitro examination by a device for performance study and/or by a device used for control purposes;(48) "investigator" means an individual responsible for the conduct of a performance study at a performance study site; (49) "diagnostic specificity" means the ability of a device to recognise the absence of a target marker associated with a particular disease or condition; (50) "diagnostic sensitivity" means the ability of a device to identify the presence of a target marker associated with a particular disease or condition; (51) "predictive value" means the probability that a person with a positive device test result has a given condition under investigation, or that a person with a negative device test result does not have a given condition; (52) "positive predictive value" means the ability of a device to separate true positive results from false positive results for a given attribute in a given population; (53) "negative predictive value" means the ability of a device to separate true negative results from false negative results for a given attribute in a given population; (54) "likelihood ratio" means the likelihood of a given result arising in an individual with the target clinical condition or physiological state compared to the likelihood of the same result arising in an individual without that clinical condition or physiological state; (55) "calibrator" means a measurement reference material used in the calibration of a device; (56) "control material" means a substance, material or article intended by its manufacturer to be used to verify the performance characteristics of a device; (57) "sponsor" means any individual, company, institution or organisation which takes responsibility for the initiation, for the management and setting up of the financing of the performance study; (58) "informed consent" means a subject's free and voluntary expression of his or her willingness to participate in a particular performance study, after having been informed of all aspects of the performance study that are relevant to the subject's decision to participate or, in the case of minors and of incapacitated subjects, an authorisation or agreement from their legally designated representative to include them in the performance study; (59) "ethics committee" means an independent body established in a Member State in accordance with the law of that Member State and empowered to give opinions for the purposes of this Regulation, taking into account the views of laypersons, in particular patients or patients' organisations; (60) "adverse event" means any untoward medical occurrence, inappropriate patient management decision, unintended disease or injury or any untoward clinical signs, including an abnormal laboratory finding, in subjects, users or other persons, in the context of a performance study, whether or not related to the device for performance study; (61) "serious adverse event" means any adverse event that led to any of the following: (a) a patient management decision resulting in death or an imminent life-threatening situation for the individual being tested, or in the death of the individual's offspring, (b) death, (c) serious deterioration in the health of the individual being tested or the recipient of tested donations or materials, that resulted in any of the following: (i) life-threatening illness or injury, (ii) permanent impairment of a body structure or a body function, (iii) hospitalisation or prolongation of patient hospitalisation, (iv) medical or surgical intervention to prevent life-threatening illness or injury or permanent impairment to a body structure or a body function, (v) chronic disease,
(d) foetal distress, foetal death or a congenital physical or mental impairment or birth defect;
(62) "device deficiency" means any inadequacy in the identity, quality, durability, reliability, safety or performance of a device for performance study, including malfunction, use errors or inadequacy in information supplied by the manufacturer; (63) "post-market surveillance" means all activities carried out by manufacturers in cooperation with other economic operators to institute and keep up to date a systematic procedure to proactively collect and review experience gained from devices they place on the market, make available on the market or put into service for the purpose of identifying any need to immediately apply any necessary corrective or preventive actions; (64) "market surveillance" means the activities carried out and measures taken by public authorities to check and ensure that devices comply with the requirements set out in the relevant Union harmonisation legislation and do not endanger health, safety or any other aspect of public interest protection; (65) "recall" means any measure aimed at achieving the return of a device that has already been made available to the end user; (66) "withdrawal" means any measure aimed at preventing a device in the supply chain from being further made available on the market; (67) "incident" means any malfunction or deterioration in the characteristics or performance of a device made available on the market, including use-error due to ergonomic features, as well as any inadequacy in the information supplied by the manufacturer and any harm as a consequence of a medical decision, action taken or not taken on the basis of information or result(s) provided by the device; (68) "serious incident" means any incident that directly or indirectly led, might have led or might lead to any of the following: (a) the death of a patient, user or other person, (b) the temporary or permanent serious deterioration of a patient's, user's or other person's state of health, (c) a serious public health threat;
(69) "serious public health threat" means an event which could result in imminent risk of death, serious deterioration in a person's state of health, or serious illness, that may require prompt remedial action, and that may cause significant morbidity or mortality in humans, or that is unusual or unexpected for the given place and time; (70) "corrective action" means action taken to eliminate the cause of a potential or actual non-conformity or other undesirable situation; (71) "field safety corrective action" means corrective action taken by a manufacturer for technical or medical reasons to prevent or reduce the risk of a serious incident in relation to a device made available on the market; (72) "field safety notice" means a communication sent by a manufacturer to users or customers in relation to a field safety corrective action; (73) "harmonised standard" means a European standard as defined in point (1)(c) of Article 2 of Regulation (EU) No 1025/2012; (74) "common specifications" (CS) means a set of technical and/or clinical requirements, other than a standard, that provides a means of complying with the legal obligations applicable to a device, process or system.
(a) the devices are not transferred to another legal entity; (b) manufacture and use of the devices occur under appropriate quality management systems; (c) the laboratory of the health institution is compliant with standard EN ISO 15189 or where applicable national provisions, including national provisions regarding accreditation; (d) the health institution justifies in its documentation that the target patient group's specific needs cannot be met, or cannot be met at the appropriate level of performance by an equivalent device available on the market; (e) the health institution provides information upon request on the use of such devices to its competent authority, which shall include a justification of their manufacturing, modification and use; (f) the health institution draws up a declaration which it shall make publicly available, including: (i) the name and address of the manufacturing health institution, (ii) the details necessary to identify the devices, (iii) a declaration that the devices meet the general safety and performance requirements set out in Annex I to this Regulation and, where applicable, information on which requirements are not fully met with a reasoned justification therefor;
(g) as regards class D devices in accordance with the rules set out in Annex VIII, the health institution draws up documentation that makes it possible to have an understanding of the manufacturing facility, the manufacturing process, the design and performance data of the devices, including the intended purpose, and that is sufficiently detailed to enable the competent authority to ascertain that the general safety and performance requirements set out in Annex I to this Regulation are met. Member States may apply this provision also to class A, B or C devices in accordance with the rules set out in Annex VIII; (h) the health institution takes all necessary measures to ensure that all devices are manufactured in accordance with the documentation referred to in point (g); and (i) the health institution reviews experience gained from clinical use of the devices and takes all necessary corrective actions.
(a) ascribing functions and properties to the device which the device does not have; (b) creating a false impression regarding treatment or diagnosis, functions or properties which the device does not have; (c) failing to inform the user or the patient of a likely risk associated with the use of the device in line with its intended purpose; (d) suggesting uses for the device other than those stated to form part of the intended purpose for which the conformity assessment was carried out.
(a) a strategy for regulatory compliance, including compliance with conformity assessment procedures and procedures for management of modifications to the devices covered by the system; (b) identification of applicable general safety and performance requirements and exploration of options to address those requirements; (c) responsibility of the management; (d) resource management, including selection and control of suppliers and sub-contractors; (e) risk management as set out in Section 3 of Annex I; (f) performance evaluation, in accordance with Article 56 and Annex XIII, including PMPF; (g) product realisation, including planning, design, development, production and service provision; (h) verification of the UDI assignments made in accordance with Article 24(3) to all relevant devices and ensuring consistency and validity of information provided in accordance with Article 26; (i) setting-up, implementation and maintenance of a post-market surveillance system, in accordance with Article 78; (j) handling communication with competent authorities, notified bodies, other economic operators, customers and/or other stakeholders; (k) processes for reporting of serious incidents and field safety corrective actions in the context of vigilance; (l) management of corrective and preventive actions and verification of their effectiveness; (m) processes for monitoring and measurement of output, data analysis and product improvement.
(a) verify that the EU declaration of conformity and technical documentation have been drawn up and, where applicable, that an appropriate conformity assessment procedure has been carried out by the manufacturer; (b) keep available a copy of the technical documentation, the EU declaration of conformity and, if applicable, a copy of the relevant certificate, including any amendments and supplements, issued in accordance with Article 51, at the disposal of competent authorities for the period referred to in Article 10(7); (c) comply with the registration obligations laid down in Article 28 and verify that the manufacturer has complied with the registration obligations laid down in Article 26; (d) in response to a request from a competent authority, provide that competent authority with all the information and documentation necessary to demonstrate the conformity of a device, in an official Union language determined by the Member State concerned; (e) forward to the manufacturer any request by a competent authority of the Member State in which the authorised representative has its registered place of business for samples, or access to a device and verify that the competent authority receives the samples or is given access to the device; (f) cooperate with the competent authorities on any preventive or corrective action taken to eliminate or, if that is not possible, mitigate the risks posed by devices; (g) immediately inform the manufacturer about complaints and reports from healthcare professionals, patients and users about suspected incidents related to a device for which they have been designated; (h) terminate the mandate if the manufacturer acts contrary to its obligations under this Regulation.
(a) the date of termination of the mandate of the outgoing authorised representative and date of beginning of the mandate of the incoming authorised representative; (b) the date until which the outgoing authorised representative may be indicated in the information supplied by the manufacturer, including any promotional material; (c) the transfer of documents, including confidentiality aspects and property rights; (d) the obligation of the outgoing authorised representative after the end of the mandate to forward to the manufacturer or incoming authorised representative any complaints or reports from healthcare professionals, patients or users about suspected incidents related to a device for which it had been designated as authorised representative.
(a) the device has been CE marked and that the EU declaration of conformity of the device has been drawn up; (b) a manufacturer is identified and that an authorised representative in accordance with Article 11 has been designated by the manufacturer; (c) the device is labelled in accordance with this Regulation and accompanied by the required instructions for use; (d) where applicable, a UDI has been assigned by the manufacturer in accordance with Article 24.
(a) the device has been CE marked and the EU declaration of conformity of the device has been drawn up; (b) the device is accompanied by the information to be supplied by the manufacturer in accordance with Article 10(10); (c) for imported devices, the importer has complied with the requirements set out in Article 13(3); (d) that, where applicable, a UDI has been assigned by the manufacturer.
(a) a diploma, certificate or other evidence of formal qualification, awarded on completion of a university degree or of a course of study recognised as equivalent by the Member State concerned, in law, medicine, pharmacy, engineering or another relevant scientific discipline, and at least one year of professional experience in regulatory affairs or in quality management systems relating to in vitro diagnostic medical devices;(b) four years of professional experience in regulatory affairs or in quality management systems relating to in vitro diagnostic medical devices.
(a) the conformity of the devices is appropriately checked, in accordance with the quality management system under which the devices are manufactured, before a device is released; (b) the technical documentation and the EU declaration of conformity are drawn up and kept up-to-date; (c) the post-market surveillance obligations are complied with in accordance with Article 10(9); (d) the reporting obligations referred to in Articles 82 to 86 are fulfilled; (e) in the case of devices for performance studies intended to be used in the context of interventional clinical performance studies or other performance studies involving risks for the subjects, the statement referred to in Section 4.1 of Annex XIV is issued.
(a) a diploma, certificate or other evidence of formal qualification, awarded on completion of a university degree or of a course of study recognised as equivalent by the Member State concerned, in law, medicine, pharmacy, engineering or another relevant scientific discipline, and at least one year of professional experience in regulatory affairs or in quality management systems relating to in vitro diagnostic medical devices;(b) four years of professional experience in regulatory affairs or in quality management systems relating to in vitro diagnostic medical devices.
(a) makes available on the market a device under its own name, registered trade name or registered trade mark, except in cases where a distributor or importer enters into an agreement with a manufacturer whereby the manufacturer is identified as such on the label and is responsible for meeting the requirements placed on manufacturers in this Regulation; (b) changes the intended purpose of a device already placed on the market or put into service; (c) modifies a device already placed on the market or put into service in such a way that compliance with the applicable requirements may be affected.
(a) provision, including translation, of the information supplied by the manufacturer, in accordance with Section 20 of Annex I, relating to a device already placed on the market and of further information which is necessary in order to market the device in the relevant Member State; (b) changes to the outer packaging of a device already placed on the market, including a change of pack size, if the repackaging is necessary in order to market the device in the relevant Member State and if it is carried out in such conditions that the original condition of the device cannot be affected by it. In the case of devices placed on the market in sterile condition, it shall be presumed that the original condition of the device is adversely affected if the packaging that is necessary for maintaining the sterile condition is opened, damaged or otherwise negatively affected by the repackaging.
(a) any economic operator to whom they have directly supplied a device; (b) any economic operator who has directly supplied them with a device; (c) any health institution or healthcare professional to which they have directly supplied a device.
(a) production of a UDI that comprises the following: (i) a UDI device identifier ("UDI-DI") specific to a manufacturer and a device, providing access to the information laid down in Part B of Annex VI; (ii) a UDI production identifier ("UDI-PI") that identifies the unit of device production and if applicable the packaged devices, as specified in Part C of Annex VI;
(b) placing of the UDI on the label of the device or on its packaging; (c) storage of the UDI by economic operators, health institutions and healthcare professionals, in accordance with the conditions laid down in paragraphs 8 and 9 respectively; (d) establishment of an electronic system for Unique Device Identification ("UDI database") in accordance with Article 28 of Regulation (EU) 2017/745.
(a) the entity is an organisation with legal personality; (b) its system for the assignment of UDIs is adequate to identify a device throughout its distribution and use in accordance with the requirements of this Regulation; (c) its system for the assignment of UDIs conforms to the relevant international standards; (d) the entity gives access to its system for the assignment of UDIs to all interested users in accordance with a set of predetermined and transparent terms and conditions; (e) the entity undertakes to do the following: (i) operate its system for the assignment of UDIs for at least 10 years after its designation; (ii) make available to the Commission and to the Member States, upon request, information concerning its system for the assignment of UDIs; (iii) remain in compliance with the criteria for designation and the terms of designation.
(a) amending the list of information set out in Part B of Annex VI in the light of technical progress; and (b) amending Annex VI in the light of international developments and technical progress in the field of Unique Device Identification.
(a) determining the devices, categories or groups of devices to which the obligation laid down in paragraph 8 is to apply; (b) specifying the data to be included in the UDI-PI of specific devices or device groups.
(a) confidentiality and data protection as referred to in Articles 102 and 103 respectively; (b) the risk-based approach; (c) the cost-effectiveness of the measures; (d) the convergence of UDI systems developed at international level; (e) the need to avoid duplications in the UDI system; (f) the needs of the health care systems of the Member States, and where possible, compatibility with other medical device identification systems that are used by stakeholders.
(a) the identification of the device and the manufacturer, including the Basic UDI-DI and, if already issued, the SRN; (b) the intended purpose of the device and any indications, contra-indications and target populations; (c) a description of the device, including a reference to previous generation(s) or variants if such exist, and a description of the differences, as well as, where relevant, a description of any accessories, other devices and products, which are intended to be used in combination with the device; (d) reference to any harmonised standards and CS applied; (e) the summary of the performance evaluation as referred to in Annex XIII, and relevant information on the PMPF; (f) the metrological traceability of assigned values; (g) suggested profile and training for users; (h) information on any residual risks and any undesirable effects, warnings and precautions.
(a) the electronic system for registration of devices referred to in Article 26; (b) the UDI database referred to in Article 25; (c) the electronic system on registration of economic operators referred to in Article 27; (d) the electronic system on notified bodies and on certificates referred to in Article 52; (e) the electronic system on performance studies referred to in Article 69, (f) the electronic system on vigilance and post-market surveillance referred to in Article 87; (g) the electronic system on market surveillance referred to in Article 95.
the result of the assessment, confirmation that the corrective and preventive actions have been appropriately addressed and, where required, implemented, any remaining diverging opinion with the joint assessment team, and, where applicable, the recommended scope of designation.
(a) assess the impact on the certificates issued by the notified body; (b) submit a report on its findings to the Commission and the other Member States within three months of having notified the changes to the designation; (c) require the notified body to suspend or withdraw, within a reasonable period of time determined by the authority, any certificates which were unduly issued to ensure the safety of devices on the market; (d) enter in the electronic system referred to in Article 52 information in relation to certificates of which it has required their suspension or withdrawal; (e) inform the competent authority for in vitro diagnostic medical devices of the Member State in which the manufacturer has its registered place of business through the electronic system referred to in Article 52 of the certificates for which it has required suspension or withdrawal. That competent authority shall take the appropriate measures, where necessary to avoid a potential risk to the health or safety of patients, users or others.
(a) the authority responsible for notified bodies has confirmed, within one month of the suspension or restriction, that there is no safety issue in relation to certificates affected by the suspension or restriction and the authority responsible for notified bodies has outlined a timeline and actions anticipated to remedy the suspension or restriction; or (b) the authority responsible for notified bodies has confirmed that no certificates relevant to the suspension will be issued, amended or re-issued during the course of the suspension or restriction, and states whether the notified body has the capability of continuing to monitor, and remain responsible for, existing certificates issued for the period of the suspension or restriction. In the event that the authority responsible for notified bodies determines that the notified body does not have the capability to support existing certificates issued, the manufacturer shall provide, to the competent authority for in vitro diagnostic medical devices of the Member State in which the manufacturer of the device covered by the certificate has its registered place of business, within three months of the suspension or restriction, a written confirmation that another qualified notified body is temporarily assuming the functions of the notified body to monitor and remain responsible for the certificates during the period of suspension or restriction.
(a) where the competent authority for in vitro diagnostic medical devices of the Member State in which the manufacturer of the device covered by the certificate has its registered place of business has confirmed that there is no safety issue associated with the devices in question; and(b) another notified body has confirmed in writing that it will assume immediate responsibilities for those devices and will have completed assessment of them within twelve months of the withdrawal of the designation.
(a) development of best practice documents relating to the activities of the authorities responsible for notified bodies; (b) development of guidance documents for notified bodies in relation to the implementation of this Regulation; (c) training and qualification of the experts referred to in Article 36; (d) monitoring of trends relating to changes to notified body designations and notifications, and trends in certificate withdrawals and transfers between notified bodies; (e) monitoring of the application and applicability of scope codes referred to in Article 38(13); (f) development of a mechanism for peer reviews between authorities and the Commission; (g) methods of communication to the public on the monitoring and surveillance activities of authorities and the Commission on notified bodies.
(a) application of Annex VIII to a given device, or category or group of devices, with a view to determining the classification of such devices; (b) that a device, or category or group of devices, shall for reasons of public health based on new scientific evidence, or based on any information which becomes available in the course of the vigilance and market surveillance activities be reclassified, by way of derogation from Annex VIII.
(a) the frequency and the sampling basis of the assessment of the technical documentation on a representative basis as set out in third paragraph of Section 2.3. and in Section 3.5 of Annex IX, in the case of class C devices; (b) the minimum frequency of unannounced on-site audits and sample tests to be conducted by notified bodies in accordance with Section 3.4 of Annex IX, taking into account the risk-class and the type of device; (c) the frequency of samples of the manufactured devices or batches of class D devices to be sent to an EU reference laboratory designated under Article 100 in accordance with Section 4.12 of Annex IX and Section 5.1 of Annex XI; or (d) the physical, laboratory or other tests to be carried out by notified bodies in the context of sample tests, assessment of technical documentation and type examination in accordance with Sections 3.4 and 4.3 of Annex IX and points (f) and (g) of Section 3. of Annex X.
(a) the list of subsidiaries referred to in Article 33(2); (b) the list of experts referred to in Article 36(2); (c) the information relating to the notification referred to in Article 38(10) and the amended notifications referred to in Article 42(2); (d) the list of notified bodies referred to in Article 39(2); (e) the summary of the report referred to in Article 40(12); (f) the notifications for conformity assessments and certificates referred to in Article 50(1); (g) withdrawal or refusals of applications for the certificates as referred to in Article 49(2) and Section 4.3 of Annex VII; (h) the information regarding certificates referred to in Article 51(5); (i) the summary of safety and performance referred to in Article 29.
(a) the date on which the certificates issued by the outgoing notified body become invalid; (b) the date until which the identification number of the outgoing notified body may be indicated in the information supplied by the manufacturer, including any promotional material; (c) the transfer of documents, including confidentiality aspects and property rights; (d) the date after which the conformity assessment tasks of the outgoing notified body is assigned to the incoming notified body; (e) the last serial number or lot number for which the outgoing notified body is responsible.
(a) scientific validity; (b) analytical performance; (c) clinical performance.
(a) in which surgically invasive sample-taking is done only for the purpose of the performance study; (b) that is an interventional clinical performance study as defined in point (46) of Article 2; or (c) where the conduct of the study involves additional invasive procedures or other risks for the subjects of the studies,
(a) the performance study is the subject of an authorisation by the Member State(s) in which the performance study is to be conducted, in accordance with this Regulation, unless otherwise stated; (b) an ethics committee, set up in accordance with national law, has not issued a negative opinion in relation to the performance study, which is valid for that entire Member State under its national law; (c) the sponsor or its legal representative or a contact person pursuant to paragraph 4 is established in the Union; (d) vulnerable populations and subjects are appropriately protected in accordance with Articles 59 to 64; (e) the anticipated benefits to the subjects or to public health justify the foreseeable risks and inconveniences and compliance with this condition is constantly monitored; (f) the subject or, where the subject is not able to give informed consent, his or her legally designated representative has given informed consent, in accordance with Article 59; (g) the subject or, where the subject is not able to give informed consent, his or her legally designated representative, has been provided with the contact details of an entity where further information can be received in case of need; (h) the rights of the subject to physical and mental integrity, to privacy and to the protection of the data concerning him or her in accordance with Directive 95/46/EC are safeguarded; (i) the performance study has been designed to involve as little pain, discomfort, fear and any other foreseeable risk as possible for the subjects, and both the risk threshold and the degree of distress are specifically defined in the performance study plan and constantly monitored; (j) the medical care provided to the subjects is the responsibility of an appropriately qualified medical doctor or, where appropriate, any other person entitled by national law to provide the relevant patient care under performance study conditions; (k) no undue influence, including that of a financial nature, is exerted on the subject, or, where applicable, on his or her legally designated representatives, to participate in the performance study; (l) where appropriate, biological safety testing reflecting the latest scientific knowledge or any other test deemed necessary in the light of the device's intended purpose has been conducted; (m) in the case of clinical performance studies, the analytical performance has been demonstrated, taking into consideration the state of the art; (n) in the case of interventional clinical performance studies, the analytical performance and scientific validity has been demonstrated, taking into consideration the state of the art. Where, for companion diagnostics, the scientific validity is not established, the scientific rationale for the use of the biomarker shall be provided; (o) the technical safety of the device with regard to its use has been proven, taking into consideration the state of the art as well as provisions in the field of occupational safety and accident prevention; (p) the requirements of Annex XIV are fulfilled.
(a) enable the subject or his or her legally designated representative to understand: (i) the nature, objectives, benefits, implications, risks and inconveniences of the performance study; (ii) the subject's rights and guarantees regarding his or her protection, in particular his or her right to refuse to participate in and the right to withdraw from the performance study at any time without any resulting detriment and without having to provide any justification; (iii) the conditions under which the performance study is to be conducted, including the expected duration of the subject's participation in the performance study; and (iv) the possible treatment alternatives, including the follow-up measures if the participation of the subject in the performance study is discontinued;
(b) be kept comprehensive, concise, clear, relevant, and understandable to the subject or his or her legally designated representative; (c) be provided in a prior interview with a member of the investigating team who is appropriately qualified under national law; and (d) include information about the applicable damage compensation system referred to in Article 65; (e) include the Union-wide unique single identification number for the performance study referred to in Article 66(1) and information about the availability of the performance study results in accordance with paragraph 6 of this Article.
(a) the informed consent of their legally designated representative has been obtained; (b) the incapacitated subjects have received the information referred to in Article 59(2) in a way that is adequate in view of their capacity to understand it; (c) the explicit wish of an incapacitated subject who is capable of forming an opinion and assessing the information referred to in Article 59(2) to refuse participation in, or to withdraw from, the performance study at any time, is respected by the investigator; (d) no incentives or financial inducements are given to subjects or their legally designated representatives, except for compensation for expenses and loss of earnings directly related to the participation in the performance study; (e) the performance study is essential with respect to incapacitated subjects and data of comparable validity cannot be obtained in performance studies on persons able to give informed consent, or by other research methods; (f) the performance study relates directly to a medical condition from which the subject suffers; (g) there are scientific grounds for expecting that participation in the performance study will produce: (i) a direct benefit to the incapacitated subject outweighing the risks and burdens involved; or (ii) some benefit for the population represented by the incapacitated subject concerned when the performance study will pose only minimal risk to, and will impose minimal burden on, the incapacitated subject concerned in comparison with the standard treatment of the incapacitated subject's condition.
(a) the informed consent of their legally designated representative has been obtained; (b) the minors have received the information referred to in Article 59(2) in a way adapted to their age and mental maturity and from investigators or members of the investigating team who are trained or experienced in working with children; (c) the explicit wish of a minor who is capable of forming an opinion and assessing the information referred to in Article 59(2) to refuse participation in, or to withdraw from, the performance study at any time, is respected by the investigator; (d) no incentives or financial inducements are given to subjects or their legally designated representatives, except for compensation for expenses and loss of earnings directly related to the participation in the performance study; (e) the performance study is intended to investigate treatments for a medical condition that only occurs in minors or the performance study is essential with respect to minors to validate data obtained in performance studies on persons able to give informed consent or by other research methods; (f) the performance study either relates directly to a medical condition from which the minor concerned suffers or is of such a nature that it can only be carried out on minors; (g) there are scientific grounds for expecting that participation in the performance study will produce: (i) a direct benefit to the minor subject outweighing the risks and burdens involved; or (ii) some benefit for the population represented by the minor concerned when the performance study will pose only minimal risk to, and will impose minimal burden on, the minor concerned in comparison with the standard treatment of the minor's condition;
(h) the minor shall take part in the informed consent procedure in a way adapted to his or her age and mental maturity; (i) if during a performance study the minor reaches the age of legal competence to give informed consent as defined in the national law, his or her express informed consent shall be obtained before that subject can continue to participate in the performance study.
(a) the performance study has the potential to produce a direct benefit for the pregnant or breastfeeding woman concerned, or her embryo, foetus or child after birth, outweighing the risks and burdens involved; (b) if such a performance study has no direct benefit for the pregnant or breastfeeding woman concerned, or her embryo, foetus or child after birth, it can be conducted only if: (i) a performance study of comparable effectiveness cannot be carried out on women who are not pregnant or breastfeeding; (ii) the performance study contributes to the attainment of results capable of benefitting pregnant or breastfeeding women or other women in relation to reproduction or other embryos, foetuses or children; and (iii) the performance study poses a minimal risk to, and imposes a minimal burden on, the pregnant or breastfeeding woman concerned, her embryo, foetus or child after birth;
(c) where research is undertaken on breastfeeding women, particular care is taken to avoid any adverse impact on the health of the child; (d) no incentives or financial inducements are given to subjects, except for compensation for expenses and loss of earnings directly related to the participation in the performance study.
(a) due to the urgency of the situation, caused by a sudden life-threatening or other sudden serious medical condition, the subject is unable to provide prior informed consent and to receive prior information on the performance study; (b) there are scientific grounds to expect that participation of the subject in the performance study will have the potential to produce a direct clinically relevant benefit for the subject resulting in a measurable health-related improvement alleviating the suffering and/or improving the health of the subject, or in the diagnosis of its condition; (c) it is not possible within the therapeutic window to supply all prior information to and obtain prior informed consent from his or her legally designated representative; (d) the investigator certifies that he or she is not aware of any objections to participate in the performance study previously expressed by the subject; (e) the performance study relates directly to the subject's medical condition because of which it is not possible within the therapeutic window to obtain prior informed consent from the subject or from his or her legally designated representative and to supply prior information, and the performance study is of such a nature that it may be conducted exclusively in emergency situations; (f) the performance study poses a minimal risk to, and imposes a minimal burden on, the subject in comparison with the standard treatment of the subject's condition.
(a) regarding incapacitated subjects and minors, the informed consent shall be sought by the investigator from his or her legally designated representative without undue delay and the information referred to in Article 59(2) shall be given as soon as possible to the subject and to his or her legally designated representative; (b) regarding other subjects, the informed consent shall be sought by the investigator without undue delay from the subject or his or her legally designated representative, whichever can be done sooner, and the information referred to in Article 59(2) shall be given as soon as possible to the subject or his or her legally designated representative, as applicable.
(a) in the case of performance studies carried out pursuant to point (a) of Article 58(1) and where the specimen collection does not represent a major clinical risk to the subject of the study, unless otherwise stated by national law, immediately after the validation date of application described in paragraph 5 of this Article, provided that a negative opinion which is valid for the entire Member State, under national law, has not been issued by an ethics committee in the Member State concerned in respect of the performance study; (b) in the case of performance studies carried out pursuant to points (b) and (c) of Article 58(1) and Article 58(2) or performance studies other than those referred to in point (a) of this paragraph, as soon as the Member State concerned has notified the sponsor of its authorisation and provided that a negative opinion which is valid for the entire Member State, under national law, has not been issued by an ethics committee in the Member State concerned in respect of the performance study. The Member State shall notify the sponsor of the authorisation within 45 days of the validation date of the application referred to in paragraph 5. The Member State may extend this period by a further 20 days for the purpose of consulting with experts.
(a) the demonstration of compliance of the device(s) for performance study with the applicable general safety and performance requirements, apart from the aspects covered by the performance study, and whether, with regard to those aspects, every precaution has been taken to protect the health and safety of the subjects. This includes, in case of performance studies, the evaluation of the analytical performance, and in case of interventional clinical performance studies, the evaluation of the analytical performance, clinical performance and scientific validity, taking into consideration the state of the art; (b) whether the risk-minimisation solutions employed by the sponsor are described in harmonised standards and, in those cases where the sponsor does not use harmonised standards, whether the risk-minimisation solutions provide a level of protection that is equivalent to that provided by harmonised standards; (c) whether the measures planned for the safe installation, putting into service and maintenance of the device for performance study are adequate; (d) the reliability and robustness of the data generated in the performance study, taking account of statistical approaches, design of the performance study and methodological aspects, including sample size, comparator and endpoints; (e) whether the requirements of Annex XIV are met.
(a) the application dossier submitted pursuant to Article 66(3) remains incomplete; (b) the device or the submitted documents, especially the performance study plan and the investigator's brochure, do not correspond to the state of scientific knowledge, and the performance study, in particular, is not suitable for providing evidence for the safety, performance characteristics or benefit of the device on subjects or patients; (c) the requirements of Article 58 are not met; or (d) any assessment under paragraph 3 is negative.
(a) the objective and methodology of the performance study; and (b) the degree of deviation of the intervention from normal clinical practice.
(a) to create the single identification numbers for performance studies referred to in Article 66(1); (b) to be used as an entry point for the submission of all applications or notifications for performance studies referred to in Articles 66, 70, 71 and 74 and for all other submission of data, or processing of data in this context; (c) for the exchange of information relating to performance studies in accordance with this Regulation between the Member States and between them and the Commission including the exchange of information referred to in to Articles 72 and 74; (d) for information to be provided by the sponsor in accordance with Article 73, including the performance study report and its summary as required in paragraph 5 of that Article; (e) for reporting on serious adverse events and device deficiencies, and related updates referred to in Article 76.
(a) protection of personal data in accordance with Regulation (EC) No 45/2001; (b) protection of commercially confidential information, especially in the investigators brochure, in particular through taking into account the status of the conformity assessment for the device, unless there is an overriding public interest in disclosure; (c) effective supervision of the conduct of the performance study by the Member State(s) concerned.
(a) the Member State in which the performance study is being or is to be conducted has notified the sponsor of its refusal based on the grounds referred to in Article 67(4) or on considerations of public health, of subject and user safety or health, or of public policy; or (b) an ethics committee in that Member State has issued a negative opinion in relation to the substantial modification to the performance study, which, in accordance with national law, is valid for that entire Member State.
(a) revoke the authorisation for the performance study; (b) suspend or terminate the performance study; (c) require the sponsor to modify any aspect of the performance study.
(a) within six days of receipt of the single application, notify the sponsor that it is the coordinating Member State ("notification date"); (b) for the purpose of the validation of the application, take into account any considerations submitted within seven days of the notification date by any Member State concerned; (c) within 10 days of the notification date, assess whether the performance study falls within the scope of this Regulation and whether the application is complete and shall notify the sponsor accordingly. Article 66(1) and (3) to (5) shall apply to the coordinating Member State in relation to that assessment; (d) establish the results of its assessment in a draft assessment report to be transmitted within 26 days of the validation date to the Member States concerned. By day 38 after the validation date, the other Member States concerned shall transmit their comments and proposals on the draft assessment report and the underlying application to the coordinating Member State which shall take due account of those comments and proposals in its finalisation of the final assessment report, to be transmitted within 45 days of the validation date to the sponsor and the other Member States concerned.
(a) when it considers that participation in the performance study would lead to a subject receiving treatment inferior to that received in normal clinical practice in that Member State concerned; (b) infringement of national law; or (c) considerations as regards subject safety and data reliability and robustness submitted under point (d) of paragraph 4.
(a) any adverse event of a type identified in the performance study plan as being critical to the evaluation of the results of that performance study; (b) any serious adverse event; (c) any device deficiency that might have led to a serious adverse event if appropriate action had not been taken, intervention had not occurred, or circumstances had been less fortunate; (d) any new findings in relation to any event referred to in points (a) to (c).
(a) any serious adverse event that has a causal relationship with the device, the comparator or the study procedure or where such causal relationship is reasonably possible; (b) any device deficiency that might have led to a serious adverse event if appropriate action had not been taken, intervention had not occurred, or circumstances had been less fortunate; (c) any new findings in relation to any event referred to in points (a) and (b).
(a) harmonised electronic forms for the application for performance studies and their assessment as referred to in Articles 66 and 74, taking into account specific categories or groups of devices; (b) the functioning of the electronic system referred to in Article 69; (c) harmonised electronic forms for the notification of PMPF studies as referred to in Article 70(1), and of substantial modifications as referred to in Article 71; (d) the exchange of information between Member States as referred to in Article 72; (e) harmonised electronic forms for the reporting of serious adverse events and device deficiencies as referred to in Article 76; (f) the timelines for the reporting of serious adverse events and device deficiencies, taking into account the severity of the event to be reported as referred to in Article 76; (g) uniform application of the requirements regarding the clinical evidence/data needed to demonstrate compliance with the general safety and performance requirements set out in Annex I.
(a) to update the benefit-risk determination and to improve the risk management as referred to in Chapter I of Annex I; (b) to update the design and manufacturing information, the instructions for use and the labelling; (c) to update the performance evaluation; (d) to update the summary of safety and performance referred to in Article 29; (e) for the identification of needs for preventive, corrective or field safety corrective action; (f) for the identification of options to improve the usability, performance and safety of the device; (g) when relevant, to contribute to the post-market surveillance of other devices; and (h) to detect and report trends in accordance with Article 83.
(a) the conclusions of the benefit-risk determination; (b) the main findings of the PMPF; and (c) the volume of sales of the device and an estimate of the size and other characteristics of the population using the device and, where practicable, the usage frequency of the device.
(a) any serious incident involving devices made available on the Union market, except expected erroneous results which are clearly documented and quantified in the product information and in the technical documentation and are subject to trend reporting pursuant to Article 83; (b) any field safety corrective action in respect of devices made available on the Union market, including any field safety corrective action undertaken in a third country in relation to a device which is also legally made available on the Union market, if the reason for the field safety corrective action is not limited to the device made available in the third country.
(a) where there is concern regarding a particular serious incident or cluster of serious incidents relating to the same device or type of device of the same manufacturer in more than one Member State; (b) where the appropriateness of a field safety corrective action that is proposed by a manufacturer in more than one Member State is in question.
designation of a coordinating competent authority on a case by case basis, when required; defining the coordinated assessment process, including the tasks and responsibilities of the coordinating competent authority and the involvement of other competent authorities.
(a) the typology of serious incidents and field safety corrective actions in relation to specific devices, or categories or groups of devices; (b) the reporting of serious incidents and field safety corrective actions and field safety notices, and the provision of periodic summary reports, post-market surveillance reports, PSURs and trend reports by manufacturers as referred to in Articles 80, 81, 82, 83 and 84 respectively; (c) standard structured forms for electronic and non-electronic reporting, including a minimum data set for reporting of suspected serious incidents by healthcare professionals, users and patients; (d) timelines for the reporting of field safety corrective actions, and for the provision by manufacturers of periodic summary reports and trend reports, taking into account the severity of the incident to be reported as referred to in Article 82; (e) harmonised forms for the exchange of information between competent authorities as referred to in Article 84; (f) procedures for the designation of a coordinating competent authority; the coordinated evaluation process, including tasks and responsibilities of the coordinating competent authority and involvement of other competent authorities in this process.
(a) reports by manufacturers on serious incidents and field safety corrective actions referred to in Article 82(1) and Article 84(5); (b) the periodic summary reports by manufacturers referred to in Article 82(9); (c) the reports by manufacturers on trends referred to in Article 83; (d) the PSURs referred to in Article 81; (e) the field safety notices by manufacturers referred to in Article 84(8); (f) the information to be exchanged between the competent authorities of the Member States and between them and the Commission in accordance with Article 84(7) and (9).
(a) the Member State in which the field safety corrective action is being or is to be undertaken; (b) the Member State in which the manufacturer has its registered place of business.
(a) the Member State or Member States participating in the coordination procedure in accordance with Article 84(9) and which have agreed on the periodic summary report; (b) the Member State in which the manufacturer has its registered place of business.
(a) may require economic operators to, inter alia , make available the documentation and information necessary for the purpose of carrying out the authorities' activities and, where justified, to provide the necessary samples of devices or access to devices free of charge; and(b) shall carry out both announced and, if necessary, unannounced inspections of the premises of economic operators, as well as suppliers and/or subcontractors, and, where necessary, at the facilities of professional users.
(a) may present an unacceptable risk to the health or safety of patients, users or other persons, or to other aspects of the protection of public health; or (b) otherwise does not comply with the requirements laid down in this Regulation,
(a) summaries of the results of the surveillance activities referred to in Article 88(4); (b) the final inspection report as referred to in Article 88(7); (c) information in relation to devices presenting an unacceptable risk to health and safety as referred to in Article 90(2), (4) and (6); (d) information in relation to non-compliance of products as referred to in Article 92(2); (e) information in relation to the preventive health protection measures referred to in Article 93(2); (f) summaries of the results of the reviews and assessments of the market surveillance activities of the Member States referred to in Article 88(8).
(a) to contribute to the assessment of applicant conformity assessment bodies and notified bodies pursuant to the provisions set out in Chapter IV; (b) to advise the Commission, at its request, in matters concerning the coordination group of notified bodies as established pursuant to Article 45; (c) to contribute to the development of guidance aimed at ensuring effective and harmonised implementation of this Regulation, in particular regarding the designation and monitoring of notified bodies, application of the general safety and performance requirements and conduct of performance evaluations by manufacturers, assessment by notified bodies and vigilance activities; (d) to contribute to the continuous monitoring of technical progress and assessment of whether the general safety and performance requirements laid down in this Regulation and Regulation (EU) 2017/745 are adequate to ensure safety and performance of devices, and thereby contribute to identifying whether there is a need to amend Annex I to this Regulation; (e) to contribute to the development of device standards and of CS; (f) to assist the competent authorities of the Member States in their coordination activities in particular in the fields of classification and the determination of the regulatory status of devices, performance studies, vigilance and market surveillance including the development and maintenance of a framework for a European market surveillance programme with the objective of achieving efficiency and harmonisation of market surveillance in the Union, in accordance with Article 88; (g) to provide advice, either on its own initiative or at request of the Commission, in the assessment of any issue related to the implementation of this Regulation; (h) to contribute to harmonised administrative practice with regard to devices in the Member States.
(a) to verify the performance claimed by the manufacturer and the compliance of class D devices with the applicable CS, when available, or with other solutions chosen by the manufacturer to ensure a level of safety and performance that is at least equivalent, as provided for in the third subparagraph of Article 48(3); (b) to carry out appropriate tests on samples of manufactured class D devices or batches of class D devices, as provided for in the Section 4.12 of Annex IX and in Section 5.1 of Annex XI; (c) to provide scientific and technical assistance to the Commission, the MDCG, the Member States and notified bodies in relation to the implementation of this Regulation; (d) to provide scientific advice regarding the state of the art in relation to specific devices, or a category or group of devices; (e) to set up and manage a network of national reference laboratories after consulting with the national authorities and publish a list of the participating national reference laboratories and their respective tasks; (f) to contribute to the development of appropriate testing and analysis methods to be applied for conformity assessment procedures and market surveillance; (g) to collaborate with notified bodies in the development of best practices for the performance of conformity assessment procedures; (h) to provide recommendations on suitable reference materials and reference measurement procedures of higher metrological order; (i) to contribute to the development of CS and of international standards; (j) to provide scientific opinions in response to consultations by notified bodies in accordance with this Regulation and publish them by electronic means having considered national provisions on confidentiality.
(a) have adequate and appropriately qualified staff with adequate knowledge and experience in the field of the in vitro diagnostic medical devices for which they are designated;(b) possess the necessary equipment and reference material to carry out the tasks assigned to them; (c) have the necessary knowledge of international standards and best practices; (d) have an appropriate administrative organisation and structure; (e) ensure that their staff observe the confidentiality of information and data obtained in carrying out their tasks; (f) act in the public interest and in an independent manner; (g) ensure that their staff do not have financial or other interests in the in vitro diagnostic medical device industry which could affect their impartiality, declare any other direct and indirect interests they may have in thein vitro diagnostic medical device industry and update this declaration whenever a relevant change occurs.
(a) applying coordinated methods, procedures and processes; (b) agreeing on the use of same reference materials and common test samples and seroconversion panels; (c) establishing common assessment and interpretation criteria; (d) using common testing protocols and assessing the test results using standardised and coordinated evaluation methods; (e) using standardised and coordinated test reports; (f) developing, applying and maintaining a peer review system; (g) organizing regular quality assessment tests (including mutual checks on the quality and comparability of test results); (h) agreeing on joint guidelines, instructions, procedural instructions or standard operational procedures; (i) coordinating the introduction of testing methods for new technologies and according to new or amended CS; (j) reassessing the state of the art on the basis of comparative test results or by further studies, as requested by a Member State or by the Commission.
(a) detailed rules to facilitate the application of paragraph 2 of this Article and detailed rules to ensure compliance with the criteria referred to in paragraph 4 of this Article. (b) the structure and the level of the fees referred to in paragraph 7 of this Article which may be levied by an EU reference laboratory for providing scientific opinions in response to consultations by notified bodies and Member States in accordance with this Regulation, taking into account the objectives of human health and safety protection, support of innovation and cost-effectiveness.
(a) personal data in accordance with Article 103; (b) commercially confidential information and trade secrets of a natural or legal person, including intellectual property rights unless disclosure is in the public interest; (c) the effective implementation of this Regulation, in particular for the purpose of inspections, investigations or audits.
(a) before the date of expiry of the certificate, the manufacturer and a notified body have signed a written agreement in accordance with Section 4.3, second subparagraph, of Annex VII to this Regulation for the conformity assessment in respect of the device covered by the expired certificate or in respect of a device intended to substitute that device; (b) a competent authority of a Member State has granted a derogation from the applicable conformity assessment procedure in accordance with Article 54(1) of this Regulation or has required the manufacturer, in accordance with Article 92(1) of this Regulation, to carry out the applicable conformity assessment procedure.
(a) 31 December 2027 , for class D devices;(b) 31 December 2028 , for class C devices;(c) 31 December 2029 , for class B devices and for class A devices placed on the market in sterile condition.
(a) those devices continue to comply with Directive 98/79/EC; (b) there are no significant changes in the design and intended purpose; (c) the devices do not present an unacceptable risk to the health or safety of patients, users or other persons, or to other aspects of the protection of public health; (d) no later than 26 May 2025 , the manufacturer has put in place a quality management system in accordance with Article 10(8);(e) the manufacturer or the authorised representative has lodged a formal application with a notified body in accordance with Section 4.3, first subparagraph, of Annex VII for conformity assessment in respect of a device referred to in paragraph 3a or 3b of this Article or in respect of a device intended to substitute that device, no later than: (i) 26 May 2025 , for devices referred to in paragraph 3a and paragraph 3b, point (a), of this Article;(ii) 26 May 2026 , for devices referred to in paragraph 3b, point (b), of this Article;(iii) 26 May 2027 , for devices referred to in paragraph 3b, point (c), of this Article;
(f) the notified body and the manufacturer have signed a written agreement in accordance with Section 4.3, second subparagraph, of Annex VII no later than: (i) 26 September 2025 , for devices referred to in paragraph 3a and paragraph 3b, point (a), of this Article;(ii) 26 September 2026 , for devices referred to in paragraph 3b, point (b), of this Article;(iii) 26 September 2027 , for devices referred to in paragraph 3b, point (c), of this Article.
(a) Article 11, Article 12(1), point (c), and Article 12(2) and (3) of Directive 98/79/EC, and the obligations relating to vigilance and performance studies provided for in the corresponding Annexes to that Directive, which are repealed, as applicable, with effect from the date referred to in Article 113(3), point (f), of this Regulation in respect of the application of the obligations and requirements that relate to the electronic systems referred to in Article 30(2), points (e) and (f), respectively, of this Regulation; (b) Article 10, Article 12(1), points (a) and (b), and Article 15(5) of Directive 98/79/EC, and the obligations relating to registration of devices and economic operators, and certificate notifications provided for in the corresponding Annexes to that Directive, which are repealed, as applicable, with effect from the date referred to in Article 113(3), point (f), of this Regulation in respect of the application of the obligations and requirements that relate to the electronic systems referred to in Article 30(2), points (a) to (d), respectively, of this Regulation.
(a) Articles 26(3) and 51(5) shall apply from 18 months after the later of the dates referred to in point (f); (b) Articles 31 to 46 and Article 96 shall apply from 26 November 2017 . However, from that date until26 May 2022 the obligations on notified bodies pursuant to Articles 31 to 46 shall apply only to those bodies which submit an application for designation in accordance with Article 34;(c) Article 97 shall apply from 26 May 2018 ;(d) Article 100 shall apply from 25 November 2020 ;(e) for class D devices, Article 24(4) shall apply from 26 May 2023 . For class B and class C devices Article 24(4) shall apply from26 May 2025 . For class A devices Article 24(4) shall apply from26 May 2027 ;(f) without prejudice to the obligations of the Commission pursuant to Article 34 of Regulation (EU) 2017/745, the obligations and requirements that relate to any of the electronic systems referred to in Article 30(2) of this Regulation shall apply from the date corresponding to 6 months from the date of publication of the notice referred to in Article 34(3) of Regulation (EU) 2017/745, informing that the relevant electronic system is functional and meets the functional specifications drawn up pursuant to Article 34(1) of that Regulation. The provisions referred to in the preceding sentence are: Article 26, Article 28, Article 29, the second sentence of Article 36(2), Article 38(10), Article 39(2), the second subparagraph of Article 40(12), points (d) and (e) of Article 42(7), Article 49(2), Article 50(1), Article 51(5), Articles 66 to 73, Article 74(1) to (13), without prejudice to Article 74(14), Articles 75 to 77, Article 81(2), Articles 82 and 83, Article 84(5) and (7) and the third subparagraph of Article 84(8), Article 85, Article 88(4), (7) and (8), Article 90(2) and (4), the last sentence of Article 92(2), Article 94(4), Article 110(3d).
Until the date of application of the provisions referred to in the first subparagraph of this point, the corresponding provisions of Directive 98/79/EC regarding information on vigilance reporting, performance studies, registration of devices and economic operators, and certificate notifications shall continue to apply. (fa) no later than 6 months from the date set out in point (f), first subparagraph, of this paragraph, manufacturers shall ensure that the information to be entered in Eudamed in accordance with Article 26 is entered in the electronic system referred to in Article 30(2), points (a) and (b), including regarding the following devices, provided that those devices are also placed on the market from the date set out in point (f), first subparagraph, of this paragraph: (i) devices for which the manufacturer has undertaken a conformity assessment in accordance with Article 48; (ii) devices placed on the market pursuant to Article 110(3), (3a) or (3b), unless the device, for which the manufacturer has undertaken a conformity assessment in accordance with Article 48, is already registered in Eudamed;
(fb) no later than 12 months from the date set out in point (f), first subparagraph, of this paragraph, notified bodies shall ensure that the information to be entered in Eudamed in accordance with Article 51(5) is entered in the electronic system referred to in Article 30(2), point (d), including regarding devices referred to in point (fa)(i) of this paragraph; for those devices, only the latest relevant certificate and, where applicable, any subsequent decision taken by the notified body related to such certificate shall be entered; (fc) by way of derogation from point (f), first subparagraph, of this paragraph, the obligations to upload the summary of safety and performance in accordance with Article 29(1) and to notify competent authorities in accordance with Article 50(1), through the electronic system referred in Article 30(2), point (d), shall apply to devices referred to in point (fa) of this paragraph when the certificate is entered in Eudamed in accordance with point (fb) of this paragraph; (fd) without prejudice to point (f), first subparagraph, of this paragraph, when a manufacturer is to submit a PSUR in accordance with Article 81(2) of this Regulation, to report a serious incident or a field safety corrective action in accordance with Article 82 of this Regulation, or to submit a trend report in accordance with Article 83 of this Regulation through the electronic system referred to in Article 30(2), point (f), of this Regulation, it shall also register the device, which is the subject of the PSUR or the vigilance reporting, in the electronic system referred to in Article 30(2), points (a) and (b), of this Regulation, except if such device was placed on the market in accordance with Directive 98/79/EC; (g) the procedure set out in Article 74 shall apply from 26 May 2029 without prejudice to Article 74(14);(h) Article 110(10) shall apply from 26 May 2019 ;(i) Article 5(5), points (b) and (c) and (e) to (i), shall apply from 26 May 2024 ;(j) Article 5(5), point (d), shall apply from 31 December 2030 .
(a) establish and document a risk management plan for each device; (b) identify and analyse the known and foreseeable hazards associated with each device; (c) estimate and evaluate the risks associated with, and occurring during, the intended use and during reasonably foreseeable misuse; (d) eliminate or control the risks referred to in point (c) in accordance with the requirements of Section 4; (e) evaluate the impact of information from the production phase and, in particular, from the post-market surveillance system, on hazards and the frequency of occurrence thereof, on estimates of their associated risks, as well as on the overall risk, the benefit-risk ratio and risk acceptability; and (f) based on the evaluation of the impact of the information referred to in point (e), if necessary amend control measures in line with the requirements of Section 4.
(a) eliminate or reduce risks as far as possible through safe design and manufacture; (b) where appropriate, take adequate protection measures, including alarms if necessary, in relation to risks that cannot be eliminated; and (c) provide information for safety (warnings/precautions/contra-indications) and, where appropriate, training to users.
(a) reduce as far as possible the risks related to the ergonomic features of the device and the environment in which the device is intended to be used (design for patient safety), and (b) give consideration to the technical knowledge, experience, education, training and use environment, where applicable, and the medical and physical conditions of intended users (design for lay, professional, disabled or other users).
(a) the analytical performance, such as, analytical sensitivity, analytical specificity, trueness (bias), precision (repeatability and reproducibility), accuracy (resulting from trueness and precision), limits of detection and quantitation, measuring range, linearity, cut-off, including determination of appropriate criteria for specimen collection and handling and control of known relevant endogenous and exogenous interference, cross-reactions; and (b) the clinical performance, such as diagnostic sensitivity, diagnostic specificity, positive predictive value, negative predictive value, likelihood ratio, expected values in normal and affected populations.
(a) for devices for self-testing, performances obtained by laypersons; (b) for devices for near-patient testing, performances obtained in relevant environments (for example, patient home, emergency units, ambulances).
(a) allow easy and safe handling; (b) reduce as far as possible any microbial leakage from the device and/or microbial exposure during use; and, where necessary (c) prevent microbial contamination of the device during use and, in the case of specimen receptacles, the risk of contamination of the specimen.
(a) the risk of injury, in connection with their physical features, including the volume/pressure ratio, dimensional and where appropriate ergonomic features; (b) risks connected with reasonably foreseeable external influences or environmental conditions, such as magnetic fields, external electrical and electromagnetic effects, electrostatic discharge, radiation associated with diagnostic or therapeutic procedures, pressure, humidity, temperature, variations in pressure and acceleration or radio signal interferences; (c) the risks associated with the use of the device when it comes into contact with materials, liquids, and substances, including gases, to which it is exposed during normal conditions of use; (d) the risks associated with the possible negative interaction between software and the IT environment within which it operates and interacts; (e) the risks of accidental ingress of substances into the device; (f) the risk of incorrect identification of specimens and the risk of erroneous results due to, for example, confusing colour and/or numeric and/or character codings on specimen receptacles, removable parts and/or accessories used with devices in order to perform the test or assay as intended; (g) the risks of any foreseeable interference with other devices.
(a) designed and manufactured in such a way as to ensure that the characteristics and the quantity of radiation emitted can be controlled and/or adjusted; and (b) fitted with visual displays and/or audible warnings of such emissions.
(a) ensure that the device can be used safely and accurately by the intended user at all stages of the procedure if necessary after appropriate training and/or information; and (b) reduce as far as possible the risk of error by the intended user in the handling of the device and, if applicable, the specimen, and also in the interpretation of the results.
(a) can verify that, at the time of use, the device will perform as intended by the manufacturer; and (b) be warned if the device has failed to provide a valid result.
(a) The medium, format, content, legibility, and location of the label and instructions for use shall be appropriate to the particular device, its intended purpose and the technical knowledge, experience, education or training of the intended user(s). In particular, instructions for use shall be written in terms readily understood by the intended user and, where appropriate, supplemented with drawings and diagrams. (b) The information required on the label shall be provided on the device itself. If this is not practicable or appropriate, some or all of the information may appear on the packaging for each unit. If individual full labelling of each unit is not practicable, the information shall be set out on the packaging of multiple devices. (c) Labels shall be provided in a human-readable format and may be supplemented by machine-readable information, such as radio-frequency identification or bar codes. (d) Instructions for use shall be provided together with devices. However, in duly justified and exceptional cases instructions for use shall not be required or may be abbreviated if the device can be used safely and as intended by the manufacturer without any such instructions for use. (e) Where multiple devices, with the exception of devices intended for self-testing or near-patient testing, are supplied to a single user and/or location, a single copy of the instructions for use may be provided if so agreed by the purchaser who in any case may request further copies to be provided free of charge. (f) When the device is intended for professional use only, instructions for use may be provided to the user in non-paper format (e.g. electronic), except when the device is intended for near-patient testing. (g) Residual risks which are required to be communicated to the user and/or other person shall be included as limitations, contra-indications, precautions or warnings in the information supplied by the manufacturer. (h) Where appropriate, the information supplied by the manufacturer shall take the form of internationally recognised symbols, taking into account the intended users. Any symbol or identification colour used shall conform to the harmonised standards or CS. In areas for which no harmonised standards or CS exist, the symbols and colours shall be described in the documentation supplied with the device. (i) In the case of devices containing a substance or a mixture which may be considered as being dangerous, taking account of the nature and quantity of its constituents and the form under which they are present, relevant hazard pictograms and labelling requirements of Regulation (EC) No 1272/2008 shall apply. Where there is insufficient space to put all the information on the device itself or on its label, the relevant hazard pictograms shall be put on the label and the other information required by Regulation (EC) No 1272/2008 shall be given in the instructions for use. (j) The provisions of Regulation (EC) No 1907/2006 on the safety data sheet shall apply, unless all relevant information, as appropriate, is already made available in the instructions for use.
(a) the name or trade name of the device; (b) the details strictly necessary for a user to identify the device and, where it is not obvious for the user, the intended purpose of the device; (c) the name, registered trade name or registered trade mark of the manufacturer and the address of its registered place of business; (d) if the manufacturer has its registered place of business outside the Union, the name of its authorised representative and the address of the registered place of business of the authorised representative; (e) an indication that the device is an in vitro diagnostic medical device, or if the device is a "device for performance study", an indication of that fact;(f) the lot number or the serial number of the device preceded by the words LOT NUMBER or SERIAL NUMBER or an equivalent symbol, as appropriate; (g) the UDI carrier as referred to in Article 24 and Part C of Annex VI; (h) an unambiguous indication of the time limit for using the device safely, without degradation of performance, expressed at least in terms of year and month and, where relevant, the day, in that order; (i) where there is no indication of the date until when it may be used safely, the date of manufacture. This date of manufacture may be included as part of the lot number or serial number, provided the date is clearly identifiable; (j) where relevant, an indication of the net quantity of contents, expressed in terms of weight or volume, numerical count, or any combination of thereof, or other terms which accurately reflect the contents of the package; (k) an indication of any special storage and/or handling condition that applies; (l) where appropriate, an indication of the sterile state of the device and the sterilisation method, or a statement indicating any special microbial state or state of cleanliness; (m) warnings or precautions to be taken that need to be brought to the immediate attention of the user of the device or to any other person. This information may be kept to a minimum in which case more detailed information shall appear in the instructions for use, taking into account the intended users; (n) if the instructions for use are not provided in paper form in accordance with point (f) of Section 20.1, a reference to their accessibility (or availability), and where applicable the website address where they can be consulted; (o) where applicable, any particular operating instructions; (p) if the device is intended for single use, an indication of that fact. A manufacturer's indication of single use shall be consistent across the Union; (q) if the device is intended for self-testing or near-patient testing, an indication of that fact; (r) where rapid assays are not intended for self-testing or near-patient testing, the explicit exclusion hereof; (s) where device kits include individual reagents and articles that are made available as separate devices, each of those devices shall comply with the labelling requirements contained in this Section and with the requirements of this Regulation; (t) the devices and separate components shall be identified, where applicable in terms of batches, to allow all appropriate action to detect any potential risk posed by the devices and detachable components. As far as practicable and appropriate, the information shall be set out on the device itself and/or, where appropriate, on the sales packaging; (u) the label for devices for self-testing shall bear the following particulars: (i) the type of specimen(s) required to perform the test (e.g. blood, urine or saliva); (ii) the need for additional materials for the test to function properly; (iii) contact details for further advice and assistance.
The name of devices for self-testing shall not reflect an intended purpose other than that specified by the manufacturer.
(a) an indication permitting the sterile packaging to be recognised as such, (b) a declaration that the device is in a sterile condition, (c) the method of sterilisation, (d) the name and address of the manufacturer, (e) a description of the device, (f) the month and year of manufacture, (g) an unambiguous indication of the time limit for using the device safely, expressed at least in terms of year and month and, where relevant, the day, in that order, (h) an instruction to check the instructions for use for what to do if the sterile packaging is damaged or unintentionally opened before use.
(a) the name or trade name of the device; (b) the details strictly necessary for the user to uniquely identify the device; (c) the device's intended purpose: (i) what is detected and/or measured; (ii) its function (e.g. screening, monitoring, diagnosis or aid to diagnosis, prognosis, prediction, companion diagnostic); (iii) the specific information that is intended to be provided in the context of: a physiological or pathological state; congenital physical or mental impairments; the predisposition to a medical condition or a disease; the determination of the safety and compatibility with potential recipients; the prediction of treatment response or reactions; the definition or monitoring of therapeutic measures;
(iv) whether it is automated or not; (v) whether it is qualitative, semi-quantitative or quantitative; (vi) the type of specimen(s) required; (vii) where applicable, the testing population; and (viii) for companion diagnostics, the International Non-proprietary Name (INN) of the associated medicinal product for which it is a companion test.
(d) an indication that the device is an in vitro diagnostic medical device, or, if the device is a "device for performance study", an indication of that fact;(e) the intended user, as appropriate (e.g. self-testing, near patient and laboratory professional use, healthcare professionals); (f) the test principle; (g) a description of the calibrators and controls and any limitation upon their use (e.g. suitable for a dedicated instrument only); (h) a description of the reagents and any limitation upon their use (e.g. suitable for a dedicated instrument only) and the composition of the reagent product by nature and amount or concentration of the active ingredient(s) of the reagent(s) or kit as well as a statement, where appropriate, that the device contains other ingredients which might influence the measurement; (i) a list of materials provided and a list of special materials required but not provided; (j) for devices intended for use in combination with or installed with or connected to other devices and/or general purpose equipment: information to identify such devices or equipment, in order to obtain a validated and safe combination, including key performance characteristics, and/or information on any known restrictions to combinations of devices and equipment.
(k) an indication of any special storage (e.g. temperature, light, humidity, etc.) and/or handling conditions which apply; (l) in-use stability which may include the storage conditions, and shelf life following the first opening of the primary container, together with the storage conditions and stability of working solutions, where this is relevant; (m) if the device is supplied as sterile, an indication of its sterile state, the sterilisation method and instructions in the event of the sterile packaging being damaged before use; (n) information that allows the user to be informed of any warnings, precautions, measures to be taken and limitations of use regarding the device. That information shall cover, where appropriate: (i) warnings, precautions and/or measures to be taken in the event of malfunction of the device or its degradation as suggested by changes in its appearance that may affect performance, (ii) warnings, precautions and/or measures to be taken as regards the exposure to reasonably foreseeable external influences or environmental conditions, such as magnetic fields, external electrical and electromagnetic effects, electrostatic discharge, radiation associated with diagnostic or therapeutic procedures, pressure, humidity, or temperature, (iii) warnings, precautions and/or measures to be taken as regards the risks of interference posed by the reasonably foreseeable presence of the device during specific diagnostic investigations, evaluations, therapeutic treatment or other procedures such as electromagnetic interference emitted by the device affecting other equipment, (iv) precautions related to materials incorporated into the device that contain or consist of CMR substances, or endocrine disrupting substances or that could result in sensitisation or an allergic reaction by the patient or user, (v) if the device is intended for single use, an indication of that fact. A manufacturer's indication of single use shall be consistent across the Union, (vi) if the device is reusable, information on the appropriate processes to allow reuse, including cleaning, disinfection, decontamination, packaging and, where appropriate, the validated method of re-sterilisation. Information shall be provided to identify when the device should no longer be reused, such as signs of material degradation or the maximum number of allowable reuses;
(o) any warnings and/or precautions related to potentially infectious material that is included in the device; (p) where relevant, requirements for special facilities, such as a clean room environment, or special training, such as on radiation safety, or particular qualifications of the intended user; (q) conditions for collection, handling, and preparation of the specimen; (r) details of any preparatory treatment or handling of the device before it is ready for use, such as sterilisation, final assembly, calibration, etc., for the device to be used as intended by the manufacturer; (s) the information needed to verify whether the device is properly installed and is ready to perform safely and as intended by the manufacturer, together with, where relevant: details of the nature, and frequency, of preventive and regular maintenance, including cleaning and disinfection; identification of any consumable components and how to replace them; information on any necessary calibration to ensure that the device operates properly and safely during its intended lifetime; methods for mitigating the risks encountered by persons involved in installing, calibrating or servicing devices.
(t) where applicable, recommendations for quality control procedures; (u) the metrological traceability of values assigned to calibrators and control materials, including identification of applied reference materials and/or reference measurement procedures of higher order and information regarding maximum (self-allowed) batch to batch variation provided with relevant figures and units of measure; (v) assay procedure including calculations and interpretation of results and where relevant if any confirmatory testing shall be considered; where applicable, the instructions for use shall be accompanied by information regarding batch to batch variation provided with relevant figures and units of measure; (w) analytical performance characteristics, such as analytical sensitivity, analytical specificity, trueness (bias), precision (repeatability and reproducibility), accuracy (resulting from trueness and precision), limits of detection and measurement range, (information needed for the control of known relevant interferences, cross-reactions and limitations of the method), measuring range, linearity and information about the use of available reference measurement procedures and materials by the user; (x) clinical performance characteristics as defined in Section 9.1 of this Annex; (y) the mathematical approach upon which the calculation of the analytical result is made; (z) where relevant, clinical performance characteristics, such as threshold value, diagnostic sensitivity and diagnostic specificity, positive and negative predictive value; (aa) where relevant, reference intervals in normal and affected populations; (ab) information on interfering substances or limitations ( e.g. visual evidence of hyperlipidaemia or haemolysis, age of specimen) that may affect the performance of the device;(ac) warnings or precautions to be taken in order to facilitate the safe disposal of the device, its accessories, and the consumables used with it, if any. This information shall cover, where appropriate: (i) infection or microbial hazards, such as consumables contaminated with potentially infectious substances of human origin; (ii) environmental hazards such as batteries or materials that emit potentially hazardous levels of radiation); (iii) physical hazards such as explosion.
(ad) the name, registered trade name or registered trade mark of the manufacturer and the address of its registered place of business at which he can be contacted and its location be established, together with a telephone number and/or fax number and/or website address to obtain technical assistance; (ae) date of issue of the instructions for use or, if they have been revised, date of issue and identifier of the latest revision of the instructions for use, with a clear indication of the introduced modifications; (af) a notice to the user that any serious incident that has occurred in relation to the device shall be reported to the manufacturer and the competent authority of the Member State in which the user and/or the patient is established; (ag) where device kits include individual reagents and articles that may be made available as separate devices, each of these devices shall comply with the instructions for use requirements contained in this Section and with the requirements of this Regulation; (ah) for devices that incorporate electronic programmable systems, including software, or software that are devices in themselves, minimum requirements concerning hardware, IT networks characteristics and IT security measures, including protection against unauthorised access, necessary to run the software as intended.
(a) details of the test procedure shall be given, including any reagent preparation, specimen collection and/or preparation and information on how to run the test and interpret the results; (b) specific particulars may be omitted provided that the other information supplied by the manufacturer is sufficient to enable the user to use the device and to understand the result(s) produced by the device; (c) the device's intended purpose shall provide sufficient information to enable the user to understand the medical context and to allow the intended user to make a correct interpretation of the results; (d) the results shall be expressed and presented in a way that is readily understood by the intended user; (e) information shall be provided with advice to the user on action to be taken (in case of positive, negative or indeterminate result), on the test limitations and on the possibility of false positive or false negative result. Information shall also be provided as to any factors that can affect the test result such as age, gender, menstruation, infection, exercise, fasting, diet or medication; (f) the information provided shall include a statement clearly directing that the user should not take any decision of medical relevance without first consulting the appropriate healthcare professional, information on disease effects and prevalence, and, where available, information specific to the Member State(s) where the device is placed on the market on where a user can obtain further advice such as national helplines, websites; (g) for devices intended for self-testing used for the monitoring of a previously diagnosed existing disease or condition, the information shall specify that the patient should only adapt the treatment if he has received the appropriate training to do so.
(a) product or trade name and a general description of the device including its intended purpose and intended users; (b) the Basic UDI-DI as referred to in Part C of Annex VI assigned by the manufacturer to the device in question, as soon as identification of this device becomes based on a UDI system, or otherwise a clear identification by means of product code, catalogue number or other unambiguous reference allowing traceability; (c) the intended purpose of the device which may include information on: (i) what is to be detected and/or measured; (ii) its function such as screening, monitoring, diagnosis or aid to diagnosis, prognosis, prediction, companion diagnostic; (iii) the specific disorder, condition or risk factor of interest that it is intended to detect, define or differentiate; (iv) whether it is automated or not; (v) whether it is qualitative, semi-quantitative or quantitative; (vi) the type of specimen(s) required; (vii) where applicable, the testing population; (viii) the intended user; (ix) in addition, for companion diagnostics, the relevant target population and the associated medicinal product(s).
(d) the description of the principle of the assay method or the principles of operation of the instrument; (e) the rationale for the qualification of the product as a device; (f) the risk class of the device and the justification for the classification rule(s) applied in accordance with Annex VIII; (g) the description of the components and where appropriate, the description of the reactive ingredients of relevant components such as antibodies, antigens, nucleic acid primers; and where applicable: (h) the description of the specimen collection and transport materials provided with the device or descriptions of specifications recommended for use; (i) for instruments of automated assays: the description of the appropriate assay characteristics or dedicated assays; (j) for automated assays: a description of the appropriate instrumentation characteristics or dedicated instrumentation; (k) a description of any software to be used with the device; (l) a description or complete list of the various configurations/variants of the device that are intended to be made available on the market; (m) a description of the accessories for a device, other devices and other products that are not devices, which are intended to be used in combination with the device.
(a) an overview of the previous generation or generations of the device produced by the manufacturer, where such devices exist; (b) an overview of identified similar devices available on the Union or international markets, where such devices exist.
(a) the label or labels on the device and on its packaging, such as single unit packaging, sales packaging, transport packaging in the case of specific management conditions, in the languages accepted in the Member States where the device is envisaged to be sold; (b) the instructions for use in the languages accepted in the Member States where the device is envisaged to be sold.
(a) a description of the critical ingredients of the device such as antibodies, antigens, enzymes and nucleic acid primers provided or recommended for use with the device; (b) for instruments, a description of major subsystems, analytical technology such as operating principles and control mechanisms, dedicated computer hardware and software; (c) for instruments and software, an overview of the entire system; (d) for software, a description of the data interpretation methodology, namely the algorithm; (e) for devices intended for self-testing or near-patient testing, a description of the design aspects that make them suitable for self-testing or near-patient testing.
(a) information to allow the manufacturing processes such as production, assembly, final product testing, and packaging of the finished device to be understood. More detailed information shall be provided for the audit of the quality management system or other applicable conformity assessment procedures; (b) identification of all sites, including suppliers and sub-contractors, where manufacturing activities are performed.
(a) the general safety and performance requirements that apply to the device and an explanation as to why others do not apply; (b) the method or methods used to demonstrate conformity with each applicable general safety and performance requirement; (c) the harmonised standards, CS or other solutions applied; (d) the precise identity of the controlled documents offering evidence of conformity with each harmonised standard, CS or other method applied to demonstrate conformity with the general safety and performance requirements. The information referred to under this point shall incorporate a cross-reference to the location of such evidence within the full technical documentation and, if applicable, the summary technical documentation.
(a) the benefit-risk analysis referred to in Sections 1 and 8 of Annex I, and (b) the solutions adopted and the results of the risk management referred to in Section 3 of Annex I.
(a) Trueness of measurement This Section shall provide information on the trueness of the measurement procedure and summarise the data in sufficient detail to allow an assessment of the adequacy of the means selected to establish the trueness. Trueness measures apply to both quantitative and qualitative assays only when a certified reference material or certified reference method is available. (b) Precision of measurement This Section shall describe repeatability and reproducibility studies.
(a) substances used for patient treatment such as medicinal products; (b) substances ingested by the patient such as alcohol, foods; (c) substances added during specimen preparation such as preservatives, stabilisers; (d) substances encountered in specific specimen types such as haemoglobin, lipids, bilirubin, proteins; (e) analytes of similar structure such as precursors, metabolites or medical conditions unrelated to the test condition including specimens negative for the assay but positive for a condition that can mimic the test condition.
(a) the population(s) studied: demographics, selection, inclusion and exclusion criteria, number of individuals included; (b) method or mode of characterisation of specimens; and (c) statistical methods such as Receiver Operator Characteristic (ROC) to generate results and if applicable, define grey-zone/equivocal zone.
(a) the study report including the protocol, number of lots, acceptance criteria and testing intervals; (b) where accelerated studies have been performed in anticipation of the real time studies, the method used for accelerated studies shall be described; (c) the conclusions and claimed shelf life.
(a) the study report (including the protocol, acceptance criteria and testing intervals); (b) the conclusions and claimed in-use stability.
(a) the study report (including the protocol, acceptance criteria); (b) the method used for simulated conditions; (c) the conclusion and recommended shipping conditions.
(a) In the case of devices placed on the market in a sterile or defined microbiological condition, a description of the environmental conditions for the relevant manufacturing steps. In the case of devices placed on the market in a sterile condition, a description of the methods used, including the validation reports, with regard to packaging, sterilisation and maintenance of sterility. The validation report shall address bioburden testing, pyrogen testing and, if applicable, testing for sterilant residues. (b) In the case of devices containing tissues, cells and substances of animal, human or microbial origin, information on the origin of such material and on the conditions in which it was collected. (c) In the case of devices placed on the market with a measuring function, a description of the methods used in order to ensure the accuracy as given in the specifications. (d) If the device is to be connected to other equipment in order to operate as intended, a description of the resulting combination including proof that it conforms to the general safety and performance requirements set out in Annex I when connected to any such equipment having regard to the characteristics specified by the manufacturer.
(a) The post-market surveillance plan shall address the collection and utilisation of available information, in particular: information concerning serious incidents, including information from PSURs, and field safety corrective actions, records referring to non-serious incidents and data on any undesirable side-effects, information from trend reporting, relevant specialist or technical literature, databases and/or registers, information, including feedbacks and complaints, provided by users, distributors and importers, and publicly-available information about similar medical devices.
(b) The post-market surveillance plan shall cover at least: a proactive and systematic process to collect any information referred to in point (a). The process shall allow a correct characterisation of the performance of the devices and shall also allow a comparison to be made between the device and similar products available on the market; effective and appropriate methods and processes to assess the collected data; suitable indicators and threshold values that shall be used in the continuous reassessment of the benefit-risk analysis and of the risk management as referred to in Section 3 of Annex I; effective and appropriate methods and tools to investigate complaints and analyse market-related experience collected in the field; methods and protocols to manage the incidents subject to the trend report as provided for in Article 83, including the methods and protocols to be used to establish any statistically significant increase in the frequency or severity of incidents as well as the observation period; methods and protocols to communicate effectively with competent authorities, notified bodies, economic operators and users; reference to procedures to fulfil the manufacturers obligations laid down in Articles 78, 79 and 81; systematic procedures to identify and initiate appropriate measures including corrective actions; effective tools to trace and identify devices for which corrective actions might be necessary; and a PMPF plan as referred to in Part B of Annex XIII, or a justification as to why a PMPF is not applicable.
1. Name, registered trade name or registered trade mark and, if already issued, SRN referred to in Article 28 of the manufacturer, and, if applicable, its authorised representative, and the address of their registered place of business where they can be contacted and their location be established; 2. A statement that the EU declaration of conformity is issued under the sole responsibility of the manufacturer; 3. The Basic UDI-DI as referred to in Part C of Annex VI; 4. Product and trade name, product code, catalogue number or other unambiguous reference allowing identification and traceability of the device covered by the EU declaration of conformity, such as a photograph, where appropriate, as well as its intended purpose. Except for the product or trade name, the information allowing identification and traceability may be provided by the Basic UDI-DI referred to in point 3; 5. Risk class of the device in accordance with the rules set out in Annex VIII; 6. A statement that the device that is covered by the present declaration is in conformity with this Regulation and, if applicable, with any other relevant Union legislation that provides for the issuing of an EU declaration of conformity; 7. References to any CS used and in relation to which conformity is declared; 8. Where applicable, the name and identification number of the notified body, a description of the conformity assessment procedure performed and identification of the certificate or certificates issued; 9. Where applicable, additional information; 10. Place and date of issue of the declaration, name and function of the person who signed it as well as an indication for, and on behalf of whom, that person signed, signature.
1.1. type of economic operator (manufacturer, authorised representative, or importer), 1.2. name, address and contact details of the economic operator, 1.3. where submission of information is carried out by another person on behalf of any of the economic operators mentioned under Section 1.1, the name, address and contact details of that person, 1.4. name address and contact details of the person or persons responsible for regulatory compliance referred to in Article 15,
2.1. Basic UDI-DI, 2.2. type, number and expiry date of the certificate issued by the notified body and the name or identification number of that notified body and the link to the information that appears on the certificate and was entered by the notified body in the electronic system on notified bodies and certificates, 2.3. Member State in which the device shall or has been placed on the market in the Union, 2.4. in the case of class B, class C or class D devices: Member States where the device is or is to be made available, 2.5. presence of tissues, cells, or, their derivatives, of human origin (y/n), 2.6. presence of tissues, cells or their derivatives of animal origin as referred to in Regulation (EU) No 722/2012(y/n), 2.7. presence of cells or substances of microbial origin (y/n), 2.8. risk class of the device, 2.9. where applicable, the single identification number of the performance study, 2.10. in the case of devices designed and manufactured by another legal or natural person as referred in Article 10(14), the name, address and contact details of that legal or natural person, 2.11. in the case of class C or D devices, the summary of safety and performance, 2.12. status of the device (on the market, no longer placed on the market, recalled, field safety corrective Action initiated), 2.13. indication as to whether the device is a "new" device. A device shall be considered to be "new" if: (a) there has been no such device continuously available on the Union market during the previous three years for the relevant analyte or other parameter; (b) the procedure involves analytical technology not continuously used in connection with a given analyte or other parameter on the Union market during the previous three years.
2.14. indication as to whether the device is intended for self-testing or near-patient testing.
1. quantity per package configuration, 2. the Basic UDI-DI as referred to in Article 24(6) and any additional UDI-DIs, 3. the manner in which production of the device is controlled (expiry date or manufacturing date, lot number, serial number), 4. if applicable, the "unit of use" UDI-DI (where a UDI is not labelled on the device at the level of its "unit of use", a "unit of use" UDI-DI shall be assigned so as to associate the use of a device with a patient), 5. name and address of the manufacturer, as indicated on the label, 6. the SRN issued in accordance with Article 28(2), 7. if applicable, name and address of the authorised representative (as indicated on the label), 8. the medical device nomenclature code as provided for in Article 23, 9. risk class of the device, 10. if applicable, name or trade name, 11. if applicable, device model, reference, or catalogue number, 12. additional product description (optional), 13. if applicable, storage and/or handling conditions (as indicated on the label or in the instructions for use), 14. if applicable, additional trade names of the device, 15. labelled as a single use device (y/n), 16. if applicable, the maximum number of reuses, 17. device labelled sterile (y/n), 18. need for sterilisation before use (y/n), 19. URL for additional information, such as electronic instructions for use (optional), 20. if applicable, critical warnings or contra-indications, 21. status of the device (on the market, no longer placed on the market, recalled, field safety action initiated).
(a) Name or trade name, (b) device version or model, (c) labelled as single use, (d) packaged sterile, (e) need for sterilization before use, (f) quantity of devices provided in a package, (g) critical warnings or contra-indications.
(a) the original performance, (b) the safety or the intended use of the software. (c) interpretation of data.
Minor software revisions are generally associated with bug fixes, usability enhancements that are not for safety purposes, security patches or operating efficiency. Minor software revisions shall be identified by a manufacturer-specific form of identification.
(a) where the software is delivered on a physical medium, for example via a CD or DVD, each packaging level shall bear the human readable and AIDC representation of the complete UDI. The UDI that is applied to the physical medium containing the software and its packaging shall be identical to the UDI assigned to the system level software;(b) the UDI shall be provided on a readily accessible screen for the user in an easily-readable plain-text format such as an "about" file, or included on the start-up screen; (c) software lacking a user interface such as middleware for image conversion, shall be capable of transmitting the UDI through an application programming interface (API); (d) only the human readable portion of the UDI shall be required in electronic displays of the software. The marking of UDI using AIDC shall not be required in the electronic displays such as "about" menu, splash screen, etc.; (e) the human readable format of the UDI for the software shall include the application identifiers (AI) for the standard used by the issuing entities, so as to assist the user in identifying the UDI and determining which standard is being used to create the UDI.
(a) provision of adequate resources for conformity assessment activities; (b) development of procedures and policies for the operation of the notified body; (c) supervision of implementation of the procedures, policies and quality management systems of the notified body; (d) supervision of the notified body's finances; (e) activities and decisions taken by the notified body, including contractual agreements; (f) delegation of authority to personnel and/or committees, where necessary, for the performance of defined activities; (g) interaction with the authority responsible for notified bodies and the obligations regarding communications with other competent authorities, the Commission and other notified bodies.
(a) be the designer, manufacturer, supplier, installer, purchaser, owner or maintainer of devices which they assess, nor the authorised representative of any of those parties. Such restriction shall not preclude the purchase and use of assessed devices that are necessary for the operations of the notified body and the conduct of the conformity assessment, or the use of such devices for personal purposes; (b) be involved in the design, manufacture or construction, marketing, installation and use, or maintenance of the devices for which they are designated, nor represent the parties engaged in those activities; (c) engage in any activity that may conflict with their independence of judgement or integrity in relation to conformity assessment activities for which they are designated; (d) offer or provide any service which may jeopardise the confidence in their independence, impartiality or objectivity. In particular, they shall not offer or provide consultancy services to the manufacturer, its authorised representative, a supplier or a commercial competitor as regards the design, construction, marketing or maintenance of the devices or processes under assessment; and (e) be linked to any organisation which itself provides consultancy services as referred to in the point (d). Such restriction shall not preclude general training activities that are not client specific and that relate to regulation of devices s or to related standards.
(a) management system structure and documentation, including policies and objectives for its activities; (b) policies for assignment of activities and responsibilities to personnel; (c) assessment and decision-making processes in accordance with the tasks, responsibilities and role of the notified body's personnel and top-level management; (d) the planning, conduct, evaluation and, if necessary, adaptation of its conformity assessment procedures; (e) control of documents; (f) control of records; (g) management reviews; (h) internal audits; (i) corrective and preventive actions; (j) complaints and appeals; (k) continuous training.
biological safety, performance evaluation, devices for self and near patient testing, companion diagnostics, functional safety, software, packaging, and the different types of sterilisation processes.
Union devices legislation and relevant guidance documents; the conformity assessment procedures provided for in this Regulation; a broad base of knowledge of device technologies and the design and manufacture of devices; the notified body's quality management system, related procedures and the required qualification criteria; training relevant to personnel involved in conformity assessment activities in relation to devices; adequate experience in conformity assessments under this Regulation or previously applicable law within a notified body.
identify when specialist input is required for the assessment of the performance evaluation conducted by the manufacturer and identify appropriately qualified experts; appropriately train external clinical experts in the relevant requirements of this Regulation, CS, guidance and harmonised standards and ensure that the external clinical experts are fully aware of the context and implications of their assessment and the advice they provide; be able to review and scientifically challenge the clinical data contained within the performance evaluation, and any associated performance study, and appropriately guide external clinical experts in the assessment of the performance evaluation presented by the manufacturer; be able to scientifically evaluate and, if necessary, challenge the performance evaluation presented, and the results of the external clinical experts' assessment of the manufacturer's performance evaluation; be able to ascertain the comparability and consistency of the assessments of performance evaluation conducted by clinical experts; be able to make an assessment of the manufacturer's performance evaluation and a clinical judgement of the opinion provided by any external expert and make a recommendation to the notified body's decision maker; and be able to draw up records and reports demonstrating that the relevant conformity assessment activities have been appropriately carried out.
successful completion of a university or a technical college degree or an equivalent qualification in relevant studies, such as medicine, pharmacy, engineering or other relevant sciences; four years' professional experience in the field of healthcare products or related activities, such as in manufacturing, auditing, or research, of which two years shall be in the design, manufacture, testing or use of devices or technology to be assessed or related to the scientific aspects to be assessed; knowledge of device legislation, including the general safety and performance requirements set out in Annex I; appropriate knowledge and experience of relevant harmonised standards, CS and guidance documents; appropriate knowledge and experience of risk management and related device standards and guidance documents; appropriate knowledge and experience of performance evaluation; appropriate knowledge of the devices which they are assessing; appropriate knowledge and experience of the conformity assessment procedures laid down in Annexes IX to XI, in particular of the aspects of those procedures for which they are responsible, and adequate authorisation for carrying out those assessments; the ability to draw up records and reports demonstrating that the relevant conformity assessment activities have been appropriately carried out.
successful completion of a university or a technical college degree or equivalent qualification in relevant studies, such as medicine, pharmacy, engineering or other relevant sciences; four years' professional experience in the field of healthcare products or related activities, such as in manufacturing, auditing or research, of which two years shall be in the area of quality management; appropriate knowledge of devices legislation as well as related harmonised standards, CS and guidance documents; appropriate knowledge and experience of risk management and related device standards and guidance documents; appropriate knowledge of quality management systems and related l devices standards and guidance documents; appropriate knowledge and experience of the conformity assessment procedures laid down in Annexes IX to XI, in particular of the aspects of those procedures for which they are responsible, and adequate authorisation for carrying out those audits; training in auditing techniques enabling them to challenge quality management systems; the ability to draw up records and reports demonstrating that the relevant conformity assessment activities have been appropriately carried out.
devices legislation and relevant guidance documents; device conformity assessments relevant to this Regulation; the types of qualifications, experience and expertise relevant to device conformity assessment; a broad base of knowledge of device technologies, including sufficient experience of the conformity assessment of devices being reviewed for certification, the device industry and the design and manufacture of devices; the notified body's quality system, related procedures and the required qualifications for personnel involved; the ability to draw up records and reports demonstrating that the conformity assessment activities have been appropriately carried out.
a matrix detailing the authorisations and responsibilities of the personnel in respect of conformity assessment activities; records attesting to the required knowledge and experience for the conformity assessment activity for which they are authorised. The records shall contain a rationale for defining the scope of the responsibilities for each of the assessment personnel and records of the conformity assessment activities carried out by each of them.
review of the qualifications and monitoring of the performance of external experts; auditing and certification activities where the subcontracting in question is to auditing or certification organisations; allocation of work to external experts for specific conformity assessment activities; final review and decision-making functions.
the subcontractor meets the relevant requirements of this Annex; subcontractors and external experts do not further subcontract work to organisations or personnel; the natural or legal person that applied for conformity assessment has been informed of the requirements referred to in the first and second indent.
are aware of the Union and national law in force on devices, relevant harmonised standards, CS, guidance documents and the results of the coordination activities referred to in Section 1.6; take part in the internal exchange of experience and the continuous training and education programme referred to in Section 3.1.2.
(a) publish a publicly available description of the application procedure by which manufacturers can obtain certification from it. That description shall include which languages are acceptable for submission of documentation and for any related correspondence; (b) have documented procedures relating to, and documented details about, fees charged for specific conformity assessment activities and any other financial conditions relating to notified bodies' assessment activities for devices;, (c) have documented procedures in relation to advertising of its conformity assessment services. Those procedures shall ensure that advertising or promotional activities in no way imply or are capable of leading to an inference that their conformity assessment will offer manufacturers earlier market access or be quicker, easier or less stringent than that of other notified bodies; (d) have documented procedures requiring the review of pre-application information including the preliminary verification that the product is covered by this Regulation and its classification prior to issuing any quotation to the manufacturer relating to a specific conformity assessment; (e) ensure that all contracts relating to the conformity assessment activities covered by this Regulation are concluded directly between the manufacturer and the notified body and not with any other organisation.
(a) the completeness of those applications with respect to the requirements of the relevant conformity assessment procedure, as referred to in the corresponding Annex, under which approval has been sought, (b) the verification of the qualification of products covered by those applications as devices and their respective classifications, (c) whether the conformity assessment procedures chosen by the applicant are applicable to the device in question under this Regulation, (d) the ability of the notified body to assess the application based on its designation, and (e) the availability of sufficient and appropriate resources.
to appropriately plan the conduct of each individual project, to ensure that the composition of the assessment teams is such that there is sufficient experience in relation to the technology concerned, and that there is continuous objectivity and independence, and to provide for rotation of the members of the assessment team at appropriate intervals, to specify the rationale for fixing time limits for completion of conformity assessment activities, to assess the manufacturer's technical documentation and the solutions adopted to meet the requirements laid down in Annex I, to review the manufacturer's procedures and documentation relating to performance evaluation, to address the interface between the manufacturer's risk management process and its appraisal and analysis of the performance evaluation and to evaluate their relevance for the demonstration of conformity with the relevant requirements in Annex I, to carry out the "specific procedures" referred to in Section 5 of Annex IX, in the case of class B or C devices, to assess the technical documentation of devices selected on a representative basis, to plan and periodically carry out appropriate surveillance audits and assessments, to carry out or request certain tests to verify the proper functioning of the quality management system and to perform unannounced on site audits, relating to the sampling of devices verify that the manufactured device is in conformity with the technical documentation; such requirements shall define the relevant sampling criteria and testing procedure prior to sampling, to evaluate and verify a manufacturer's compliance with relevant Annexes.
(a) As part of the assessment of the quality management system a, a notified body shall prior to an audit and in accordance with its documented procedures: assess the documentation submitted in accordance with the relevant conformity assessment Annex, and draw up an audit programme which clearly identifies the number and sequence of activities required to demonstrate complete coverage of a manufacturer's quality management system and to determine whether it meets the requirements of this Regulation, identify links between and allocation of responsibilities among, the various manufacturing sites, and identify relevant suppliers and/or subcontractors of the manufacturer, and consider the need to specifically audit any of those suppliers or subcontractors or both, clearly define, for each audit identified in the audit programme, the objectives, criteria and scope of the audit, and draw up an audit plan that adequately addresses and takes account of the specific requirements for the devices, technologies and processes involved, draw up and keep up to date, for class B and class C devices, a sampling plan for the assessment of technical documentation as referred to in Annexes II and III covering the range of such devices covered by the manufacturer's application. That plan shall ensure that the entire range of devices covered by the certificate is sampled over the period of validity of the certificate, and select and assign appropriately qualified and authorised personnel for conducting the individual audits. The respective roles, responsibilities and authorities of the team members shall be clearly defined and documented.
(b) Based on the audit programme it has drawn up, the notified body shall, in accordance with its documented procedures: audit the manufacturer's quality management system in order to verify that the quality management system ensures that the devices covered conform to the relevant provisions of this Regulation which apply to devices at every stage, from design through final quality control to ongoing surveillance, and shall determine whether the requirements of this Regulation are met, based on relevant technical documentation, and in order to determine whether the manufacturer meets the requirements referred to in the relevant conformity assessment Annex, review and audit the manufacturer's processes and subsystems,– in particular for: design and development, production and process controls, product documentation, purchasing controls including verification of purchased devices, corrective and preventive actions including for post-market surveillance, and PMPF,
and review and audit requirements and provisions adopted by the manufacturer, including those in relation to fulfilling the general safety and performance requirements set out in Annex I, the documentation shall be sampled in such as a manner as to reflect the risks associated with the intended use of the device, the complexity of the manufacturing technologies, the range and classes of devices produced and any available post-market surveillance information, if not already covered by the audit programme, audit the control of processes on the premises of the manufacturer's suppliers, when the conformity of finished devices is significantly influenced by the activity of suppliers and, in particular when the manufacturer cannot demonstrate sufficient control over its suppliers, conduct assessments of the technical documentation based on its sampling plan and taking account of Section 4.5.4. for performance evaluation, the notified body shall ensure that audit findings are appropriately and consistently classified in accordance with the requirements of this Regulation and with relevant standards, or with best practice documents developed or adopted by the MDCG.
the allocation of appropriately qualified and authorised personnel for the examination of individual aspects, such as use of the device, biocompatibility, performance evaluation, risk management and sterilisation, and the assessment of conformity of the design with this Regulation, and taking account of Sections 4.5.4. and 4.5.5. This assessment shall include the examination of the implementation by manufacturers of incoming, in-process and final checks and the results thereof. If further tests or other evidence is required for the assessment of conformity with the requirements of this Regulation, the notified body in question shall carry out adequate physical or laboratory tests in relation to the device or request the manufacturer to carry out such tests.
examine and assess the technical documentation, taking account of Sections 4.5.4. and 4.5.5, and verify that the type has been manufactured in conformity with that documentation; establish a test plan identifying all relevant and critical parameters which need to be tested by the notified body or under its responsibility; document its rationale for the selection of those parameters; carry out the appropriate examinations and tests in order to verify that the solutions adopted by the manufacturer meet the general safety and performance requirements set out in Annex I. Such examinations and tests shall include all tests necessary to verify that the manufacturer has in fact applied the relevant standards it has opted to use; agree with the applicant as to where the necessary tests will be performed if they are not to be carried out directly by the notified body; assume full responsibility for test results. Test reports submitted by the manufacturer shall only be taken into account if they have been issued by conformity assessment bodies which are competent and independent of the manufacturer.
(a) have documented procedures, sufficient expertise and facilities for the verification by examination and testing of every product batch in accordance with Annexes IX and XI; (b) establish a test plan identifying all relevant and critical parameters which need to be tested by the notified body or under its responsibility in order to: verify, for class C devices, the conformity of the device with the type described in the EU type-examination certificate and with the requirements of this Regulation which apply to those devices, confirm, for class B devices, the conformity with the technical documentation referred to in Annexes II and III and with the requirements of this Regulation which apply to those devices,
(c) document its rationale for the selection of the parameters referred to in point (b); (d) have documented procedures to carry out the appropriate assessments and tests in order to verify the conformity of the device with the requirements of this Regulation by examining and testing every product batch as specified in Section 5 of Annex XI; (e) have documented procedures providing for the reaching of an agreement with the applicant concerning when and where necessary tests that are not to be carried out by the notified body itself are to be performed; (f) assume full responsibility for test results in accordance with documented procedures; test reports submitted by the manufacturer shall only be taken into account if they have been issued by conformity assessment bodies which are competent and independent of the manufacturer.
(a) the planning, conduct, assessment, reporting and updating of the performance evaluation as referred to in Annex XIII, (b) post-market surveillance and post-market performance follow up, (c) the interface with the risk management process, (d) the appraisal and analysis of the available data and its relevance with regard to demonstrating conformity with the relevant requirements in Annex I, (e) the conclusions drawn with regard to the clinical evidence and drawing up of the performance evaluation report.
the intended use specified by the manufacturer and claims for the device defined by it, the planning of the performance evaluation, the methodology for the literature search, relevant documentation from the literature search, the performance studies, post-market surveillance and post-market performance follow up, validity of equivalence claimed in relation to other devices, the demonstration of equivalence, the suitability and conclusions data from equivalent and similar devices, the performance evaluation report, justifications in relation to non-performance of performance studies or PMPF.
ensure that all steps of the conformity assessment are documented so that the conclusions of the assessment are clear and demonstrate compliance with the requirements of this Regulation and can represent objective evidence of such compliance to persons that are not themselves involved in the assessment, for example personnel in designating authorities, ensure that records that are sufficient to provide a discernible audit trail are available for quality management system audits, clearly document the conclusions of its assessment of performance evaluation in a performance evaluation assessment report, for each specific project, provide a detailed report which shall be based on a standard format containing a minimum set of elements determined by the MDCG.
clearly document the outcome of its assessment and draw clear conclusions from the verification of the manufacturer's conformity with the requirements of this Regulation, make a recommendation for a final review and for a final decision to be taken by the notified body; this recommendation shall be signed off by the member of personnel responsible in the notified body, be provided to the manufacturer in question.
ensure that the personnel assigned for the final review and decision making on specific projects are appropriately authorised and are different from the personnel who have conducted the assessments, verify that the report or reports and supporting documentation needed for decision making, including concerning resolution of non-conformities noted during assessment, are complete and sufficient with respect to the scope of the application, and verify whether there are any unresolved non-conformities preventing issuance of a certificate.
decide, based on the assessment documentation and additional information available whether the requirements of this Regulation are fulfilled, decide, based on the results of its assessment of the performance evaluation and risk management whether the post-market surveillance plan, including the PMPF plan, is adequate, decide on specific milestones for further review by the notified body of the up to date performance evaluation, decide whether specific conditions or provisions need to be defined for the certification, decide, based on the novelty, risk classification, performance evaluation and conclusions from the risk analysis of the device, on a period of certification not exceeding five years, clearly document decision making and approval steps including approval by signature of the members of personnel responsible, clearly document responsibilities and mechanisms for communication of decisions, in particular, where the final signatory of a certificate differs from the decision maker or decision makers or does not fulfil the requirements laid down in Section 3.2.7., issue a certificate or certificates in accordance with the minimum requirements laid down in Annex XII for a period of validity not exceeding five years and shall indicate whether there are specific conditions or limitations associated with the certification, issue a certificate or certificates for the applicant alone and shall not issue certificates covering multiple entities, ensure that the manufacturer is notified of the outcome of the assessment and the resultant decision and that they are entered into the electronic system referred to in Article 52.
the approved quality management system or systems or to the product-range covered, the approved design of a device, the approved type of a device, any substance incorporated in or utilised for the manufacturing of a device and being subject to the specific procedures in accordance with Section 4.5.5.
ensure that manufacturers submit for prior approval plans for changes as referred to in the first paragraph and relevant information relating to such changes, assess the changes proposed and verify whether, after these changes, the quality management system, or the design of a device or type of a device, still meets the requirements of this Regulation, notify the manufacturer of its decision and provide a report or, as applicable, a supplementary report, which shall contain the justified conclusions of its assessment.
defining how and when surveillance activities of manufacturers are to be conducted. Those procedures shall include arrangements for unannounced on-site audits of manufacturers and, where applicable, subcontractors and suppliers carrying out product tests and the monitoring of compliance with any conditions binding manufacturers and associated with certification decisions, such as updates to clinical data at defined intervals, for screening relevant sources of scientific and clinical data and post-market information relating to the scope of their designation. Such information shall be taken into account in the planning and conduct of surveillance activities, to review vigilance data to which they have access under to Article 87 in order to estimate its impact, if any, on the validity of existing certificates. The results of the evaluation and any decisions taken shall be thoroughly documented.
not to take action on the basis that the vigilance case is clearly not related to the certification granted, observe the manufacturer's and competent authorities' activities and the results of the manufacturer's investigation so as to determine whether the certification granted is at risk or whether adequate corrective action has been taken, perform extraordinary surveillance measures, such as document reviews, short-notice or unannounced audits and product testing, where it is likely that the certification granted is at risk, increase the frequency of surveillance audits, review specific products or processes on the occasion of the next audit of the manufacturer, or take any other relevant measure.
conduct surveillance audits of the manufacturer on at least an annual basis which shall be planned and conducted in line with the relevant requirements in Section 4.5., ensure that it adequately assesses the manufacturer's documentation on, and application of, the provisions on vigilance, the post-market surveillance and PMPF, sample and test devices and technical documentation, during audits, according to pre-defined sampling criteria and testing procedures to ensure that the manufacturer continuously applies the approved quality management system, ensure that the manufacturer complies with the documentation and information obligations laid down in the relevant Annexes and that its procedures take into account best practices in the implementation of quality management systems, ensure that the manufacturer does not use quality management system or device approvals in a misleading manner, gather sufficient information to determine if the quality management system continues to comply with the requirements of this Regulation, ask the manufacturer, if non-conformities are detected, for corrections, corrective actions, and where applicable preventive actions, and where necessary, impose specific restrictions on the relevant certificate, or suspend or withdraw it.
conduct an in-depth review of the performance evaluation as most recently updated by the manufacturer based on the manufacturer's post-market surveillance, on its PMPF and on clinical literature relevant to the condition being treated with the device or on clinical literature relevant to similar devices, clearly document the outcome of the in-depth review and address any specific concerns to the manufacturer or impose any specific conditions on it, ensure that the performance evaluation as most recently updated is appropriately reflected in the instructions for use and, where applicable, the summary of safety and performance.
(a) all changes to the originally approved device, including changes not yet notified, (b) experience gained from post-market surveillance, (c) experience from risk-management, (d) experience from updating the proof of compliance with the general safety and performance requirements set out in Annex I, (e) experience from reviews of the performance evaluation, including the results of any performance studies and PMPF, (f) changes to the requirements, to components of the device or to the scientific or regulatory environment, (g) changes to applied or new harmonised standards, CS or equivalent documents, and (h) changes in medical, scientific and technical knowledge, such as: new treatments, changes in test methods, new scientific findings on materials and components, including findings on their biocompatibility, experience from studies on comparable devices, data from registers and registries, experience from performance studies with comparable devices.
detection of the presence of, or exposure to, a transmissible agent in blood, blood components, cells, tissues or organs, or in any of their derivatives, in order to assess their suitability for transfusion, transplantation or cell administration; detection of the presence of, or exposure to, a transmissible agent that causes a life-threatening disease with a high or suspected high risk of propagation; determining the infectious load of a life-threatening disease where monitoring is critical in the process of patient management.
ABO system [A (ABO1), B (ABO2), AB (ABO3)]; Rhesus system [RH1 (D), RHW1, RH2 (C), RH3 (E), RH4 (c), RH5 (e)]; Kell system [Kel1 (K)]; Kidd system [JK1 (Jka), JK2 (Jkb)]; Duffy system [FY1 (Fya), FY2 (Fyb)];
(a) for detecting the presence of, or exposure to, a sexually transmitted agent; (b) for detecting the presence in cerebrospinal fluid or blood of an infectious agent without a high or suspected high risk of propagation; (c) for detecting the presence of an infectious agent, if there is a significant risk that an erroneous result would cause death or severe disability to the individual, foetus or embryo being tested, or to the individual's offspring; (d) for pre-natal screening of women in order to determine their immune status towards transmissible agents; (e) for determining infective disease status or immune status, where there is a risk that an erroneous result would lead to a patient management decision resulting in a life-threatening situation for the patient or for the patient's offspring; (f) to be used as companion diagnostics; (g) to be used for disease staging, where there is a risk that an erroneous result would lead to a patient management decision resulting in a life-threatening situation for the patient or for the patient's offspring; (h) to be used in screening, diagnosis, or staging of cancer; (i) for human genetic testing; (j) for monitoring of levels of medicinal products, substances or biological components, when there is a risk that an erroneous result will lead to a patient management decision resulting in a life-threatening situation for the patient or for the patient's offspring; (k) for management of patients suffering from a life-threatening disease or condition; (l) for screening for congenital disorders in the embryo or foetus; (m) for screening for congenital disorders in new-born babies where failure to detect and treat such disorders could lead to life-threatening situations or severe disabilities.
(a) Devices intended for self-testing are classified as class C, except for devices for the detection of pregnancy, for fertility testing and for determining cholesterol level, and devices for the detection of glucose, erythrocytes ,leucocytes and bacteria in urine, which are classified as class B.(b) Devices intended for near-patient testing are classified in their own right.
(a) products for general laboratory use, accessories which possess no critical characteristics, buffer solutions, washing solutions, and general culture media and histological stains, intended by the manufacturer to make them suitable for in vitro diagnostic procedures relating to a specific examination;(b) instruments intended by the manufacturer specifically to be used for in vitro diagnostic procedures;(c) specimen receptacles.
the name of the manufacturer and address of its registered place of business and any additional manufacturing site covered by the quality management system, and, if the manufacturer's application is lodged by its authorised representative the name of the authorised representative and the address of the authorised representative's registered place of business, all relevant information on the device or group of devices covered by the quality management system, a written declaration that no application has been lodged with any other notified body for the same device-related quality management system, or information about any previous application for the same device-related quality management system, a draft of an EU declaration of conformity in accordance with Article 17 and Annex IV for the device model covered by the conformity assessment procedure, the documentation on the manufacturer's quality management system, a documented description of the procedures in place to fulfil the obligations arising from the quality management system and required under this Regulation and of the undertaking by the manufacturer in question to apply those procedures, a description of the procedures in place to ensure that the quality management system remains adequate and effective, and the undertaking by the manufacturer to apply those procedures, the documentation on the manufacturer's post-market surveillance system, and, where applicable, on the PMPF plan, and the procedures put in place to ensure compliance with the obligations resulting from the provisions on vigilance set out in Articles 82 to 87, a description of the procedures in place to keep up to date the post-market surveillance system and, where applicable, the PMPF plan, and the procedures ensuring compliance with the obligations resulting from the provisions on vigilance set out in Articles 82 to 87, as well as the undertaking by the manufacturer to apply those procedures, documentation on the performance evaluation plan, and a description of the procedures in place to keep up to date the performance evaluation plan, taking into account the state of the art.
(a) the manufacturer's quality objectives; (b) the organisation of the business and in particular: the organisational structures with the assignment of staff responsibilities in relation to critical procedures, the responsibilities of the managerial staff and their organisational authority, the methods of monitoring whether the operation of the quality management system is efficient and in particular the ability of that system to achieve the desired design and device quality, including control of devices which fail to conform, where the design, manufacture, and/or final verification and testing of the devices, or parts of any of those processes, is carried out by another party, the methods of monitoring the efficient operation of the quality management system and in particular the type and extent of control applied to the other party, where the manufacturer does not have a registered place of business in a Member State, the draft mandate for the designation of an authorised representative and a letter of intention from the authorised representative to accept the mandate;
(c) the procedures and techniques for monitoring, verifying, validating and controlling the design of the devices, and the corresponding documentation as well as the data and records arising from those procedures and techniques. Those procedures and techniques shall specifically cover: the strategy for regulatory compliance, including processes for identification of relevant legal requirements, qualification, classification, handling of equivalence, choice of, and compliance with, conformity assessment procedures, identification of applicable general safety and performance requirements and solutions to fulfil those requirements, taking applicable CS into account and, where opted for, harmonised standards, risk management as referred to in Section 3 of Annex I, the performance evaluation, pursuant to Article 56 and Annex XIII, including PMPF, solutions for fulfilling the applicable specific requirements regarding design and construction, including appropriate pre-clinical evaluation, in particular the requirements of Chapter II of Annex I, solutions for fulfilling the applicable specific requirements regarding the information to be supplied with the device, in particular the requirements of Chapter III of Annex I, the device identification procedures drawn up and kept up to date from drawings, specifications or other relevant documents at every stage of manufacture, and management of design or quality management system changes;
(d) the verification and quality assurance techniques at the manufacturing stage and in particular the processes and procedures which are to be used, particularly as regards sterilisation, and the relevant documents, and (e) the appropriate tests and trials which are to be carried out before, during and after manufacture, the frequency with which they are to take place, and the test equipment to be used; it shall be possible to trace back adequately the calibration of that test equipment.
the documentation on its quality management system, the documentation on any findings and conclusions resulting from the application of the post-market surveillance plan, including the PMPF plan, for a representative sample of devices, and of the provisions on vigilance set out in Articles 82 to 87, the data stipulated in the part of the quality management system relating to design, such as the results of analyses, calculations, tests and the solutions adopted regarding the risk-management as referred to in Section 4 of Annex I, the data stipulated in the part of the quality management system relating to manufacture, such as quality control reports and test data, calibration data, and records on the qualifications of the personnel concerned.
(a) The manufacturer of class B, C and D devices for self-testing and near-patient testing shall lodge with the notified body an application for the assessment of the technical documentation. (b) The application shall enable the design of the device characteristics and performance(s) to be understood and shall enable conformity with the design-related requirements of this Regulation to be assessed. It shall include: (i) test reports, including results of studies carried out with intended users; (ii) where practicable, an example of the device; if required, the device shall be returned on completion of the technical documentation assessment; (iii) data showing the suitability of the device in view of its intended purpose for self-testing or near patient-testing; (iv) the information to be provided with the device on its label and its instructions for use.
The notified body may require the application to be completed by carrying out further tests or by providing further proof to allow assessment of conformity with the requirements of this Regulation. (c) The notified body shall verify the compliance of the device with the relevant requirements set out in Annex I of this Regulation. (d) The notified body shall assess the application, by using staff, employed by it, with proven knowledge and experience regarding the technology concerned and the intended purpose of the device and provide the manufacturer with a technical documentation assessment report. (e) If the device conforms to the relevant provisions of this Regulation, the notified body shall issue an EU technical documentation assessment certificate. The certificate shall contain the conclusions of the assessment, the conditions of its validity, the data needed for the identification of the approved devices and, where appropriate, a description of the intended purpose of the device. (f) Changes to the approved device shall require approval from the notified body which issued the EU technical documentation assessment certificate, where such changes could affect the safety and performance of the device or the conditions prescribed for use of the device. Where the manufacturer plans to introduce any of the above-mentioned changes, it shall inform the notified body which issued the EU technical documentation assessment certificate thereof. The notified body shall assess the planned changes and decide whether the planned changes require a new conformity assessment in accordance with Article 48 or whether they could be addressed by means of a supplement to the EU technical documentation assessment certificate. In the latter case, the notified body shall assess the changes, notify the manufacturer of its decision and, where the changes are approved, provide it with a supplement to the EU technical documentation assessment certificate.
(a) The manufacturer of a companion diagnostic shall lodge with the notified body an application for the assessment of the technical documentation. The notified body shall assess that application in accordance with the procedure laid down in Sections 4.1 to 4.8 of this Annex. (b) The application shall enable the characteristics and performance of the device to be understood, and shall enable conformity with the design-related requirements of this Regulation to be assessed, in particular, with regard to the suitability of the device in relation to the medicinal product concerned. (c) The notified body shall, before issuing an EU technical documentation assessment certificate for the companion diagnostic and on the basis of the draft summary of safety and performance and the draft instructions for use, seek a scientific opinion from one of the competent authorities designated by the Member States in accordance with Directive 2001/83/EC or from the EMA, either of which to be referred to in this Section as "the medicinal products authority consulted" depending on which has been consulted under this point, regarding the suitability of the device in relation to the medicinal product concerned. Where the medicinal product falls exclusively within the scope of the Annex to Regulation (EC) No 726/2004 of the European Parliament and of the Council , the notified body shall seek the opinion of the EMA. If the medicinal product concerned is already authorised, or if an application for its authorisation has been submitted, the notified body shall consult the medicinal products authority, or the EMA, that is responsible for the authorisation.Regulation (EC) No 726/2004 of the European Parliament and of the Council of 31 March 2004 laying down Community procedures for the authorisation and supervision of medicinal products for human and veterinary use and establishing a European Medicines Agency (OJ L 136, 30.4.2004, p. 1 ).(d) The medicinal products authority consulted shall provide its opinion, within 60 days of receipt of all the necessary documentation. This 60-day period may be extended once for a further 60 days on justified grounds. The opinion and any possible update shall be included in the documentation of the notified body concerning the device. (e) The notified body shall give due consideration to the scientific opinion referred to in point (d) when making its decision. The notified body shall convey its final decision to the medicinal products authority consulted. The EU technical documentation assessment certificate shall be delivered in accordance with point (e) of Section 5.1. (f) Before changes affecting the performance and/or the intended use and/or the suitability of the device in relation to the medicinal product concerned are made, the manufacturer shall inform the notified body of the changes. The notified body shall assess the planned changes and decide whether the planned changes require a new conformity assessment in accordance with Article 48 or whether they could be addressed by means of a supplement to the EU technical documentation assessment certificate. In the latter case, the notified body shall assess the changes and seek the opinion of the medicinal products authority consulted. The medicinal products authority consulted shall give its opinion within 30 days of receipt of the all the necessary documentation regarding the changes. A supplement to the EU technical documentation assessment certificate shall be issued in accordance with point (f) of Section 5.1.
the EU declaration of conformity, the documentation referred to in the fifth indent of Section 2.1. and, in particular, the data and records arising from the procedures referred to in point (c) of the second paragraph of Section 2.2., information on the changes referred to in Section 2.4., the documentation referred to in Sections 4.2. and point (b) of Section 5.1., and the decisions and reports from the notified body as referred to in this Annex.
the name of the manufacturer and the address of its registered place of business and, if the application is lodged by the authorised representative, the name of the authorised representative and the address of its registered place of business, the technical documentation referred to in Annexes II and III. The applicant shall make a representative sample of the device production envisaged ("type") available to the notified body. The notified body may request other samples as necessary, in the case of devices for self-testing or near-patient testing, test reports, including results of studies carried out with intended users, and data showing the handling suitability of the device in relation to its intended purpose for self-testing or near patient-testing, where practicable, an example of the device. If required, the device shall be returned on completion of the technical documentation assessment; data showing the suitability of the device in relation to its intended purpose for self-testing or near-patient testing, the information to be provided with the device on its label and its instructions for use, and a written declaration that no application has been lodged with any other notified body for the same type, or information about any previous application for the same type that was refused by another notified body or was withdrawn by the manufacturer or its authorised representative before that other notified body made its final assessment.
(a) examine the application, by using staff with proven knowledge and experience in the evaluation of the technology, and the devices concerned and the evaluation of clinical evidence. The notified body may require the application to be completed by having further tests carried out or requesting further evidence to be provided to allow assessment of conformity with the relevant requirements of this Regulation. The notified body shall carry out adequate physical or laboratory tests in relation to the device or request the manufacturer to carry out such tests; (b) examine and assess the technical documentation for conformity with the requirements of this Regulation that are applicable to the device and verify that the type has been manufactured in conformity with that documentation; it shall also record the items designed in conformity with the applicable standards referred to in Article 8 or with applicable CS, and record items not designed on the basis of the relevant standards referred to in Article 8 or of the relevant CS; (c) review the clinical evidence presented by the manufacturer in the performance evaluation report in accordance with Section 1.3.2 of Annex XIII. The notified body shall employ device reviewers with sufficient clinical expertise and, if necessary, use external clinical experts with direct and current experience relating to the clinical application of the device in question for the purposes of that review; (d) in circumstances in which the clinical evidence is partly or totally based on data from devices which are claimed to be similar or equivalent to the device under assessment, assess the suitability of using such data, taking into account factors such as new indications and innovation. The notified body shall clearly document its conclusions on the claimed equivalence, and on the relevance and adequacy of the data for demonstrating conformity; (e) clearly document the outcome of its assessment in the performance evaluation assessment report referred to in Section 4.8 of Annex IX; (f) carry out or arrange for the appropriate assessments and the physical or laboratory tests necessary to verify whether the solutions adopted by the manufacturer meet the general safety and performance requirements laid down in this Regulation in the event that the standards referred to in Article 8 or the CS have not been applied. Where the device has to be connected to another device or devices in order to operate as intended, proof shall be provided that it conforms to the general safety and performance requirements when connected to any such device or devices having the characteristics specified by the manufacturer; (g) carry out or arrange for the appropriate assessments and the physical or laboratory tests necessary to verify whether, in the event that the manufacturer has chosen to apply the relevant harmonised standards, those standards have actually been applied; (h) agree with the applicant on the place where the necessary assessments and tests are to be carried out; (i) draw up an EU type-examination report on the results of the assessments and tests carried out under points (a) to (g); (j) in the case of class D devices, request the EU reference laboratory, where designated in accordance with Article 100, to verify the performance claimed by the manufacturer and the compliance of the device with the CS, where available, or with other solutions chosen by the manufacturer to ensure a level of safety and performance that is at least equivalent. The verification shall include laboratory tests by the EU reference laboratory in accordance with Article 48(5). In addition, the notified body shall, in the cases referred to in Article 48(6) of this Regulation, consult the relevant experts referred to in Article 106 of Regulation (EU) 2017/745 following the procedure laid down in Article 48(6) of this Regulation on the performance evaluation report of the manufacturer. The EU reference laboratory shall provide a scientific opinion within 60 days. The scientific opinion of the EU reference laboratory and, where the procedure laid down in Article 48(6) is applicable, the views of the experts consulted, and any possible updates shall be included in the documentation of the notified body concerning the device. The notified body shall give due consideration to the views expressed in the scientific opinion of the EU reference laboratory, and, where applicable, to the views expressed by the experts consulted in accordance with Article 48(6), when making its decision. The notified body shall not deliver the certificate if the scientific opinion of the EU reference laboratory is unfavourable; (k) for companion diagnostics, seek the opinion, on the basis of the draft summary of safety and performance and the draft instructions for use, of one of the competent authorities designated by the Member States in accordance with Directive 2001/83/EC or the EMA (either of which to be hereinafter referred to as "the medicinal products authority consulted" depending on which has been consulted under this point) on the suitability of the device in relation to the medicinal product concerned. Where the medicinal product falls exclusively within the scope of the Annex of Regulation (EC) No 726/2004, the notified body shall consult the EMA. If the medicinal product concerned is already authorised, or if an application for its authorisation has been submitted, the notified body shall consult the medicinal products competent authority, or the EMA, that is responsible for the authorisation. The medicinal products authority consulted shall deliver its opinion within 60 days of receipt of all the necessary documentation. This 60-day period may be extended once for a further 60 days on justified grounds. The opinion of the medicinal products authority consulted and any possible update shall be included in the documentation of the notified body concerning the device. The notified body shall give due consideration to the opinion expressed by the medicinal products authority consulted when making its decision. It shall convey its final decision to the medicinal products authority consulted; and (l) draw up an EU type-examination report on the results of the assessments and tests carried out, and scientific opinions provided under, points (a) to (k), including a performance evaluation assessment report for class C or class D devices or covered by the third indent of Section 2.
the documentation referred to in the second indent of Section 2, information on the changes referred to in Section 5, copies of EU type-examination certificates, scientific opinions and reports and their additions/supplements.
all elements listed in Section 2.1 of Annex IX, the technical documentation referred to in Annexes II and III for the types approved, a copy of the EU type-examination certificates referred to in Section 4 of Annex X; if the EU type-examination certificates have been issued by the same notified body with which the application is lodged, a reference to the technical documentation and its updates and the certificates issued shall also be included in the application.
the EU declaration of conformity, the documentation referred to in the fifth indent of Section 2.1 of Annex IX, the documentation referred to in the eighth indent of Section 2.1 of Annex IX, including the EU type-examination certificate referred to in Annex X, information on the changes referred to in Section 2.4 of Annex IX, and the decisions and reports from the notified body as referred to in Sections 2.3., 3.3. and 3.4. of Annex IX.
1. Certificates shall be drawn up in one of the official languages of the Union. 2. Each certificate shall refer to only one conformity assessment procedure. 3. Certificates shall only be issued to one manufacturer. The name and address of the manufacturer included in the certificate shall be the same as that registered in the electronic system referred to in Article 27. 4. The scope of the certificates shall unambiguously describe the device or devices covered: (a) EU technical documentation assessment certificates and EU type-examination certificates shall include a clear identification, including the name, model and type, of the device or devices, the intended purpose as indicated by the manufacturer in the instructions for use and in relation to which the device has been assessed in the conformity assessment procedure, risk classification and the Basic UDI-DI as referred to in Article 24(6). (b) EU quality management system certificates and EU production quality assurance certificates shall include the identification of the devices or groups of devices, the risk classification and the intended purpose.
5. The notified body shall be able to demonstrate on request, which (individual) devices are covered by the certificate. The notified body shall set up a system that enables the determination of the devices, including their classification, covered by the certificate. 6. Certificates shall contain, if applicable, a note that, for the placing on the market of the device or devices it covers, another certificate issued in accordance with this Regulation is required. 7. EU quality management system certificates and EU production quality assurance certificates for class A sterile devices shall include a statement that the audit by the notified body was limited to the aspects of manufacture concerned with securing and maintaining sterile conditions. 8. Where a certificate is supplemented, modified or re-issued, the new certificate shall contain a reference to the preceding certificate and its date of issue with identification of the changes.
1. name, address and identification number of the notified body; 2. name and address of the manufacturer and, if applicable, of the authorised representative; 3. unique number identifying the certificate; 4. if already issued, the SRN of the manufacturer referred to in Article 28(2); 5. date of issue; 6. date of expiry; 7. data needed for the unambiguous identification of the device or devices where applicable as specified in Section 4 of this Annex; 8. if applicable, reference to any previous certificate as specified in Section 8 of Chapter I; 9. reference to this Regulation and the relevant Annex in accordance with which the conformity assessment has been carried out; 10. examinations and tests performed, e.g. reference to relevant CS, harmonised standards, test reports and audit report(s); 11. if applicable, reference to the relevant parts of the technical documentation or other certificates required for the placing on the market of the device or devices covered; 12. if applicable, information about the surveillance by the notified body; 13. conclusions of the notified body's conformity assessment with regard to the relevant Annex; 14. conditions for or limitations to the validity of the certificate; 15. legally binding signature of the notified body in accordance with the applicable national law.
a specification of the intended purpose of the device; a specification of the characteristics of the device as described in Section 9 of Chapter II of Annex I and in point (c) of Section 20.4.1. of Chapter III of Annex I; a specification of the analyte or marker to be determined by the device; a specification of the intended use of the device; identification of certified reference materials or reference measurement procedures to allow for metrological traceability; a clear identification of specified target patient groups with clear indications, limitations and contra-indications; an identification of the general safety and performance requirements as laid down in Sections 1 to 9 of Annex I that require support from relevant scientific validity and analytical and clinical performance data; a specification of methods, including the appropriate statistical tools, used for the examination of the analytical and clinical performance of the device and of the limitations of the device and information provided by it; a description of the state of the art, including an identification of existing relevant standards, CS, guidance or best practices documents; an indication and specification of parameters to be used to determine, based on the state of the art in medicine, the acceptability of the benefit-risk ratio for the intended purpose or purposes and for the analytical and clinical performance of the device; for software qualified as a device, an identification and specification of reference databases and other sources of data used as the basis for its decision making; an outline of the different development phases including the sequence and means of determination of the scientific validity, the analytical and clinical performance, including an indication of milestones and a description of potential acceptance criteria; the PMPF planning as referred to in Part B of this Annex.
identify through a systematic scientific literature review the available data relevant to the device and its intended purpose and identify any remaining unaddressed issues or gaps in the data; appraise all relevant data by evaluating their suitability for establishing the safety and performance of the device; generate any new or additional data necessary to address outstanding issues.
relevant information on the scientific validity of devices measuring the same analyte or marker; scientific (peer-reviewed) literature; consensus expert opinions/positions from relevant professional associations; results from proof of concept studies; results from clinical performance studies.
clinical performance studies; scientific peer-reviewed literature; published experience gained by routine diagnostic testing.
the justification for the approach taken to gather the clinical evidence; the literature search methodology and the literature search protocol and literature search report of a literature review; the technology on which the device is based, the intended purpose of the device and any claims made about the device's performance or safety; the nature and extent of the scientific validity and the analytical and clinical performance data that has been evaluated; the clinical evidence as the acceptable performances against the state of the art in medicine; any new conclusions derived from PMPF reports in accordance with Part B of this Annex.
(a) the single identification number of the clinical performance study, as referred to in Article 66(1); (b) identification of the sponsor, including the name, address of the registered place of business and contact details of the sponsor and, if applicable, the name, address of the registered place of business and contact details of its contact person or legal representative pursuant to Article 58(4) established in the Union; (c) information on the investigator or investigators, namely principal, coordinating or other investigator; qualifications; contact details, and investigation site or sites, such as number, qualification, contact details and, in the case of devices for self-testing, the location and number of lay persons involved; (d) the starting date and scheduled duration for the clinical performance study; (e) identification and description of the device, its intended purpose, the analyte or analytes or marker or markers, the metrological traceability, and the manufacturer; (f) information about the type of specimens under investigation; (g) overall synopsis of the clinical performance study, its design type, such as observational, interventional, together with the objectives and hypotheses of the study, reference to the current state of the art in diagnosis and/or medicine; (h) a description of the expected risks and benefits of the device and of the clinical performance study in the context of the state of the art in clinical practice, and with the exception of studies using left-over samples, the medical procedures involved and patient management; (i) the instructions for use of the device or test protocol, the necessary training and experience of the user, the appropriate calibration procedures and means of control, the indication of any other devices, medical devices, medicinal product or other articles to be included or excluded and the specifications on any comparator or comparative method used as reference; (j) description of and justification for the design of the clinical performance study, its scientific robustness and validity, including the statistical design, and details of measures to be taken to minimise bias, such as randomisation, and management of potential confounding factors; (k) the analytical performance in accordance with point (a) of Section 9.1 of Chapter I of Annex I with justification for any omission; (l) parameters of clinical performance in accordance with point (b) of Section 9.1 of Annex I to be determined, with justification for any omission; and with the exception of studies using left-over samples the specified clinical outcomes/endpoints (primary/secondary) used with a justification and the potential implications for individual health and/or public health management decisions; (m) information on the performance study population: specifications of the subjects, selection criteria, size of performance study population, representativity of target population and, if applicable, information on vulnerable subjects involved, such as children, pregnant women, immuno-compromised or elderly subjects; (n) information on use of data out of left over specimens banks, genetic or tissue banks, patient or disease registries etc. with description of reliability and representativity and statistical analysis approach; assurance of relevant method for determining the true clinical status of patient specimens; (o) monitoring plan; (p) data management; (q) decision algorithms; (r) policy regarding any amendments, including those in accordance with Article 71, to or deviations from the CPSP, with a clear prohibition of use of waivers from the CPSP; (s) accountability regarding the device, in particular control of access to the device, follow-up in relation to the device used in the clinical performance study and the return of unused, expired or malfunctioning devices; (t) statement of compliance with the recognised ethical principles for medical research involving humans and the principles of good clinical practice in the field of clinical performance studies as well as with the applicable regulatory requirements; (u) description of the informed consent process, including a copy of the patient information sheet and consent forms; (v) procedures for safety recording and reporting, including definitions of recordable and reportable events, and procedures and timelines for reporting; (w) criteria and procedures for suspension or early termination of the clinical performance study; (x) criteria and procedures for follow up of subjects following completion of a performance study, procedures for follow up of subjects in the case of suspension or early termination, procedures for follow up of subjects who have withdrawn their consent and procedures for subjects lost to follow up; (y) procedures for communication of test results outside the study, including communication of test results to the performance study subjects; (z) policy as regards the establishment of the clinical performance study report and publication of results in accordance with the legal requirements and the ethical principles referred to in Section 2.2; (aa) list of the technical and functional features of the device indicating those that are covered by the performance study; (ab) bibliography.
(a) confirming the safety and performance of the device throughout its expected lifetime, (b) identifying previously unknown risks or limits to performance and contra-indications, (c) identifying and analysing emergent risks on the basis of factual evidence, (d) ensuring the continued acceptability of the clinical evidence and of the benefit-risk ratio referred to in Sections 1 and 8 of Chapter I of Annex I, and (e) identifying possible systematic misuse.
(a) the general methods and procedures of the PMPF to be applied, such as gathering of clinical experience gained, feedback from users, screening of scientific literature and of other sources of performance or scientific data; (b) the specific methods and procedures of PMPF to be applied, such as ring trials and other quality assurance activities, epidemiological studies, evaluation of suitable patient or disease registers, genetic databanks or post-market clinical performance studies; (c) a rationale for the appropriateness of the methods and procedures referred to in points (a) and (b); (d) a reference to the relevant parts of the performance evaluation report referred to in Section 1.3 of this Annex and to the risk management referred to in Section 3 of Annex I; (e) the specific objectives to be addressed by the PMPF; (f) an evaluation of the performance data relating to equivalent or similar devices, and the current state of the art; (g) reference to any relevant CS, harmonised standards when used by the manufacturer, and relevant guidance on PMPF, and; (h) a detailed and adequately justified time schedule for PMPF activities, such as analysis of PMPF data and reporting, to be undertaken by the manufacturer.
1. Application form The application form shall be duly filled in, containing the following information: 1.1. name, address and contact details of the sponsor and, if applicable, name, address and contact details of its contact person or legal representative in accordance with Article 58(4) established in the Union; 1.2. if different from those in Section 1.1, name, address and contact details of the manufacturer of the device intended for performance evaluation and, if applicable, of its authorised representative; 1.3. title of the performance study; 1.4. single identification number in accordance with Article 66(1); 1.5. status of the performance study, such as. the first submission, resubmission, significant amendment; 1.6. details and/ or reference to the performance study plan, such as including details of the design phase of the performance study; 1.7. if the application is a resubmission with regard to a device for which an application has been already submitted, the date or dates and reference number or numbers of the earlier application or in the case of significant amendment, reference to the original application. The sponsor shall identify all of the changes from the previous application together with a rationale for those changes, in particular, whether any changes have been made to address conclusions of previous competent authority or ethics committee reviews; 1.8. if the application is submitted in parallel with an application for a clinical trial in accordance with Regulation (EU) No 536/2014, reference to the official registration number of the clinical trial; 1.9. identification of the Member States and third countries in which the clinical performance study is to be conducted as part of a multicentre or multinational study at the time of application; 1.10. brief description of the device for performance study, its classification and other information necessary for the identification of the device and device type; 1.11. summary of the performance study plan; 1.12. if applicable, information regarding a comparator device, its classification and other information necessary for the identification of the comparator device; 1.13. evidence from the sponsor that the clinical investigator and the investigational site are capable of conducting the clinical performance study in accordance with the performance study plan; 1.14. details of the anticipated start date and duration of the performance study; 1.15. details to identify the notified body, if already involved at the stage of application for the performance study; 1.16. confirmation that the sponsor is aware that the competent authority may contact the ethics committee that is assessing or has assessed the application; 1.17. the statement referred to in Section 4.1.
2. Investigator's brochure The investigator's brochure (IB) shall contain the information on the device for performance study that is relevant for the study and available at the time of application. Any updates to the IB or other relevant information that is newly available shall be brought to the attention of the investigators in a timely manner. The IB shall be clearly identified and contain in particular the following information: 2.1. Identification and description of the device, including information on the intended purpose, the risk classification and applicable classification rule pursuant to Annex VIII, design and manufacturing of the device and reference to previous and similar generations of the device. 2.2. Manufacturer's instructions for installation, maintenance, maintaining hygiene standards and for use, including storage and handling requirements, as well as, to the extent that such information is available, information to be placed the label, and instructions for use to be provided with the device when placed on the market. In addition, information relating to any relevant training required. 2.3. Analytical performance. 2.4. Existing clinical data, in particular: from relevant peer-reviewed scientific literature and available consensus expert opinions or positions from relevant professional associations relating to the safety, performance, clinical benefits to patients, design characteristics, scientific validity, clinical performance and intended purpose of the device and/or of equivalent or similar devices; other relevant clinical data available relating to the safety, scientific validity, clinical performance, clinical benefits to patients, design characteristics and intended purpose of similar devices, including details of their similarities and differences with the device in question.
2.5. Summary of the benefit-risk analysis and the risk management, including information regarding known or foreseeable risks and warnings. 2.6. In the case of devices that include tissues, cells and substances of human, animal or microbial origins detailed information on the tissues, cells and substances, and on the compliance with the relevant general safety and performance requirements and the specific risk management in relation to those tissues, cells and substances. 2.7. A list detailing the fulfilment of the relevant general safety and performance requirements set out in Annex I, including the standards and CS applied, in full or in part, as well as a description of the solutions for fulfilling the relevant general safety and performance requirements, in so far as those standards and CS have not or have only been partly fulfilled or are lacking. 2.8. A detailed description of the clinical procedures and diagnostic tests used in the course of the performance study and in particular information on any deviation from normal clinical practice.
3. Performance study plan as referred to in Sections 2 and 3 of Annex XIII. 4. Other information 4.1. A signed statement by the natural or legal person responsible for the manufacture of the device for performance study that the device in question conforms to the general safety and performance requirements laid down in Annex I apart from the aspects covered by the clinical performance study and that, with regard to those aspects, every precaution has been taken to protect the health and safety of the subject. 4.2. Where applicable according to national law, a copy of the opinion or opinions of the ethics committee or committees concerned. Where under national law the opinion or opinions of the ethics committee or committees is not required at the time of the submission of the application, a copy of the opinion or opinions shall be submitted as soon as available. 4.3. Proof of insurance cover or indemnification of subjects in case of injury, pursuant to Article 65 and the corresponding national law. 4.4. Documents to be used to obtain informed consent, including the patient information sheet and the informed consent document. 4.5 Description of the arrangements to comply with the applicable rules on the protection and confidentiality of personal data, in particular: organisational and technical arrangements that will be implemented to avoid unauthorised access, disclosure, dissemination, alteration or loss of information and personal data processed; a description of measures that will be implemented to ensure confidentiality of records and personal data of subjects; a description of measures that will be implemented in case of a data security breach in order to mitigate the possible adverse effects.
4.6. Full details of the available technical documentation, for example detailed risk analysis/management documentation or specific test reports shall be submitted to the competent authority reviewing an application upon request.
1. The sponsor shall undertake to keep available for the competent national authorities any documentation necessary to provide evidence for the documentation referred to in Chapter I of this Annex. If the sponsor is not the natural or legal person responsible for the manufacture of the device intended for performance study, that obligation may be fulfilled by that person on behalf of the sponsor. 2. The sponsor shall have an agreement in place to ensure that any serious adverse events or any other event as referred to in Article 76(2) are reported by the investigator or investigators to the sponsor in a timely manner. 3. The documentation mentioned in this Annex shall be kept for a period of time of at least 10 years after the clinical performance study with the device in question has ended, or, in the event that the device is subsequently placed on the market, for at least 10 years after the last device has been placed on the market. Each Member State shall require that the documentation referred to in this Annex is kept at the disposal of the competent authorities for the period indicated in the first subparagraph in case the sponsor, or his contact person, established within its territory, goes bankrupt or ceases its activity prior to the end of this period. 4. The sponsor shall appoint a monitor that is independent of the investigation site to ensure that the clinical performance study is conducted in accordance with the Clinical Performance Study Plan, the principles of good clinical practice and this Regulation. 5. The sponsor shall complete the follow-up of investigation subjects.
Directive 98/79/EC | This Regulation |
---|---|
Article 1(1) | Article 1(1) |
Article 1(2) | Article 2 |
Article 1(3) | points (54) and (55) of Article 2 |
Article 1(4) | — |
Article 1(5) | Article 5(4) and (5) |
Article 1(6) | Article 1(9) |
Article 1(7) | Article 1(5) |
Article 2 | Article 5(1) |
Article 3 | Article 5(2) |
Article 4(1) | Article 21 |
Article 4(2) | Article 19(1) and (2) |
Article 4(3) | Article 19(3) |
Article 4(4) | Article 10(10) |
Article 4(5) | Article 18(6) |
Article 5(1) | Article 8(1) |
Article 5(2) | — |
Article 5(3) | Article 9 |
Article 6 | — |
Article 7 | Article 107 |
Article 8 | Articles 89 and 92 |
Article 9(1) first subparagraph | Article 48(10) first subparagraph |
Article 9(1) second subparagraph | Article 48(3) second subparagraph, Article 48(7) second subparagraph and Article 48(9) second subparagraph |
Article 9(2) | Article 48(3) to (6) |
Article 9(3) | Article 48(3) to (9) |
Article 9(4) | Article 5(6) |
Article 9(5) | — |
Article 9(6) | Article 11(3) and (4) |
Article 9(7) | Article 10(7) |
Article 9(8) | Article 49(1) |
Article 9(9) | Article 49(4) |
Article 9(10) | Article 51(2) |
Article 9(11) | Article 48(12) |
Article 9(12) | Article 54(1) |
Article 9(13) | Article 48(2) |
Article 10(1) and (2), second sentence of Article 10(3) and Article 10(4) | Articles 26(3), 27 and 28 |
Article 10(3), first sentence | Article 11(1) |
Article 11(1) | Articles 82(1) and 84(2) |
Article 11(2) | Article 82(10) and Article 82(11) first subparagraph |
Article 11(3) | Article 84(7) |
Article 11(4) | — |
Article 11(5) | Article 86 |
Article 12 | Article 30 |
Article 13 | Article 93 |
Article 14(1)(a) | — |
Article 14(1)(b) | Article 47(3) and (6) |
Article 14(2) | — |
Article 14(3) | — |
Article 15(1) | Article 38 and Article 39 |
Article 15(2) | Article 32 |
Article 15(3) | Article 40(2) and (4) |
Article 15(4) | — |
Article 15(5) | Article 51(5) |
Article 15(6) | Article 51(4) |
Article 15(7) | Article 34(2) and Article 40(2) |
Article 16 | Article 18 |
Article 17 | Articles 89 to 92 |
Article 18 | Article 94 |
Article 19 | Article 102 |
Article 20 | Article 97 |
Article 21 | — |
Article 22 | — |
Article 23 | — |
Article 24 | — |