(a) in vitro diagnostic medical devices covered by Regulation (EU) 2017/746;(b) medicinal products as defined in point 2 of Article 1 of Directive 2001/83/EC. In deciding whether a product falls under Directive 2001/83/EC or under this Regulation, particular account shall be taken of the principal mode of action of the product; (c) advanced therapy medicinal products covered by Regulation (EC) No 1394/2007; (d) human blood, blood products, plasma or blood cells of human origin or devices which incorporate, when placed on the market or put into service, such blood products, plasma or cells, except for devices referred to in paragraph 8 of this Article; (e) cosmetic products covered by Regulation (EC) No 1223/2009; (f) transplants, tissues or cells of animal origin, or their derivatives, or products containing or consisting of them; however this Regulation does apply to devices manufactured utilising tissues or cells of animal origin, or their derivatives, which are non-viable or are rendered non-viable; (g) transplants, tissues or cells of human origin, or their derivatives, covered by Directive 2004/23/EC, or products containing or consisting of them; however this Regulation does apply to devices manufactured utilising derivatives of tissues or cells of human origin which are non-viable or are rendered non-viable; (h) products, other than those referred to in points (d), (f) and (g), that contain or consist of viable biological material or viable organisms, including living micro-organisms, bacteria, fungi or viruses in order to achieve or support the intended purpose of the product; (i) food covered by Regulation (EC) No 178/2002.
Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC (Text with EEA relevance. )
Modified by
- Regulation (EU) 2020/561 of the European Parliament and of the Councilof 23 April 2020amending Regulation (EU) 2017/745 on medical devices, as regards the dates of application of certain of its provisions(Text with EEA relevance), 32020R0561, April 24, 2020
- Commission Delegated Regulation (EU) 2023/502of 1 December 2022amending Regulation (EU) 2017/745 of the European Parliament and of the Council as regards the frequency of complete re-assessments of notified bodies(Text with EEA relevance), 32023R0502, March 8, 2023
- Regulation (EU) 2023/607 of the European Parliament and of the Councilof 15 March 2023amending Regulations (EU) 2017/745 and (EU) 2017/746 as regards the transitional provisions for certain medical devices and in vitro diagnostic medical devices(Text with EEA relevance), 32023R0607, March 20, 2023
- Regulation (EU) 2024/1860 of the European Parliament and of the Councilof 13 June 2024amending Regulations (EU) 2017/745 and (EU) 2017/746 as regards a gradual roll-out of Eudamed, the obligation to inform in case of interruption or discontinuation of supply, and transitional provisions for certain in vitro diagnostic medical devices(Text with EEA relevance), 32024R1860, July 9, 2024
Corrected by
- Corrigendum to Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC, 32017R0745R(01), May 3, 2019
- Corrigendum to Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC, 32017R0745R(02), December 27, 2019
(1) "medical device" means any instrument, apparatus, appliance, software, implant, reagent, material or other article intended by the manufacturer to be used, alone or in combination, for human beings for one or more of the following specific medical purposes: diagnosis, prevention, monitoring, prediction, prognosis, treatment or alleviation of disease, diagnosis, monitoring, treatment, alleviation of, or compensation for, an injury or disability, investigation, replacement or modification of the anatomy or of a physiological or pathological process or state, providing information by means of in vitro examination of specimens derived from the human body, including organ, blood and tissue donations,
and which does not achieve its principal intended action by pharmacological, immunological or metabolic means, in or on the human body, but which may be assisted in its function by such means. The following products shall also be deemed to be medical devices: devices for the control or support of conception; products specifically intended for the cleaning, disinfection or sterilisation of devices as referred to in Article 1(4) and of those referred to in the first paragraph of this point.
(2) "accessory for a medical device" means an article which, whilst not being itself a medical device, is intended by its manufacturer to be used together with one or several particular medical device(s) to specifically enable the medical device(s) to be used in accordance with its/their intended purpose(s) or to specifically and directly assist the medical functionality of the medical device(s) in terms of its/their intended purpose(s); (3) "custom-made device" means any device specifically made in accordance with a written prescription of any person authorised by national law by virtue of that person's professional qualifications which gives, under that person's responsibility, specific design characteristics, and is intended for the sole use of a particular patient exclusively to meet their individual conditions and needs. However, mass-produced devices which need to be adapted to meet the specific requirements of any professional user and devices which are mass-produced by means of industrial manufacturing processes in accordance with the written prescriptions of any authorised person shall not be considered to be custom-made devices; (4) "active device" means any device, the operation of which depends on a source of energy other than that generated by the human body for that purpose, or by gravity, and which acts by changing the density of or converting that energy. Devices intended to transmit energy, substances or other elements between an active device and the patient, without any significant change, shall not be deemed to be active devices. Software shall also be deemed to be an active device; (5) "implantable device" means any device, including those that are partially or wholly absorbed, which is intended: to be totally introduced into the human body, or to replace an epithelial surface or the surface of the eye,
by clinical intervention and which is intended to remain in place after the procedure. Any device intended to be partially introduced into the human body by clinical intervention and intended to remain in place after the procedure for at least 30 days shall also be deemed to be an implantable device; (6) "invasive device" means any device which, in whole or in part, penetrates inside the body, either through a body orifice or through the surface of the body; (7) "generic device group" means a set of devices having the same or similar intended purposes or a commonality of technology allowing them to be classified in a generic manner not reflecting specific characteristics; (8) "single-use device" means a device that is intended to be used on one individual during a single procedure; (9) "falsified device" means any device with a false presentation of its identity and/or of its source and/or its CE marking certificates or documents relating to CE marking procedures. This definition does not include unintentional non-compliance and is without prejudice to infringements of intellectual property rights; (10) "procedure pack" means a combination of products packaged together and placed on the market with the purpose of being used for a specific medical purpose; (11) "system" means a combination of products, either packaged together or not, which are intended to be inter-connected or combined to achieve a specific medical purpose; (12) "intended purpose" means the use for which a device is intended according to the data supplied by the manufacturer on the label, in the instructions for use or in promotional or sales materials or statements and as specified by the manufacturer in the clinical evaluation; (13) "label" means the written, printed or graphic information appearing either on the device itself, or on the packaging of each unit or on the packaging of multiple devices; (14) "instructions for use" means the information provided by the manufacturer to inform the user of a device's intended purpose and proper use and of any precautions to be taken; (15) "Unique Device Identifier" ("UDI") means a series of numeric or alphanumeric characters that is created through internationally accepted device identification and coding standards and that allows unambiguous identification of specific devices on the market; (16) "non-viable" means having no potential for metabolism or multiplication; (17) "derivative" means a "non-cellular substance" extracted from human or animal tissue or cells through a manufacturing process. The final substance used for manufacturing of the device in this case does not contain any cells or tissues; (18) "nanomaterial" means a natural, incidental or manufactured material containing particles in an unbound state or as an aggregate or as an agglomerate and where, for 50 % or more of the particles in the number size distribution, one or more external dimensions is in the size range 1-100 nm; Fullerenes, graphene flakes and single-wall carbon nanotubes with one or more external dimensions below 1 nm shall also be deemed to be nanomaterials; (19) "particle", for the purposes of the definition of nanomaterial in point (18), means a minute piece of matter with defined physical boundaries; (20) "agglomerate", for the purposes of the definition of nanomaterial in point (18), means a collection of weakly bound particles or aggregates where the resulting external surface area is similar to the sum of the surface areas of the individual components; (21) "aggregate", for the purposes of the definition of nanomaterial in point (18), means a particle comprising of strongly bound or fused particles; (22) "performance" means the ability of a device to achieve its intended purpose as stated by the manufacturer; (23) "risk" means the combination of the probability of occurrence of harm and the severity of that harm; (24) "benefit-risk determination" means the analysis of all assessments of benefit and risk of possible relevance for the use of the device for the intended purpose, when used in accordance with the intended purpose given by the manufacturer; (25) "compatibility" is the ability of a device, including software, when used together with one or more other devices in accordance with its intended purpose, to: (a) perform without losing or compromising the ability to perform as intended, and/or (b) integrate and/or operate without the need for modification or adaption of any part of the combined devices, and/or (c) be used together without conflict/interference or adverse reaction.
(26) "interoperability" is the ability of two or more devices, including software, from the same manufacturer or from different manufacturers, to: (a) exchange information and use the information that has been exchanged for the correct execution of a specified function without changing the content of the data, and/or (b) communicate with each other, and/or (c) work together as intended.
(27) "making available on the market" means any supply of a device, other than an investigational device, for distribution, consumption or use on the Union market in the course of a commercial activity, whether in return for payment or free of charge; (28) "placing on the market" means the first making available of a device, other than an investigational device, on the Union market; (29) "putting into service" means the stage at which a device, other than an investigational device, has been made available to the final user as being ready for use on the Union market for the first time for its intended purpose; (30) "manufacturer" means a natural or legal person who manufactures or fully refurbishes a device or has a device designed, manufactured or fully refurbished, and markets that device under its name or trade mark; (31) "fully refurbishing", for the purposes of the definition of manufacturer, means the complete rebuilding of a device already placed on the market or put into service, or the making of a new device from used devices, to bring it into conformity with this Regulation, combined with the assignment of a new lifetime to the refurbished device; (32) "authorised representative" means any natural or legal person established within the Union who has received and accepted a written mandate from a manufacturer, located outside the Union, to act on the manufacturer's behalf in relation to specified tasks with regard to the latter's obligations under this Regulation; (33) "importer" means any natural or legal person established within the Union that places a device from a third country on the Union market; (34) "distributor" means any natural or legal person in the supply chain, other than the manufacturer or the importer, that makes a device available on the market, up until the point of putting into service; (35) "economic operator" means a manufacturer, an authorised representative, an importer, a distributor or the person referred to in Article 22(1) and 22(3); (36) "health institution" means an organisation the primary purpose of which is the care or treatment of patients or the promotion of public health; (37) "user" means any healthcare professional or lay person who uses a device; (38) "lay person" means an individual who does not have formal education in a relevant field of healthcare or medical discipline; (39) "reprocessing" means a process carried out on a used device in order to allow its safe reuse including cleaning, disinfection, sterilisation and related procedures, as well as testing and restoring the technical and functional safety of the used device; (40) "conformity assessment" means the process demonstrating whether the requirements of this Regulation relating to a device have been fulfilled; (41) "conformity assessment body" means a body that performs third-party conformity assessment activities including calibration, testing, certification and inspection; (42) "notified body" means a conformity assessment body designated in accordance with this Regulation; (43) "CE marking of conformity" or "CE marking" means a marking by which a manufacturer indicates that a device is in conformity with the applicable requirements set out in this Regulation and other applicable Union harmonisation legislation providing for its affixing; (44) "clinical evaluation" means a systematic and planned process to continuously generate, collect, analyse and assess the clinical data pertaining to a device in order to verify the safety and performance, including clinical benefits, of the device when used as intended by the manufacturer; (45) "clinical investigation" means any systematic investigation involving one or more human subjects, undertaken to assess the safety or performance of a device; (46) "investigational device" means a device that is assessed in a clinical investigation; (47) "clinical investigation plan" means a document that describes the rationale, objectives, design, methodology, monitoring, statistical considerations, organisation and conduct of a clinical investigation; (48) "clinical data" means information concerning safety or performance that is generated from the use of a device and is sourced from the following: clinical investigation(s) of the device concerned, clinical investigation(s) or other studies reported in scientific literature, of a device for which equivalence to the device in question can be demonstrated, reports published in peer reviewed scientific literature on other clinical experience of either the device in question or a device for which equivalence to the device in question can be demonstrated, clinically relevant information coming from post-market surveillance, in particular the post-market clinical follow-up;
(49) "sponsor" means any individual, company, institution or organisation which takes responsibility for the initiation, for the management and setting up of the financing of the clinical investigation; (50) "subject" means an individual who participates in a clinical investigation; (51) "clinical evidence" means clinical data and clinical evaluation results pertaining to a device of a sufficient amount and quality to allow a qualified assessment of whether the device is safe and achieves the intended clinical benefit(s), when used as intended by the manufacturer; (52) "clinical performance" means the ability of a device, resulting from any direct or indirect medical effects which stem from its technical or functional characteristics, including diagnostic characteristics, to achieve its intended purpose as claimed by the manufacturer, thereby leading to a clinical benefit for patients, when used as intended by the manufacturer; (53) "clinical benefit" means the positive impact of a device on the health of an individual, expressed in terms of a meaningful, measurable, patient-relevant clinical outcome(s), including outcome(s) related to diagnosis, or a positive impact on patient management or public health; (54) "investigator" means an individual responsible for the conduct of a clinical investigation at a clinical investigation site; (55) "informed consent" means a subject's free and voluntary expression of his or her willingness to participate in a particular clinical investigation, after having been informed of all aspects of the clinical investigation that are relevant to the subject's decision to participate or, in the case of minors and of incapacitated subjects, an authorisation or agreement from their legally designated representative to include them in the clinical investigation; (56) "ethics committee" means an independent body established in a Member State in accordance with the law of that Member State and empowered to give opinions for the purposes of this Regulation, taking into account the views of laypersons, in particular patients or patients' organisations; (57) "adverse event" means any untoward medical occurrence, unintended disease or injury or any untoward clinical signs, including an abnormal laboratory finding, in subjects, users or other persons, in the context of a clinical investigation, whether or not related to the investigational device; (58) "serious adverse event" means any adverse event that led to any of the following: (a) death, (b) serious deterioration in the health of the subject, that resulted in any of the following: (i) life-threatening illness or injury, (ii) permanent impairment of a body structure or a body function, (iii) hospitalisation or prolongation of patient hospitalisation, (iv) medical or surgical intervention to prevent life-threatening illness or injury or permanent impairment to a body structure or a body function, (v) chronic disease,
(c) foetal distress, foetal death or a congenital physical or mental impairment or birth defect;
(59) "device deficiency" means any inadequacy in the identity, quality, durability, reliability, safety or performance of an investigational device, including malfunction, use errors or inadequacy in information supplied by the manufacturer; (60) "post-market surveillance" means all activities carried out by manufacturers in cooperation with other economic operators to institute and keep up to date a systematic procedure to proactively collect and review experience gained from devices they place on the market, make available on the market or put into service for the purpose of identifying any need to immediately apply any necessary corrective or preventive actions; (61) "market surveillance" means the activities carried out and measures taken by competent authorities to check and ensure that devices comply with the requirements set out in the relevant Union harmonisation legislation and do not endanger health, safety or any other aspect of public interest protection; (62) "recall" means any measure aimed at achieving the return of a device that has already been made available to the end user; (63) "withdrawal" means any measure aimed at preventing a device in the supply chain from being further made available on the market; (64) "incident" means any malfunction or deterioration in the characteristics or performance of a device made available on the market, including use-error due to ergonomic features, as well as any inadequacy in the information supplied by the manufacturer and any undesirable side-effect; (65) "serious incident" means any incident that directly or indirectly led, might have led or might lead to any of the following: (a) the death of a patient, user or other person, (b) the temporary or permanent serious deterioration of a patient's, user's or other person's state of health, (c) a serious public health threat;
(66) "serious public health threat" means an event which could result in imminent risk of death, serious deterioration in a person's state of health, or serious illness, that may require prompt remedial action, and that may cause significant morbidity or mortality in humans, or that is unusual or unexpected for the given place and time; (67) "corrective action" means action taken to eliminate the cause of a potential or actual non-conformity or other undesirable situation; (68) "field safety corrective action" means corrective action taken by a manufacturer for technical or medical reasons to prevent or reduce the risk of a serious incident in relation to a device made available on the market; (69) "field safety notice" means a communication sent by a manufacturer to users or customers in relation to a field safety corrective action; (70) "harmonised standard" means a European standard as defined in point (1)(c) of Article 2 of Regulation (EU) No 1025/2012; (71) "common specifications" (CS) means a set of technical and/or clinical requirements, other than a standard, that provides a means of complying with the legal obligations applicable to a device, process or system.
(a) the devices are not transferred to another legal entity, (b) manufacture and use of the devices occur under appropriate quality management systems, (c) the health institution justifies in its documentation that the target patient group's specific needs cannot be met, or cannot be met at the appropriate level of performance by an equivalent device available on the market, (d) the health institution provides information upon request on the use of such devices to its competent authority, which shall include a justification of their manufacturing, modification and use; (e) the health institution draws up a declaration which it shall make publicly available, including: (i) the name and address of the manufacturing health institution; (ii) the details necessary to identify the devices; (iii) a declaration that the devices meet the general safety and performance requirements set out in Annex I to this Regulation and, where applicable, information on which requirements are not fully met with a reasoned justification therefor,
(f) the health institution draws up documentation that makes it possible to have an understanding of the manufacturing facility, the manufacturing process, the design and performance data of the devices, including the intended purpose, and that is sufficiently detailed to enable the competent authority to ascertain that the general safety and performance requirements set out in Annex I to this Regulation are met; (g) the health institution takes all necessary measures to ensure that all devices are manufactured in accordance with the documentation referred to in point (f), and (h) the health institution reviews experience gained from clinical use of the devices and takes all necessary corrective actions.
(a) ascribing functions and properties to the device which the device does not have; (b) creating a false impression regarding treatment or diagnosis, functions or properties which the device does not have; (c) failing to inform the user or the patient of a likely risk associated with the use of the device in line with its intended purpose; (d) suggesting uses for the device other than those stated to form part of the intended purpose for which the conformity assessment was carried out.
(a) a strategy for regulatory compliance, including compliance with conformity assessment procedures and procedures for management of modifications to the devices covered by the system; (b) identification of applicable general safety and performance requirements and exploration of options to address those requirements; (c) responsibility of the management; (d) resource management, including selection and control of suppliers and sub-contractors; (e) risk management as set out in in Section 3 of Annex I; (f) clinical evaluation in accordance with Article 61 and Annex XIV, including PMCF; (g) product realisation, including planning, design, development, production and service provision; (h) verification of the UDI assignments made in accordance with Article 27(3) to all relevant devices and ensuring consistency and validity of information provided in accordance with Article 29; (i) setting-up, implementation and maintenance of a post-market surveillance system, in accordance with Article 83; (j) handling communication with competent authorities, notified bodies, other economic operators, customers and/or other stakeholders; (k) processes for reporting of serious incidents and field safety corrective actions in the context of vigilance; (l) management of corrective and preventive actions and verification of their effectiveness; (m) processes for monitoring and measurement of output, data analysis and product improvement.
(a) verify that the EU declaration of conformity and technical documentation have been drawn up and, where applicable, that an appropriate conformity assessment procedure has been carried out by the manufacturer; (b) keep available a copy of the technical documentation, the EU declaration of conformity and, if applicable, a copy of the relevant certificate, including any amendments and supplements, issued in accordance with Article 56, at the disposal of competent authorities for the period referred to in Article 10(8); (c) comply with the registration obligations laid down in Article 31 and verify that the manufacturer has complied with the registration obligations laid down in Articles 27 and 29; (d) in response to a request from a competent authority, provide that competent authority with all the information and documentation necessary to demonstrate the conformity of a device, in an official Union language determined by the Member State concerned; (e) forward to the manufacturer any request by a competent authority of the Member State in which the authorised representative has its registered place of business for samples, or access to a device and verify that the competent authority receives the samples or is given access to the device; (f) cooperate with the competent authorities on any preventive or corrective action taken to eliminate or, if that is not possible, mitigate the risks posed by devices; (g) immediately inform the manufacturer about complaints and reports from healthcare professionals, patients and users about suspected incidents related to a device for which they have been designated; (h) terminate the mandate if the manufacturer acts contrary to its obligations under this Regulation.
(a) the date of termination of the mandate of the outgoing authorised representative and date of beginning of the mandate of the incoming authorised representative; (b) the date until which the outgoing authorised representative may be indicated in the information supplied by the manufacturer, including any promotional material; (c) the transfer of documents, including confidentiality aspects and property rights; (d) the obligation of the outgoing authorised representative after the end of the mandate to forward to the manufacturer or incoming authorised representative any complaints or reports from healthcare professionals, patients or users about suspected incidents related to a device for which it had been designated as authorised representative.
(a) the device has been CE marked and that the EU declaration of conformity of the device has been drawn up; (b) a manufacturer is identified and that an authorised representative in accordance with Article 11 has been designated by the manufacturer; (c) the device is labelled in accordance with this Regulation and accompanied by the required instructions for use; (d) where applicable, a UDI has been assigned by the manufacturer in accordance with Article 27.
(a) the device has been CE marked and that the EU declaration of conformity of the device has been drawn up; (b) the device is accompanied by the information to be supplied by the manufacturer in accordance with Article 10(11); (c) for imported devices, the importer has complied with the requirements set out in Article 13(3); (d) that, where applicable, a UDI has been assigned by the manufacturer.
(a) a diploma, certificate or other evidence of formal qualification, awarded on completion of a university degree or of a course of study recognised as equivalent by the Member State concerned, in law, medicine, pharmacy, engineering or another relevant scientific discipline, and at least one year of professional experience in regulatory affairs or in quality management systems relating to medical devices; (b) four years of professional experience in regulatory affairs or in quality management systems relating to medical devices.
(a) the conformity of the devices is appropriately checked, in accordance with the quality management system under which the devices are manufactured, before a device is released; (b) the technical documentation and the EU declaration of conformity are drawn up and kept up-to-date; (c) the post-market surveillance obligations are complied with in accordance with Article 10(10); (d) the reporting obligations referred to in Articles 87 to 91 are fulfilled; (e) in the case of investigational devices, the statement referred to in Section 4.1 of Chapter II of Annex XV is issued.
(a) a diploma, certificate or other evidence of formal qualification, awarded on completion of a university degree or of a course of study recognised as equivalent by the Member State concerned, in law, medicine, pharmacy, engineering or another relevant scientific discipline, and at least one year of professional experience in regulatory affairs or in quality management systems relating to medical devices; (b) four years of professional experience in regulatory affairs or in quality management systems relating to medical devices.
(a) makes available on the market a device under its name, registered trade name or registered trade mark, except in cases where a distributor or importer enters into an agreement with a manufacturer whereby the manufacturer is identified as such on the label and is responsible for meeting the requirements placed on manufacturers in this Regulation; (b) changes the intended purpose of a device already placed on the market or put into service; (c) modifies a device already placed on the market or put into service in such a way that compliance with the applicable requirements may be affected.
(a) provision, including translation, of the information supplied by the manufacturer, in accordance with Section 23 of Annex I, relating to a device already placed on the market and of further information which is necessary in order to market the device in the relevant Member State; (b) changes to the outer packaging of a device already placed on the market, including a change of pack size, if the repackaging is necessary in order to market the device in the relevant Member State and if it is carried out in such conditions that the original condition of the device cannot be affected by it. In the case of devices placed on the market in sterile condition, it shall be presumed that the original condition of the device is adversely affected if the packaging that is necessary for maintaining the sterile condition is opened, damaged or otherwise negatively affected by the repackaging.
(a) the safety and performance of the reprocessed device is equivalent to that of the original device and the requirements in points (a), (b), (d), (e), (f), (g) and (h) of Article 5(5) are complied with; (b) the reprocessing is performed in accordance with CS detailing the requirements concerning: risk management, including the analysis of the construction and material, related properties of the device (reverse engineering) and procedures to detect changes in the design of the original device as well as of its planned application after reprocessing, the validation of procedures for the entire process, including cleaning steps, the product release and performance testing, the quality management system, the reporting of incidents involving devices that have been reprocessed, and the traceability of reprocessed devices.
(a) the reprocessing of single-use devices and the transfer of single-use devices to another Member State or to a third country with a view to their reprocessing; (b) the making available or further use of reprocessed single-use devices.
(a) information allowing the identification of the device, including the device name, serial number, lot number, the UDI, the device model, as well as the name, address and the website of the manufacturer; (b) any warnings, precautions or measures to be taken by the patient or a healthcare professional with regard to reciprocal interference with reasonably foreseeable external influences, medical examinations or environmental conditions; (c) any information about the expected lifetime of the device and any necessary follow-up; (d) any other information to ensure safe use of the device by the patient, including the information in point (u) of Section 23.4 of Annex I.
(a) investigational devices being supplied to an investigator for the purpose of a clinical investigation if they meet the conditions laid down in Articles 62 to 80 and Article 82, in the implementing acts adopted pursuant to Article 81 and in Annex XV; (b) custom-made devices being made available on the market if Article 52(8) and Annex XIII have been complied with.
(a) other devices bearing the CE marking; (b) in vitro diagnostic medical devices bearing the CE marking in conformity with Regulation (EU) 2017/746;(c) other products which are in conformity with legislation that applies to those products only where they are used within a medical procedure or their presence in the system or procedure pack is otherwise justified.
(a) they verified the mutual compatibility of the devices and, if applicable other products, in accordance with the manufacturers' instructions and have carried out their activities in accordance with those instructions; (b) they packaged the system or procedure pack and supplied relevant information to users incorporating the information to be supplied by the manufacturers of the devices or other products which have been put together; (c) the activity of combining devices and, if applicable, other products as a system or procedure pack was subject to appropriate methods of internal monitoring, verification and validation.
(a) any economic operator to whom they have directly supplied a device; (b) any economic operator who has directly supplied them with a device; (c) any health institution or healthcare professional to which they have directly supplied a device.
(a) production of a UDI that comprises the following: (i) a UDI device identifier ("UDI-DI") specific to a manufacturer and a device, providing access to the information laid down in Part B of Annex VI; (ii) a UDI production identifier ("UDI-PI") that identifies the unit of device production and if applicable the packaged devices, as specified in Part C of Annex VI;
(b) placing of the UDI on the label of the device or on its packaging; (c) storage of the UDI by economic operators, health institutions and healthcare professionals, in accordance with the conditions laid down in paragraphs 8 and 9 of this Article respectively; (d) establishment of an electronic system for Unique Device Identification ("UDI database") in accordance with Article 28.
(a) the entity is an organisation with legal personality; (b) its system for the assignment of UDIs is adequate to identify a device throughout its distribution and use in accordance with the requirements of this Regulation; (c) its system for the assignment of UDIs conforms to the relevant international standards; (d) the entity gives access to its system for the assignment of UDIs to all interested users in accordance with a set of predetermined and transparent terms and conditions; (e) the entity undertakes to do the following: (i) operate its system for the assignment of UDIs for at least 10 years after its designation; (ii) make available to the Commission and to the Member States, upon request, information concerning its system for the assignment of UDIs; (iii) remain in compliance with the criteria for designation and the terms of designation.
class III implantable devices; the devices, categories or groups of devices determined by a measure referred to in point (a) of paragraph 11.
(a) amending the list of information set out in Part B of Annex VI in the light of technical progress; and (b) amending Annex VI in the light of international developments and technical progress in the field of Unique Device Identification.
(a) determining the devices, categories or groups of devices to which the obligation laid down in paragraph 8 is to apply; (b) specifying the data to be included in the UDI-PI of specific devices or device groups;
(a) confidentiality and data protection as referred to in Articles 109 and 110 respectively; (b) the risk-based approach; (c) the cost-effectiveness of the measures; (d) the convergence of UDI systems developed at international level; (e) the need to avoid duplications in the UDI system; (f) the needs of the healthcare systems of the Member States, and where possible, compatibility with other medical device identification systems that are used by stakeholders.
(a) the identification of the device and the manufacturer, including the Basic UDI-DI and, if already issued, the SRN; (b) the intended purpose of the device and any indications, contraindications and target populations; (c) a description of the device, including a reference to previous generation(s) or variants if such exist, and a description of the differences, as well as, where relevant, a description of any accessories, other devices and products, which are intended to be used in combination with the device; (d) possible diagnostic or therapeutic alternatives; (e) reference to any harmonised standards and CS applied; (f) the summary of clinical evaluation as referred to in Annex XIV, and relevant information on post-market clinical follow-up; (g) suggested profile and training for users; (h) information on any residual risks and any undesirable effects, warnings and precautions.
(a) to enable the public to be adequately informed about devices placed on the market, the corresponding certificates issued by notified bodies and about the relevant economic operators; (b) to enable unique identification of devices within the internal market and to facilitate their traceability; (c) to enable the public to be adequately informed about clinical investigations and to enable sponsors of clinical investigations to comply with obligations under Articles 62 to 80, Article 82, and any acts adopted pursuant to Article 81; (d) to enable manufacturers to comply with the information obligations laid down in Articles 87 to 90 or in any acts adopted pursuant to Article 91; (e) to enable the competent authorities of the Member States and the Commission to carry out their tasks relating to this Regulation on a well-informed basis and to enhance the cooperation between them.
(a) the electronic system for registration of devices referred to in Article 29(4); (b) the UDI-database referred to in Article 28; (c) the electronic system on registration of economic operators referred to in Article 30; (d) the electronic system on notified bodies and on certificates referred to in Article 57; (e) the electronic system on clinical investigations referred to in Article 73; (f) the electronic system on vigilance and post-market surveillance referred to in Article 92; (g) the electronic system on market surveillance referred to in Article 100.
the result of the assessment, confirmation that the corrective and preventive actions have been appropriately addressed and, where required, implemented, any remaining diverging opinion with the joint assessment team, and, where applicable, the recommended scope of designation.
(a) assess the impact on the certificates issued by the notified body; (b) submit a report on its findings to the Commission and the other Member States within three months of having notified the changes to the designation; (c) require the notified body to suspend or withdraw, within a reasonable period of time determined by the authority, any certificates which were unduly issued to ensure the safety of devices on the market; (d) enter into the electronic system referred to in Article 57 information in relation to certificates of which it has required their suspension or withdrawal; (e) inform the competent authority for medical devices of the Member State in which the manufacturer has its registered place of business through the electronic system referred to in Article 57 of the certificates for which it has required suspension or withdrawal. That competent authority shall take the appropriate measures, where necessary to avoid a potential risk to the health or safety of patients, users or others.
(a) the authority responsible for notified bodies has confirmed, within one month of the suspension or restriction, that there is no safety issue in relation to certificates affected by the suspension or restriction, and the authority responsible for notified bodies has outlined a timeline and actions anticipated to remedy the suspension or restriction; or (b) the authority responsible for notified bodies has confirmed that no certificates relevant to the suspension will be issued, amended or re-issued during the course of the suspension or restriction, and states whether the notified body has the capability of continuing to monitor and remain responsible for existing certificates issued for the period of the suspension or restriction. In the event that the authority responsible for notified bodies determines that the notified body does not have the capability to support existing certificates issued, the manufacturer shall provide, to the competent authority for medical devices of the Member State in which the manufacturer of the device covered by the certificate has its registered place of business, within three months of the suspension or restriction, a written confirmation that another qualified notified body is temporarily assuming the functions of the notified body to monitor and remain responsible for the certificates during the period of suspension or restriction.
(a) where the competent authority for medical devices of the Member State in which the manufacturer of the device covered by the certificate has its registered place of business has confirmed that there is no safety issue associated with the devices in question; and (b) another notified body has confirmed in writing that it will assume immediate responsibilities for those devices and will have completed assessment of them within twelve months of the withdrawal of the designation.
(a) development of best practice documents relating to the activities of the authorities responsible for notified bodies; (b) development of guidance documents for notified bodies in relation to the implementation of this Regulation; (c) training and qualification of the experts referred to in Article 40; (d) monitoring of trends relating to changes to notified body designations and notifications and trends in certificate withdrawals and transfers between notified bodies; (e) monitoring of the application and applicability of scope codes referred to in Article 42(13); (f) development of a mechanism for peer reviews between authorities and the Commission; (g) methods of communication to the public on the monitoring and surveillance activities of authorities and the Commission on notified bodies.
(a) application of Annex VIII to a given device, or category or group of devices, with a view to determining the classification of such devices; (b) that a device, or category or group of devices, shall for reasons of public health based on new scientific evidence, or based on any information which becomes available in the course of the vigilance and market surveillance activities be reclassified, by way of derogation from Annex VIII.
(a) in the case of devices placed on the market in sterile condition, to the aspects relating to establishing, securing and maintaining sterile conditions; (b) in the case of devices with a measuring function, to the aspects relating to the conformity of the devices with the metrological requirements; (c) in the case of reusable surgical instruments, to the aspects relating to the reuse of the device, in particular cleaning, disinfection, sterilization, maintenance and functional testing and the related instructions for use.
(a) the frequency and the sampling basis of the assessment of the technical documentation on a representative basis as set out in the third paragraph of Section 2.3 and in Section 3.5 of Annex IX in the case of class IIa and class IIb devices, and in Section 10.2 of Annex XI in the case of class IIa devices; (b) the minimum frequency of unannounced on-site audits and sample tests to be conducted by notified bodies in accordance with Section 3.4 of Annex IX, taking into account the risk-class and the type of device; (c) the physical, laboratory or other tests to be carried out by notified bodies in the context of sample tests, assessment of the technical documentation and type examination in accordance with Sections 3.4 and 4.3 of Annex IX, Section 3 of Annex X and Section 15 of Annex XI.
(a) class III implantable devices, and (b) class IIb active devices intended to administer and/or remove a medicinal product, as referred to in Section 6.4 of Annex VIII (Rule 12).
(a) in the case of renewal of a certificate issued under this Regulation; (b) where the device has been designed by modifying a device already marketed by the same manufacturer for the same intended purpose, provided that the manufacturer has demonstrated to the satisfaction of the notified body that the modifications do not adversely affect the benefit-risk ratio of the device; or (c) where the principles of the clinical evaluation of the device type or category have been addressed in a CS referred to in Article 9 and the notified body confirms that the clinical evaluation of the manufacturer for this device is in compliance with the relevant CS for clinical evaluation of that kind of device.
(a) the list of subsidiaries referred to in Article 37(3); (b) the list of experts referred to in Article 40(2); (c) the information relating to the notification referred to in Article 42(10) and the amended notifications referred to in Article 46(2); (d) the list of notified bodies referred to in Article 43(2); (e) the summary of the report referred to in Article 44(12); (f) the notifications for conformity assessments and certificates referred to in Articles 54(3) and 55(1); (g) withdrawal or refusals of applications for the certificates as referred to in Article 53(2) and Section 4.3 of Annex VII; (h) the information regarding certificates referred to in Article 56(5); (i) the summary of safety and clinical performance referred to in Article 32.
(a) the date on which the certificates issued by the outgoing notified body become invalid; (b) the date until which the identification number of the outgoing notified body may be indicated in the information supplied by the manufacturer, including any promotional material; (c) the transfer of documents, including confidentiality aspects and property rights; (d) the date after which the conformity assessment tasks of the outgoing notified body is assigned to the incoming notified body; (e) the last serial number or lot number for which the outgoing notified body is responsible.
(a) a critical evaluation of the relevant scientific literature currently available relating to the safety, performance, design characteristics and intended purpose of the device, where the following conditions are satisfied: it is demonstrated that the device subject to clinical evaluation for the intended purpose is equivalent to the device to which the data relate, in accordance with Section 3 of Annex XIV, and the data adequately demonstrate compliance with the relevant general safety and performance requirements;
(b) a critical evaluation of the results of all available clinical investigations, taking duly into consideration whether the investigations were performed under Articles 62 to 80, any acts adopted pursuant to Article 81, and Annex XV; and (c) a consideration of currently available alternative treatment options for that purpose, if any.
the device has been designed by modifications of a device already marketed by the same manufacturer, the modified device has been demonstrated by the manufacturer to be equivalent to the marketed device, in accordance with Section 3 of Annex XIV and this demonstration has been endorsed by the notified body, and the clinical evaluation of the marketed device is sufficient to demonstrate conformity of the modified device with the relevant safety and performance requirements.
the two manufacturers have a contract in place that explicitly allows the manufacturer of the second device full access to the technical documentation on an ongoing basis, and the original clinical evaluation has been performed in compliance with the requirements of this Regulation,
(a) which have been lawfully placed on the market or put into service in accordance with Directive 90/385/EEC or Directive 93/42/EEC and for which the clinical evaluation: is based on sufficient clinical data, and is in compliance with the relevant product-specific CS for the clinical evaluation of that kind of device, where such a CS is available; or
(b) that are sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips or connectors for which the clinical evaluation is based on sufficient clinical data and is in compliance with the relevant product-specific CS, where such a CS is available.
(a) to establish and verify that, under normal conditions of use, a device is designed, manufactured and packaged in such a way that it is suitable for one or more of the specific purposes listed in point (1) of Article 2, and achieves the performance intended as specified by its manufacturer; (b) to establish and verify the clinical benefits of a device as specified by its manufacturer; (c) to establish and verify the clinical safety of the device and to determine any undesirable side-effects, under normal conditions of use of the device, and assess whether they constitute acceptable risks when weighed against the benefits to be achieved by the device.
(a) the clinical investigation is the subject of an authorisation by the Member State(s) in which the clinical investigation is to be conducted, in accordance with this Regulation, unless otherwise stated; (b) an ethics committee, set up in accordance with national law, has not issued a negative opinion in relation to the clinical investigation, which is valid for that entire Member State under its national law; (c) the sponsor, or its legal representative or a contact person pursuant to paragraph 2, is established in the Union; (d) vulnerable populations and subjects are appropriately protected in accordance with Articles 64 to 68; (e) the anticipated benefits to the subjects or to public health justify the foreseeable risks and inconveniences and compliance with this condition is constantly monitored; (f) the subject or, where the subject is not able to give informed consent, his or her legally designated representative has given informed consent in accordance with Article 63; (g) the subject or, where the subject is not able to give informed consent, his or her legally designated representative, has been provided with the contact details of an entity where further information can be received in case of need; (h) the rights of the subject to physical and mental integrity, to privacy and to the protection of the data concerning him or her in accordance with Directive 95/46/EC are safeguarded; (i) the clinical investigation has been designed to involve as little pain, discomfort, fear and any other foreseeable risk as possible for the subjects, and both the risk threshold and the degree of distress are specifically defined in the clinical investigation plan and constantly monitored; (j) the medical care provided to the subjects is the responsibility of an appropriately qualified medical doctor or, where appropriate, a qualified dental practitioner or any other person entitled by national law to provide the relevant patient care under clinical investigation conditions; (k) no undue influence, including that of a financial nature, is exerted on the subject, or, where applicable, on his or her legally designated representatives, to participate in the clinical investigation; (l) the investigational device(s) in question conform(s) to the applicable general safety and performance requirements set out in Annex I apart from the aspects covered by the clinical investigation and that, with regard to those aspects, every precaution has been taken to protect the health and safety of the subjects. This includes, where appropriate, technical and biological safety testing and pre-clinical evaluation, as well as provisions in the field of occupational safety and accident prevention, taking into consideration the state of the art; (m) the requirements of Annex XV are fulfilled.
(a) enable the subject or his or her legally designated representative to understand: (i) the nature, objectives, benefits, implications, risks and inconveniences of the clinical investigations; (ii) the subject's rights and guarantees regarding his or her protection, in particular his or her right to refuse to participate in and the right to withdraw from the clinical investigation at any time without any resulting detriment and without having to provide any justification; (iii) the conditions under which the clinical investigations is to be conducted, including the expected duration of the subject's participation in the clinical investigation; and (iv) the possible treatment alternatives, including the follow-up measures if the participation of the subject in the clinical investigation is discontinued;
(b) be kept comprehensive, concise, clear, relevant, and understandable to the subject or his or her legally designated representative; (c) be provided in a prior interview with a member of the investigating team who is appropriately qualified under national law; (d) include information about the applicable damage compensation system referred to in Article 69; and (e) include the Union-wide unique single identification number of the clinical investigation referred to in Article 70(1) and information about the availability of the clinical investigation results in accordance with paragraph 6 of this Article.
(a) the informed consent of their legally designated representative has been obtained; (b) the incapacitated subjects have received the information referred to in Article 63(2) in a way that is adequate in view of their capacity to understand it; (c) the explicit wish of an incapacitated subject who is capable of forming an opinion and assessing the information referred to in Article 63(2) to refuse participation in, or to withdraw from, the clinical investigation at any time, is respected by the investigator; (d) no incentives or financial inducements are given to subjects or their legally designated representatives, except for compensation for expenses and loss of earnings directly related to the participation in the clinical investigation; (e) the clinical investigation is essential with respect to incapacitated subjects and data of comparable validity cannot be obtained in clinical investigations on persons able to give informed consent, or by other research methods; (f) the clinical investigation relates directly to a medical condition from which the subject suffers; (g) there are scientific grounds for expecting that participation in the clinical investigation will produce a direct benefit to the incapacitated subject outweighing the risks and burdens involved.
(a) the informed consent of their legally designated representative has been obtained; (b) the minors have received the information referred to in Article 63(2) in a way adapted to their age and mental maturity and from investigators or members of the investigating team who are trained or experienced in working with children; (c) the explicit wish of a minor who is capable of forming an opinion and assessing the information referred to in Article 63(2) to refuse participation in, or to withdraw from, the clinical investigation at any time, is respected by the investigator; (d) no incentives or financial inducements are given to the subject or his or her legally designated representative except for compensation for expenses and loss of earnings directly related to the participation in the clinical investigation; (e) the clinical investigation is intended to investigate treatments for a medical condition that only occurs in minors or the clinical investigation is essential with respect to minors to validate data obtained in clinical investigations on persons able to give informed consent or by other research methods; (f) the clinical investigation either relates directly to a medical condition from which the minor concerned suffers or is of such a nature that it can only be carried out on minors; (g) there are scientific grounds for expecting that participation in the clinical investigation will produce a direct benefit to the minor subject outweighing the risks and burdens involved; (h) the minor shall take part in the informed consent procedure in a way adapted to his or her age and mental maturity; (i) if during a clinical investigation the minor reaches the age of legal competence to give informed consent as defined in national law, his or her express informed consent shall be obtained before that subject can continue to participate in the clinical investigation.
(a) the clinical investigation has the potential to produce a direct benefit for the pregnant or breastfeeding woman concerned, or her embryo, foetus or child after birth, outweighing the risks and burdens involved; (b) where research is undertaken on breastfeeding women, particular care is taken to avoid any adverse impact on the health of the child; (c) no incentives or financial inducements are given to the subject except for compensation for expenses and loss of earnings directly related to the participation in the clinical investigation.
(a) due to the urgency of the situation, caused by a sudden life-threatening or other sudden serious medical condition, the subject is unable to provide prior informed consent and to receive prior information on the clinical investigation; (b) there are scientific grounds to expect that participation of the subject in the clinical investigation will have the potential to produce a direct clinically relevant benefit for the subject resulting in a measurable health-related improvement alleviating the suffering and/or improving the health of the subject, or in the diagnosis of its condition; (c) it is not possible within the therapeutic window to supply all prior information to and obtain prior informed consent from his or her legally designated representative; (d) the investigator certifies that he or she is not aware of any objections to participate in the clinical investigation previously expressed by the subject; (e) the clinical investigation relates directly to the subject's medical condition because of which it is not possible within the therapeutic window to obtain prior informed consent from the subject or from his or her legally designated representative and to supply prior information, and the clinical investigation is of such a nature that it may be conducted exclusively in emergency situations; (f) the clinical investigation poses a minimal risk to, and imposes a minimal burden on, the subject in comparison with the standard treatment of the subject's condition.
(a) regarding incapacitated subjects and minors, the informed consent shall be sought by the investigator from his or her legally designated representative without undue delay and the information referred to in Article 63(2) shall be given as soon as possible to the subject and to his or her legally designated representative; (b) regarding other subjects, the informed consent shall be sought by the investigator without undue delay from the subject or his or her legally designated representative, whichever can be done sooner, and the information referred to in Article 63(2) shall be given as soon as possible to the subject or his or her legally designated representative, as applicable.
(a) in the case of investigational class I devices or in the case of non-invasive class IIa and class IIb devices, unless otherwise stated by national law, immediately after the validation date of the application pursuant to paragraph 5, and provided that a negative opinion which is valid for the entire Member State, under national law, has not been issued by an ethics committee in the Member State concerned in respect of the clinical investigation; (b) in the case of investigational devices, other than those referred to in point (a), as soon as the Member State concerned has notified the sponsor of its authorisation, and provided that a negative opinion which is valid for the entire Member State, under national law, has not been issued by an ethics committee in the Member State concerned in respect of the clinical investigation. The Member State shall notify the sponsor of the authorisation within 45 days of the validation date referred to in paragraph 5. The Member State may extend this period by a further 20 days for the purpose of consulting with experts.
(a) the demonstration of compliance of the investigational device(s) with the applicable general safety and performance requirements, apart from the aspects covered by the clinical investigation, and whether, with regard to those aspects, every precaution has been taken to protect the health and safety of the subjects. This includes, where appropriate, assurance of technical and biological safety testing and pre-clinical evaluation; (b) whether the risk-minimisation solutions employed by the sponsor are described in harmonised standards and, in those cases where the sponsor does not use harmonised standards, whether the risk-minimisation solutions provide a level of protection that is equivalent to that provided by harmonised standards; (c) whether the measures planned for the safe installation, putting into service and maintenance of the investigational device are adequate; (d) the reliability and robustness of the data generated in the clinical investigation, taking account of statistical approaches, design of the investigation and methodological aspects, including sample size, comparator and endpoints; (e) whether the requirements of Annex XV are met; (f) in the case of devices for sterile use, evidence of the validation of the manufacturer's sterilisation procedures or information on the reconditioning and sterilisation procedures which have to be conducted by the investigation site; (g) the demonstration of the safety, quality and usefulness of any components of animal or human origin or of substances, which may be considered medicinal products in accordance with Directive 2001/83/EC.
(a) the application dossier submitted pursuant to Article 70(1) remains incomplete; (b) the device or the submitted documents, especially the investigation plan and the investigator's brochure, do not correspond to the state of scientific knowledge, and the clinical investigation, in particular, is not suitable for providing evidence for the safety, performance characteristics or benefit of the device on subjects or patients, (c) the requirements of Article 62 are not met, or (d) any assessment under paragraph 3 is negative.
(a) the objective and methodology of the clinical investigation; and (b) the degree of deviation of the intervention from normal clinical practice.
(a) to create the single identification numbers for clinical investigations referred to in Article 70(1); (b) to be used as an entry point for the submission of all applications or notifications for clinical investigations referred to in Articles 70, 74, 75 and 78 and for all other submission of data, or processing of data in this context; (c) for the exchange of information relating to clinical investigations in accordance with this Regulation between the Member States and between them and the Commission including the exchange of information referred to in Articles 70 and 76; (d) for information to be provided by the sponsor in accordance with Article 77, including the clinical investigation report and its summary as required in paragraph 5 of that Article; (e) for reporting on serious adverse events and device deficiencies and related updates referred to in Article 80.
(a) protection of personal data in accordance with Regulation (EC) No 45/2001; (b) protection of commercially confidential information, especially in the investigators brochure, in particular through taking into account the status of the conformity assessment for the device, unless there is an overriding public interest in disclosure; (c) effective supervision of the conduct of the clinical investigation by the Member State(s) concerned.
(a) the Member State in which the clinical investigation is being or is to be conducted has notified the sponsor of its refusal based on the grounds referred to in Article 71(4) or on considerations of public health, subject and user safety or health, of public policy, or (b) an ethics committee in that Member State has issued a negative opinion in relation to the substantial modification to the clinical investigation, which, in accordance with national law, is valid for that entire Member State.
(a) revoke the authorisation for the clinical investigation; (b) suspend or terminate the clinical investigation; (c) require the sponsor to modify any aspect of the clinical investigation.
(a) within six days of receipt of the single application, notify the sponsor that it is the coordinating Member State ("notification date"); (b) for the purpose of the validation of the application, take into account any considerations submitted within seven days of the notification date by any Member State concerned; (c) within 10 days of the notification date, assess whether the clinical investigation falls within the scope of this Regulation and whether the application is complete, and shall notify the sponsor accordingly. Article 70(1) and (3) to (5) shall apply to the coordinating Member State in relation to that assessment; (d) establish the results of its assessment in a draft assessment report to be transmitted within 26 days of the validation date to the Member States concerned. By day 38 after the validation date, the other Member States concerned shall transmit their comments and proposals on the draft assessment report and the underlying application to the coordinating Member State which shall take due account of those comments and proposals in its finalisation of the final assessment report, to be transmitted within 45 days of the validation date to the sponsor and the other Member States concerned.
(a) when it considers that participation in the clinical investigation would lead to a subject receiving treatment inferior to that received in normal clinical practice in that Member State concerned; (b) infringement of national law; or (c) considerations as regards subject safety and data reliability and robustness submitted under point (d) of paragraph 4.
(a) any adverse event of a type identified in the clinical investigation plan as being critical to the evaluation of the results of that clinical investigation; (b) any serious adverse event; (c) any device deficiency that might have led to a serious adverse event if appropriate action had not been taken, intervention had not occurred, or circumstances had been less fortunate; (d) any new findings in relation to any event referred to in points (a) to (c).
(a) any serious adverse event that has a causal relationship with the investigational device, the comparator or the investigation procedure or where such causal relationship is reasonably possible; (b) any device deficiency that might have led to a serious adverse event if appropriate action had not been taken, intervention had not occurred, or circumstances had been less fortunate; (c) any new findings in relation to any event referred to in points (a) and (b).
(a) harmonised electronic forms for the application for clinical investigations and their assessment as referred to in Articles 70 and 78, taking into account specific categories or groups of devices; (b) the functioning of the electronic system referred to in Article 73; (c) harmonised electronic forms for the notification of PMCF investigations as referred to in Article 74(1), and of substantial modifications as referred to in Article 75; (d) the exchange of information between Member States as referred to in Article 76; (e) harmonised electronic forms for the reporting of serious adverse events and device deficiencies as referred to in Article 80; (f) the timelines for the reporting of serious adverse events and device deficiencies, taking into account the severity of the event to be reported as referred to in Article 80; (g) uniform application of the requirements regarding the clinical evidence or data needed to demonstrate compliance with the general safety and performance requirements set out in Annex I.
(a) to update the benefit-risk determination and to improve the risk management as referred to in Chapter I of Annex I; (b) to update the design and manufacturing information, the instructions for use and the labelling; (c) to update the clinical evaluation; (d) to update the summary of safety and clinical performance referred to in Article 32; (e) for the identification of needs for preventive, corrective or field safety corrective action; (f) for the identification of options to improve the usability, performance and safety of the device; (g) when relevant, to contribute to the post-market surveillance of other devices; and (h) to detect and report trends in accordance with Article 88.
(a) the conclusions of the benefit-risk determination; (b) the main findings of the PMCF; and (c) the volume of sales of the device and an estimate evaluation of the size and other characteristics of the population using the device and, where practicable, the usage frequency of the device.
(a) any serious incident involving devices made available on the Union market, except expected side-effects which are clearly documented in the product information and quantified in the technical documentation and are subject to trend reporting pursuant to Article 88; (b) any field safety corrective action in respect of devices made available on the Union market, including any field safety corrective action undertaken in a third country in relation to a device which is also legally made available on the Union market, if the reason for the field safety corrective action is not limited to the device made available in the third country.
(a) where there is concern regarding a particular serious incident or cluster of serious incidents relating to the same device or type of device of the same manufacturer in more than one Member State; (b) where the appropriateness of a field safety corrective action that is proposed by a manufacturer in more than one Member State is in question.
designation of a coordinating competent authority on a case by case basis, when required; defining the coordinated assessment process, including the tasks and responsibilities of the coordinating competent authority and the involvement of other competent authorities.
(a) the typology of serious incidents and field safety corrective actions in relation to specific devices, or categories or groups of devices; (b) the reporting of serious incidents and field safety corrective actions and field safety notices, and the provision of periodic summary reports, post-market surveillance reports, PSURs and trend reports by manufacturers as referred to in Articles 85, 86, 87, 88 and 89 respectively; (c) standard structured forms for electronic and non-electronic reporting, including a minimum data set for reporting of suspected serious incidents by healthcare professionals, users and patients; (d) timelines for the reporting of field safety corrective actions, and for the provision by manufacturers of periodic summary reports and trend reports, taking into account the severity of the incident to be reported as referred to in Article 87; (e) harmonised forms for the exchange of information between competent authorities as referred to in Article 89; (f) procedures for the designation of a coordinating competent authority; the coordinated evaluation process, including tasks and responsibilities of the coordinating competent authority and involvement of other competent authorities in this process.
(a) the reports by manufacturers on serious incidents and field safety corrective actions referred to in Article 87(1) and Article 89(5); (b) the periodic summary reports by manufacturers referred to in Article 87(9); (c) the reports by manufacturers on trends referred to in Article 88; (d) the PSURs referred to in Article 86; (e) the field safety notices by manufacturers referred to in Article 89(8); (f) the information to be exchanged between the competent authorities of the Member States and between them and the Commission in accordance with Article 89(7) and (9).
(a) the Member States in which the field safety corrective action is being or is to be undertaken; (b) the Member State in which the manufacturer has its registered place of business.
(a) the Member State or Member States participating in the coordination procedure in accordance with Article 89(9) and which have agreed on the periodic summary report; (b) the Member State in which the manufacturer has its registered place of business.
(a) may require economic operators to, inter alia , make available the documentation and information necessary for the purpose of carrying out the authorities' activities and, where justified, to provide the necessary samples of devices or access to devices free of charge; and(b) shall carry out both announced and, if necessary, unannounced inspections of the premises of economic operators, as well as suppliers and/or subcontractors, and, where necessary, at the facilities of professional users.
(a) may present an unacceptable risk to the health or safety of patients, users or other persons, or to other aspects of the protection of public health; or (b) otherwise does not comply with the requirements laid down in this Regulation,
(a) summaries of the results of the surveillance activities referred to in Article 93(4); (b) the final inspection report referred to in Article 93(7); (c) information in relation to devices presenting an unacceptable risk to health and safety as referred to in Article 95(2), (4) and (6); (d) information in relation to non-compliance of products as referred to in Article 97(2); (e) information in relation to the preventive health protection measures referred to in Article 98(2); (f) summaries of the results of the reviews and assessments of the market surveillance activities of the Member States referred to in 93(8).
the adoption of opinions or recommendations or other positions, including in cases of urgency; the delegation of tasks to reporting and co-reporting members; the implementation of Article 107 regarding conflict of interests; the functioning of sub-groups.
(a) to contribute to the assessment of applicant conformity assessment bodies and notified bodies pursuant to the provisions set out in Chapter IV; (b) to advise the Commission, at its request, in matters concerning the coordination group of notified bodies as established pursuant to Article 49; (c) to contribute to the development of guidance aimed at ensuring effective and harmonised implementation of this Regulation, in particular regarding the designation and monitoring of notified bodies, application of the general safety and performance requirements and conduct of clinical evaluations and investigations by manufacturers, assessment by notified bodies and vigilance activities; (d) to contribute to the continuous monitoring of technical progress and assessment of whether the general safety and performance requirements laid down in this Regulation and Regulation (EU) 2017/746 are adequate to ensure safety and performance of devices, and thereby contribute to identifying whether there is a need to amend Annex I to this Regulation; (e) to contribute to the development of device standards, of CS and of scientific guidelines, including product specific guidelines, on clinical investigation of certain devices in particular implantable devices and class III devices; (f) to assist the competent authorities of the Member States in their coordination activities in particular in the fields of classification and the determination of the regulatory status of devices, clinical investigations, vigilance and market surveillance including the development and maintenance of a framework for a European market surveillance programme with the objective of achieving efficiency and harmonisation of market surveillance in the Union, in accordance with Article 93; (g) to provide advice, either on its own initiative or at request of the Commission, in the assessment of any issue related to the implementation of this Regulation; (h) to contribute to harmonised administrative practice with regard to devices in the Member States.
physico-chemical characterisation, or microbiological, biocompatibility, mechanical, electrical, electronic or non-clinical biological and toxicological testing
(a) have adequate and appropriately qualified staff with adequate knowledge and experience in the field of the devices for which they are designated; (b) possess the necessary equipment to carry out the tasks assigned to them; (c) have the necessary knowledge of international standards and best practices; (d) have an appropriate administrative organisation and structure; (e) ensure that their staff observe the confidentiality of information and data obtained in carrying out their tasks.
(a) to provide scientific, technical and clinical assistance to the Commission and the MDCG in relation to the implementation of this Regulation; (b) to contribute to the development and maintenance of appropriate guidance and CS for: clinical investigations, clinical evaluation and PMCF, performance studies, performance evaluation and post-market performance follow-up, physico-chemical characterisation, and microbiological, biocompatibility, mechanical, electrical, electronic or non-clinical toxicological testing
for specific devices, or a category or group of devices, or for specific hazards related to a category or group of devices; (c) to develop and review clinical evaluation guidance and performance evaluation guidance for performance of conformity assessment in line with the state of the art with regard to clinical evaluation, performance evaluation, physico-chemical characterisation, and microbiological, biocompatibility, mechanical, electrical, electronic or non-clinical toxicological testing; (d) to contribute to the development of standards at international level, ensuring that such standards reflect the state of the art; (e) to provide opinions in response to consultations by manufacturers in accordance with Article 61(2), notified bodies and Member States in accordance with paragraphs 11 to 13 of this Article. (f) to contribute to identification of concerns and emerging issues on the safety and performance of medical devices; (g) to provide views in accordance with Article 48(4) of Regulation (EU) 2017/746 on the performance evaluation of certain in vitro diagnostic medical devices.
(a) personal data, in accordance with Article 110; (b) commercially confidential information and trade secrets of a natural or legal person, including intellectual property rights; unless disclosure is in the public interest; (c) the effective implementation of this Regulation, in particular for the purpose of inspections, investigations or audits.
----------------------
----------------------
(a) before the date of expiry of the certificate, the manufacturer and a notified body have signed a written agreement in accordance with Section 4.3, second subparagraph, of Annex VII to this Regulation for the conformity assessment in respect of the device covered by the expired certificate or in respect of a device intended to substitute that device; (b) a competent authority of a Member State has granted a derogation from the applicable conformity assessment procedure in accordance with Article 59(1) of this Regulation or has required the manufacturer, in accordance with Article 97(1) of this Regulation, to carry out the applicable conformity assessment procedure.
(a) 31 December 2027 , for all class III devices, and for class IIb implantable devices except sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors;(b) 31 December 2028 , for class IIb devices other than those covered by point (a) of this paragraph, for class IIa devices, and for class I devices placed on the market in sterile condition or having a measuring function.
(a) those devices continue to comply with Directive 90/385/EEC or Directive 93/42/EEC, as applicable; (b) there are no significant changes in the design and intended purpose; (c) the devices do not present an unacceptable risk to the health or safety of patients, users or other persons, or to other aspects of the protection of public health; (d) no later than 26 May 2024 , the manufacturer has put in place a quality management system in accordance with Article 10(9);(e) no later than 26 May 2024 , the manufacturer or the authorised representative has lodged a formal application with a notified body in accordance with Section 4.3, first subparagraph, of Annex VII for conformity assessment in respect of a device referred to in paragraph 3a or 3b of this Article or in respect of a device intended to substitute that device, and, no later than26 September 2024 , the notified body and the manufacturer have signed a written agreement in accordance with Section 4.3, second subparagraph, of Annex VII.
Articles 8 and 10, Article 10b(1), points (b) and (c), and Article 10b(2) and (3) of Directive 90/385/EEC, Article 10, Article 14a(1), points (c) and (d), Article 14a(2) and (3) and Article 15 of Directive 93/42/EEC, and the obligations relating to vigilance and clinical investigations provided for in the corresponding Annexes to those Directives, which are repealed, as applicable, with effect from the date referred to in Article 123(3), point (d), of this Regulation in respect of the application of the obligations and requirements that relate to the electronic systems referred to in Article 33(2), points (e) and (f), respectively, of this Regulation; Article 10a, Article 10b(1), point (a), and Article 11(5) of Directive 90/385/EEC, Article 14(1) and (2), Article 14a(1), points (a) and (b), and Article 16(5) of Directive 93/42/EEC, and the obligations relating to registration of devices and economic operators, and to certificate notifications, provided for in the corresponding Annexes to those Directives, which are repealed, as applicable, with effect from the date referred to in Article 123(3), point (d), of this Regulation in respect of the application of the obligations and requirements that relate to the electronic systems referred to in Article 33(2), points (a) to (d), respectively, of this Regulation; Article 9(9) of Directive 90/385/EEC and Article 11(13) of Directive 93/42/EEC, which are repealed with effect from 24 April 2020 .
(a) Articles 35 to 50 shall apply from 26 November 2017 . However, from that date until26 May 2021 , the obligations on notified bodies pursuant to Articles 35 to 50 shall apply only to those bodies which submit an application for designation in accordance with Article 38;(b) Articles 101 and 103 shall apply from 26 November 2017 ;(c) Article 102 shall apply from 26 May 2018 ;(d) without prejudice to the obligations of the Commission pursuant to Article 34, the obligations and requirements that relate to any of the electronic systems referred to in Article 33(2) shall apply from the date corresponding to 6 months from the date of publication of the notice referred to in Article 34(3), informing that the relevant electronic system is functional and meets the functional specifications drawn up pursuant to Article 34(1). The provisions referred to in the preceding sentence are: Article 29, Article 31, Article 32, Article 33(4), the second sentence of Article 40(2), Article 42(10), Article 43(2), the second subparagraph of Article 44(12), points (d) and (e) of Article 46(7), Article 53(2), Article 54(3), Article 55(1), Article 56(5), Articles 70 to 77, Article 78(1) to (13), without prejudice to Article 78(14), Articles 79 to 82, Article 86(2), Articles 87 and 88, Article 89(5) and (7), and the third subparagraph of Article 89(8), Article 90, Article 93(4), (7) and (8), Article 95(2) and (4), the last sentence of Article 97(2), Article 99(4), Article 120(3d).
Until the date of application of the provisions referred to in the first subparagraph of this point, the corresponding provisions of Directives 90/385/EEC and 93/42/EEC regarding information on vigilance reporting, clinical investigations, registration of devices and economic operators, and certificate notifications shall continue to apply. (e) no later than 12 months from the date of publication of the notice referred to in Article 34(3) in respect of the electronic system referred to in Article 33(2), points (a) and (b), manufacturers shall ensure that the information to be entered in Eudamed in accordance with Article 29 is entered in that electronic system, including regarding the following devices, provided that those devices are also placed on the market from 6 months from the date of publication of that notice: (i) devices, other than custom-made devices, for which the manufacturer has undertaken a conformity assessment in accordance with Article 52; (ii) devices, other than custom-made devices, placed on the market pursuant to Article 120(3), (3a) or (3b), unless the device, for which the manufacturer has undertaken a conformity assessment in accordance with Article 52, is already registered in Eudamed;
(ea) no later than 18 months from the date of publication of the notice referred to in Article 34(3) in respect of the electronic system referred to in Article 33(2), point (d), notified bodies shall ensure that the information to be entered in Eudamed in accordance with Article 56(5) is entered in that electronic system, including regarding devices referred to in point (e)(i) of this paragraph; for those devices, only the latest relevant certificate and, where applicable, any subsequent decision taken by the notified body related to such certificate shall be entered; (eb) by way of derogation from point (d), first subparagraph, of this paragraph, the obligations to upload the summary of safety and clinical performance in accordance with Article 32(1) and to notify competent authorities in accordance with Article 55(1), through the electronic system referred in Article 33(2), point (d), shall apply to devices referred to in point (e) of this paragraph when the certificate is entered in Eudamed in accordance with point (ea) of this paragraph; (ec) without prejudice to point (d), first subparagraph, of this paragraph, when a manufacturer is to submit a PSUR in accordance with Article 86(2) of this Regulation, to report a serious incident or a field safety corrective action in accordance with Article 87 of this Regulation, or to submit a trend report in accordance with Article 88 of this Regulation through the electronic system referred to in Article 33(2), point (f), of this Regulation, it shall also register the device, which is the subject of the PSUR or the vigilance reporting, in the electronic system referred to in Article 33(2), points (a) and (b), of this Regulation, except if such device was placed on the market in accordance with Directive 90/385/EEC or 93/42/EEC; (f) for implantable devices and for class III devices Article 27(4) shall apply from 26 May 2021 . For class IIa and class IIb devices Article 27(4) shall apply from26 May 2023 . For class I devices Article 27(4) shall apply from26 May 2025 ;(g) with regard to reusable devices that are required to bear the UDI carrier on the device itself, Article 27(4) shall apply to: (i) implantable devices and class III devices from 26 May 2023 ;(ii) class IIa and class IIb devices from 26 May 2025 ;(iii) class I devices from 26 May 2027 ;
(h) The procedure set out in Article 78 shall apply from 26 May 2027 , without prejudice to Article 78(14);(i) Article 120(12) shall apply from 26 May 2019 ;(j) Article 59 shall apply from 24 April 2020 .
(a) establish and document a risk management plan for each device; (b) identify and analyse the known and foreseeable hazards associated with each device; (c) estimate and evaluate the risks associated with, and occurring during, the intended use and during reasonably foreseeable misuse; (d) eliminate or control the risks referred to in point (c) in accordance with the requirements of Section 4; (e) evaluate the impact of information from the production phase and, in particular, from the post-market surveillance system, on hazards and the frequency of occurrence thereof, on estimates of their associated risks, as well as on the overall risk, benefit-risk ratio and risk acceptability; and (f) based on the evaluation of the impact of the information referred to in point (e), if necessary amend control measures in line with the requirements of Section 4.
(a) eliminate or reduce risks as far as possible through safe design and manufacture; (b) where appropriate, take adequate protection measures, including alarms if necessary, in relation to risks that cannot be eliminated; and (c) provide information for safety (warnings/precautions/contra-indications) and, where appropriate, training to users.
(a) reduce as far as possible the risks related to the ergonomic features of the device and the environment in which the device is intended to be used (design for patient safety), and (b) give consideration to the technical knowledge, experience, education, training and use environment, where applicable, and the medical and physical conditions of intended users (design for lay, professional, disabled or other users).
(a) the choice of materials and substances used, particularly as regards toxicity and, where relevant, flammability; (b) the compatibility between the materials and substances used and biological tissues, cells and body fluids, taking account of the intended purpose of the device and, where relevant, absorption, distribution, metabolism and excretion; (c) the compatibility between the different parts of a device which consists of more than one implantable part; (d) the impact of processes on material properties; (e) where appropriate, the results of biophysical or modelling research the validity of which has been demonstrated beforehand; (f) the mechanical properties of the materials used, reflecting, where appropriate, matters such as strength, ductility, fracture resistance, wear resistance and fatigue resistance; (g) surface properties; and (h) the confirmation that the device meets any defined chemical and/or physical specifications.
are invasive and come into direct contact with the human body, (re)administer medicines, body liquids or other substances, including gases, to/from the body, or transport or store such medicines, body fluids or substances, including gases, to be (re)administered to the body,
(a) substances which are carcinogenic, mutagenic or toxic to reproduction ("CMR"), of category 1A or 1B, in accordance with Part 3 of Annex VI to Regulation (EC) No 1272/2008 of the European Parliament and of the Council , orRegulation (EC) No 1272/2008 of the European Parliament and of the Council of 16 December 2008 on classification, labelling and packaging of substances and mixtures, amending and repealing Directives 67/548/EEC and 1999/45/EC, and amending Regulation (EC) No 1907/2006 (OJ L 353, 31.12.2008, p. 1 ).(b) substances having endocrine-disrupting properties for which there is scientific evidence of probable serious effects to human health and which are identified either in accordance with the procedure set out in Article 59 of Regulation (EC) No 1907/2006 of the European Parliament and of the Council or, once a delegated act has been adopted by the Commission pursuant to the first subparagraph of Article 5(3) of Regulation (EU) No 528/2012 of the European Parliament and the CouncilRegulation (EC) No 1907/2006 of the European Parliament and of the Council of 18 December 2006 concerning the Registration, Evaluation, Authorisation and Restriction of Chemicals (REACH) (OJ L 396, 30.12.2006, p. 1 ). , in accordance with the criteria that are relevant to human health amongst the criteria established therein.Regulation (EU) No 528/2012 of the European Parliament and the Council of 22 May 2012 concerning the making available on the market of and use of biocidal products (OJ L 167, 27.6.2012, p. 1 ).
(a) an analysis and estimation of potential patient or user exposure to the substance; (b) an analysis of possible alternative substances, materials or designs, including, where available, information about independent research, peer-reviewed studies, scientific opinions from relevant scientific committees and an analysis of the availability of such alternatives; (c) argumentation as to why possible substance and/ or material substitutes, if available, or design changes, if feasible, are inappropriate in relation to maintaining the functionality, performance and the benefit-risk ratios of the product; including taking into account if the intended use of such devices includes treatment of children or treatment of pregnant or breastfeeding women or treatment of other patient groups considered particularly vulnerable to such substances and/or materials; and (d) where applicable and available, the latest relevant scientific committee guidelines in accordance with Sections 10.4.3. and 10.4.4.
(a) reduce as far as possible and appropriate the risks from unintended cuts and pricks, such as needle stick injuries, (b) allow easy and safe handling, (c) reduce as far as possible any microbial leakage from the device and/or microbial exposure during use, and (d) prevent microbial contamination of the device or its content such as specimens or fluids.
(a) donation, procurement and testing of the tissues and cells shall be done in accordance with Directive 2004/23/EC; (b) processing, preservation and any other handling of those tissues and cells or their derivatives shall be carried out so as to provide safety for patients, users and, where applicable, other persons. In particular, safety with regard to viruses and other transmissible agents shall be addressed by appropriate methods of sourcing and by implementation of validated methods of elimination or inactivation in the course of the manufacturing process; (c) the traceability system for those devices shall be complementary and compatible with the traceability and data protection requirements laid down in Directive 2004/23/EC and in Directive 2002/98/EC.
(a) where feasible taking into account the animal species, tissues and cells of animal origin, or their derivatives, shall originate from animals that have been subjected to veterinary controls that are adapted to the intended use of the tissues. Information on the geographical origin of the animals shall be retained by manufacturers; (b) sourcing, processing, preservation, testing and handling of tissues, cells and substances of animal origin, or their derivatives, shall be carried out so as to provide safety for patients, users and, where applicable, other persons. In particular safety with regard to viruses and other transmissible agents shall be addressed by implementation of validated methods of elimination or viral inactivation in the course of the manufacturing process, except when the use of such methods would lead to unacceptable degradation compromising the clinical benefit of the device; (c) in the case of devices manufactured utilising tissues or cells of animal origin, or their derivatives, as referred to in Regulation (EU) No 722/2012 the particular requirements laid down in that Regulation shall apply.
(a) the risk of injury, in connection with their physical features, including the volume/pressure ratio, dimensional and where appropriate ergonomic features; (b) risks connected with reasonably foreseeable external influences or environmental conditions, such as magnetic fields, external electrical and electromagnetic effects, electrostatic discharge, radiation associated with diagnostic or therapeutic procedures, pressure, humidity, temperature, variations in pressure and acceleration or radio signal interferences; (c) the risks associated with the use of the device when it comes into contact with materials, liquids, and substances, including gases, to which it is exposed during normal conditions of use; (d) the risks associated with the possible negative interaction between software and the IT environment within which it operates and interacts; (e) the risks of accidental ingress of substances into the device; (f) the risks of reciprocal interference with other devices normally used in the investigations or for the treatment given; and (g) risks arising where maintenance or calibration are not possible (as with implants), from ageing of materials used or loss of accuracy of any measuring or control mechanism.
(a) Devices shall be designed, manufactured and packaged in such a way that exposure of patients, users and other persons to radiation is reduced as far as possible, and in a manner that is compatible with the intended purpose, whilst not restricting the application of appropriate specified levels for therapeutic and diagnostic purposes. (b) The operating instructions for devices emitting hazardous or potentially hazardous radiation shall contain detailed information as to the nature of the emitted radiation, the means of protecting the patient and the user, and on ways of avoiding misuse and of reducing the risks inherent to installation as far as possible and appropriate. Information regarding the acceptance and performance testing, the acceptance criteria, and the maintenance procedure shall also be specified.
(a) Where devices are designed to emit hazardous, or potentially hazardous, levels of ionizing and/or non-ionizing radiation necessary for a specific medical purpose the benefit of which is considered to outweigh the risks inherent to the emission, it shall be possible for the user to control the emissions. Such devices shall be designed and manufactured to ensure reproducibility of relevant variable parameters within an acceptable tolerance. (b) Where devices are intended to emit hazardous, or potentially hazardous, ionizing and/or non-ionizing radiation, they shall be fitted, where possible, with visual displays and/or audible warnings of such emissions.
(a) Devices intended to emit ionizing radiation shall be designed and manufactured taking into account the requirements of the Directive 2013/59/Euratom laying down basic safety standards for protection against the dangers arising from exposure to ionising radiation. (b) Devices intended to emit ionising radiation shall be designed and manufactured in such a way as to ensure that, where possible, taking into account the intended use, the quantity, geometry and quality of the radiation emitted can be varied and controlled, and, if possible, monitored during treatment. (c) Devices emitting ionising radiation intended for diagnostic radiology shall be designed and manufactured in such a way as to achieve an image and/or output quality that are appropriate to the intended medical purpose whilst minimising radiation exposure of the patient and user. (d) Devices that emit ionising radiation and are intended for therapeutic radiology shall be designed and manufactured in such a way as to enable reliable monitoring and control of the delivered dose, the beam type, energy and, where appropriate, the quality of radiation.
(a) risks connected with the use of energy sources with particular reference, where electricity is used, to insulation, leakage currents and overheating of the devices, (b) risks connected with medical treatment, in particular those resulting from the use of defibrillators or high-frequency surgical equipment, and (c) risks which may arise where maintenance and calibration are impossible, including: excessive increase of leakage currents, ageing of the materials used, excess heat generated by the device, decreased accuracy of any measuring or control mechanism.
if applicable, the compatibility of the devices with the substances they are intended to administer, and the reliability of the source of energy.
ensure that the device can be used safely and accurately by the intended user at all stages of the procedure, if necessary after appropriate training and/or information, reduce, as far as possible and appropriate, the risk from unintended cuts and pricks such as needle stick injuries, and reduce as far as possible the risk of error by the intended user in the handling of the device and, if applicable, in the interpretation of the results.
can verify that, at the time of use, the device will perform as intended by the manufacturer, and if applicable, is warned if the device has failed to provide a valid result.
(a) The medium, format, content, legibility, and location of the label and instructions for use shall be appropriate to the particular device, its intended purpose and the technical knowledge, experience, education or training of the intended user(s). In particular, instructions for use shall be written in terms readily understood by the intended user and, where appropriate, supplemented with drawings and diagrams. (b) The information required on the label shall be provided on the device itself. If this is not practicable or appropriate, some or all of the information may appear on the packaging for each unit, and/or on the packaging of multiple devices. (c) Labels shall be provided in a human-readable format and may be supplemented by machine-readable information, such as radio-frequency identification ("RFID") or bar codes. (d) Instructions for use shall be provided together with devices. By way of exception, instructions for use shall not be required for class I and class IIa devices if such devices can be used safely without any such instructions and unless otherwise provided for elsewhere in this Section. (e) Where multiple devices are supplied to a single user and/or location, a single copy of the instructions for use may be provided if so agreed by the purchaser who in any case may request further copies to be provided free of charge. (f) Instructions for use may be provided to the user in non-paper format (e.g. electronic) to the extent, and only under the conditions, set out in Regulation (EU) No 207/2012 or in any subsequent implementing rules adopted pursuant to this Regulation. (g) Residual risks which are required to be communicated to the user and/or other person shall be included as limitations, contra-indications, precautions or warnings in the information supplied by the manufacturer. (h) Where appropriate, the information supplied by the manufacturer shall take the form of internationally recognised symbols. Any symbol or identification colour used shall conform to the harmonised standards or CS. In areas for which no harmonised standards or CS exist, the symbols and colours shall be described in the documentation supplied with the device.
(a) the name or trade name of the device; (b) the details strictly necessary for a user to identify the device, the contents of the packaging and, where it is not obvious for the user, the intended purpose of the device; (c) the name, registered trade name or registered trade mark of the manufacturer and the address of its registered place of business; (d) if the manufacturer has its registered place of business outside the Union, the name of the authorised representative and address of the registered place of business of the authorised representative; (e) where applicable, an indication that the device contains or incorporates: a medicinal substance, including a human blood or plasma derivative, or tissues or cells, or their derivatives, of human origin, or tissues or cells of animal origin, or their derivatives, as referred to in Regulation (EU) No 722/2012;
(f) where applicable, information labelled in accordance with Section 10.4.5.; (g) the lot number or the serial number of the device preceded by the words LOT NUMBER or SERIAL NUMBER or an equivalent symbol, as appropriate; (h) the UDI carrier referred to in Article 27(4) and Part C of Annex VI; (i) an unambiguous indication of t the time limit for using or implanting the device safely, expressed at least in terms of year and month, where this is relevant; (j) where there is no indication of the date until when it may be used safely, the date of manufacture. This date of manufacture may be included as part of the lot number or serial number, provided the date is clearly identifiable; (k) an indication of any special storage and/or handling condition that applies; (l) if the device is supplied sterile, an indication of its sterile state and the sterilisation method; (m) warnings or precautions to be taken that need to be brought to the immediate attention of the user of the device, and to any other person. This information may be kept to a minimum in which case more detailed information shall appear in the instructions for use, taking into account the intended users; (n) if the device is intended for single use, an indication of that fact. A manufacturer's indication of single use shall be consistent across the Union; (o) if the device is a single-use device that has been reprocessed, an indication of that fact, the number of reprocessing cycles already performed, and any limitation as regards the number of reprocessing cycles; (p) if the device is custom-made, the words "custom-made device"; (q) an indication that the device is a medical device. If the device is intended for clinical investigation only, the words "exclusively for clinical investigation"; (r) in the case of devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body via a body orifice or applied to the skin and that are absorbed by or locally dispersed in the human body, the overall qualitative composition of the device and quantitative information on the main constituent or constituents responsible for achieving the principal intended action; (s) for active implantable devices, the serial number, and for other implantable devices, the serial number or the lot number.
(a) an indication permitting the sterile packaging to be recognised as such, (b) a declaration that the device is in a sterile condition, (c) the method of sterilisation, (d) the name and address of the manufacturer, (e) a description of the device, (f) if the device is intended for clinical investigations, the words "exclusively for clinical investigations", (g) if the device is custom-made, the words "custom-made device", (h) the month and year of manufacture, (i) an unambiguous indication of the time limit for using or implanting the device safely expressed at least in terms of year and month, and (j) an instruction to check the instructions for use for what to do if the sterile packaging is damaged or unintentionally opened before use.
(a) the particulars referred to in points (a), (c), (e), (f), (k), (l), (n) and (r) of Section 23.2; (b) the device's intended purpose with a clear specification of indications, contra-indications, the patient target group or groups, and of the intended users, as appropriate; (c) where applicable, a specification of the clinical benefits to be expected. (d) where applicable, links to the summary of safety and clinical performance referred to in Article 32; (e) the performance characteristics of the device; (f) where applicable, information allowing the healthcare professional to verify if the device is suitable and select the corresponding software and accessories; (g) any residual risks, contra-indications and any undesirable side-effects, including information to be conveyed to the patient in this regard; (h) specifications the user requires to use the device appropriately, e.g. if the device has a measuring function, the degree of accuracy claimed for it; (i) details of any preparatory treatment or handling of the device before it is ready for use or during its use, such as sterilisation, final assembly, calibration, etc., including the levels of disinfection required to ensure patient safety and all available methods for achieving those levels of disinfection; (j) any requirements for special facilities, or special training, or particular qualifications of the device user and/or other persons; (k) the information needed to verify whether the device is properly installed and is ready to perform safely and as intended by the manufacturer, together with, where relevant: details of the nature, and frequency, of preventive and regular maintenance, and of any preparatory cleaning or disinfection, identification of any consumable components and how to replace them, information on any necessary calibration to ensure that the device operates properly and safely during its intended lifetime, and methods for eliminating the risks encountered by persons involved in installing, calibrating or servicing devices;
(l) if the device is supplied sterile, instructions in the event of the sterile packaging being damaged or unintentionally opened before use; (m) if the device is supplied non-sterile with the intention that it is sterilised before use, the appropriate instructions for sterilisation; (n) if the device is reusable, information on the appropriate processes for allowing reuse, including cleaning, disinfection, packaging and, where appropriate, the validated method of re-sterilisation appropriate to the Member State or Member States in which the device has been placed on the market. Information shall be provided to identify when the device should no longer be reused, e.g. signs of material degradation or the maximum number of allowable reuses; (o) an indication, if appropriate, that a device can be reused only if it is reconditioned under the responsibility of the manufacturer to comply with the general safety and performance requirements; (p) if the device bears an indication that it is for single use, information on known characteristics and technical factors known to the manufacturer that could pose a risk if the device were to be re-used. This information shall be based on a specific section of the manufacturer's risk management documentation, where such characteristics and technical factors shall be addressed in detail. If in accordance with point (d) of Section 23.1. no instructions for use are required, this information shall be made available to the user upon request; (q) for devices intended for use together with other devices and/or general purpose equipment: information to identify such devices or equipment, in order to obtain a safe combination, and/or information on any known restrictions to combinations of devices and equipment;
(r) if the device emits radiation for medical purposes: detailed information as to the nature, type and where appropriate, the intensity and distribution of the emitted radiation, the means of protecting the patient, user, or other person from unintended radiation during use of the device;
(s) information that allows the user and/or patient to be informed of any warnings, precautions, contra-indications, measures to be taken and limitations of use regarding the device. That information shall, where relevant, allow the user to brief the patient about any warnings, precautions, contra-indications, measures to be taken and limitations of use regarding the device. The information shall cover, where appropriate: warnings, precautions and/or measures to be taken in the event of malfunction of the device or changes in its performance that may affect safety, warnings, precautions and/or measures to be taken as regards the exposure to reasonably foreseeable external influences or environmental conditions, such as magnetic fields, external electrical and electromagnetic effects, electrostatic discharge, radiation associated with diagnostic or therapeutic procedures, pressure, humidity, or temperature, warnings, precautions and/or measures to be taken as regards the risks of interference posed by the reasonably foreseeable presence of the device during specific diagnostic investigations, evaluations, or therapeutic treatment or other procedures such as electromagnetic interference emitted by the device affecting other equipment, if the device is intended to administer medicinal products, tissues or cells of human or animal origin, or their derivatives, or biological substances, any limitations or incompatibility in the choice of substances to be delivered, warnings, precautions and/or limitations related to the medicinal substance or biological material that is incorporated into the device as an integral part of the device; and precautions related to materials incorporated into the device that contain or consist of CMR substances or endocrine-disrupting substances, or that could result in sensitisation or an allergic reaction by the patient or user;
(t) in the case of devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body and that are absorbed by or locally dispersed in the human body, warnings and precautions, where appropriate, related to the general profile of interaction of the device and its products of metabolism with other devices, medicinal products and other substances as well as contra-indications, undesirable side-effects and risks relating to overdose; (u) in the case of implantable devices, the overall qualitative and quantitative information on the materials and substances to which patients can be exposed; (v) warnings or precautions to be taken in order to facilitate the safe disposal of the device, its accessories and the consumables used with it, if any. This information shall cover, where appropriate: infection or microbial hazards such as explants, needles or surgical equipment contaminated with potentially infectious substances of human origin, and physical hazards such as from sharps.
If in accordance with the point (d) of Section 23.1 no instructions for use are required, this information shall be made available to the user upon request; (w) for devices intended for use by lay persons, the circumstances in which the user should consult a healthcare professional; (x) for the devices covered by this Regulation pursuant to Article 1(2), information regarding the absence of a clinical benefit and the risks related to use of the device; (y) date of issue of the instructions for use or, if they have been revised, date of issue and identifier of the latest revision of the instructions for use; (z) a notice to the user and/or patient that any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established; (aa) information to be supplied to the patient with an implanted device in accordance with Article 18; (ab) for devices that incorporate electronic programmable systems, including software, or software that are devices in themselves, minimum requirements concerning hardware, IT networks characteristics and IT security measures, including protection against unauthorised access, necessary to run the software as intended.
(a) product or trade name and a general description of the device including its intended purpose and intended users; (b) the Basic UDI-DI as referred to in Part C of Annex VI assigned by the manufacturer to the device in question, as soon as identification of this device becomes based on a UDI system, or otherwise a clear identification by means of product code, catalogue number or other unambiguous reference allowing traceability; (c) the intended patient population and medical conditions to be diagnosed, treated and/or monitored and other considerations such as patient selection criteria, indications, contra-indications, warnings; (d) principles of operation of the device and its mode of action, scientifically demonstrated if necessary; (e) the rationale for the qualification of the product as a device; (f) the risk class of the device and the justification for the classification rule(s) applied in accordance with Annex VIII; (g) an explanation of any novel features; (h) a description of the accessories for a device, other devices and other products that are not devices, which are intended to be used in combination with it; (i) a description or complete list of the various configurations/variants of the device that are intended to be made available on the market; (j) a general description of the key functional elements, e.g. its parts/components (including software if appropriate), its formulation, its composition, its functionality and, where relevant, its qualitative and quantitative composition. Where appropriate, this shall include labelled pictorial representations (e.g. diagrams, photographs, and drawings), clearly indicating key parts/components, including sufficient explanation to understand the drawings and diagrams; (k) a description of the raw materials incorporated into key functional elements and those making either direct contact with the human body or indirect contact with the body, e.g., during extracorporeal circulation of body fluids; (l) technical specifications, such as features, dimensions and performance attributes, of the device and any variants/configurations and accessories that would typically appear in the product specification made available to the user, for example in brochures, catalogues and similar publications.
(a) an overview of the previous generation or generations of the device produced by the manufacturer, where such devices exist; (b) an overview of identified similar devices available on the Union or international markets, where such devices exist.
the label or labels on the device and on its packaging, such as single unit packaging, sales packaging, transport packaging in case of specific management conditions, in the languages accepted in the Member States where the device is envisaged to be sold; and the instructions for use in the languages accepted in the Member States where the device is envisaged to be sold.
(a) information to allow the design stages applied to the device to be understood; (b) complete information and specifications, including the manufacturing processes and their validation, their adjuvants, the continuous monitoring and the final product testing. Data shall be fully included in the technical documentation; (c) identification of all sites, including suppliers and sub-contractors, where design and manufacturing activities are performed.
(a) the general safety and performance requirements that apply to the device and an explanation as to why others do not apply; (b) the method or methods used to demonstrate conformity with each applicable general safety and performance requirement; (c) the harmonised standards, CS or other solutions applied; and (d) the precise identity of the controlled documents offering evidence of conformity with each harmonised standard, CS or other method applied to demonstrate conformity with the general safety and performance requirements. The information referred to under this point shall incorporate a cross-reference to the location of such evidence within the full technical documentation and, if applicable, the summary technical documentation.
(a) the benefit-risk analysis referred to in Sections 1 and 8 of Annex I, and (b) the solutions adopted and the results of the risk management referred to in Section 3 of Annex I.
(a) results of tests, such as engineering, laboratory, simulated use and animal tests, and evaluation of published literature applicable to the device, taking into account its intended purpose, or to similar devices, regarding the pre-clinical safety of the device and its conformity with the specifications; (b) detailed information regarding test design, complete test or study protocols, methods of data analysis, in addition to data summaries and test conclusions regarding in particular: the biocompatibility of the device including the identification of all materials in direct or indirect contact with the patient or user; physical, chemical and microbiological characterisation; electrical safety and electromagnetic compatibility; software verification and validation (describing the software design and development process and evidence of the validation of the software, as used in the finished device. This information shall typically include the summary results of all verification, validation and testing performed both in-house and in a simulated or actual user environment prior to final release. It shall also address all of the different hardware configurations and, where applicable, operating systems identified in the information supplied by the manufacturer); stability, including shelf life; and performance and safety.
Where applicable, conformity with the provisions of Directive 2004/10/EC of the European Parliament and of the Council shall be demonstrated.Directive 2004/10/EC of the European Parliament and of the Council of 11 February 2004 on the harmonisation of laws, regulations and administrative provisions relating to the application of the principles of good laboratory practice and the verification of their applications for tests on chemical substances (OJ L 50, 20.2.2004, p. 44 ).Where no new testing has been undertaken, the documentation shall incorporate a rationale for that decision. An example of such a rationale would be that biocompatibility testing on identical materials was conducted when those materials were incorporated in a previous version of the device that has been legally placed on the market or put into service; (c) the clinical evaluation report and its updates and the clinical evaluation plan referred to in Article 61(12) and Part A of Annex XIV; (d) the PMCF plan and PMCF evaluation report referred to in Part B of Annex XIV or a justification why a PMCF is not applicable.
(a) Where a device incorporates, as an integral part, a substance which, if used separately, may be considered to be a medicinal product within the meaning of point 2 of Article 1 of Directive 2001/83/EC, including a medicinal product derived from human blood or human plasma, as referred to in the first subparagraph of Article 1(8), a statement indicating this fact. In this case, the documentation shall identify the source of that substance and contain the data of the tests conducted to assess its safety, quality and usefulness, taking account of the intended purpose of the device. (b) Where a device is manufactured utilising tissues or cells of human or animal origin, or their derivatives, and is covered by this Regulation in accordance with points (f) and (g) of Article 1(6, and where a device incorporates, as an integral part, tissues or cells of human origin or their derivatives that have an action ancillary to that of the device and is covered by this Regulation in accordance with the first subparagraph of Article 1(10), a statement indicating this fact. In such a case, the documentation shall identify all materials of human or animal origin used and provide detailed information concerning the conformity with Sections 13.1. or 13.2., respectively, of Annex I. (c) In the case of devices that are composed of substances or combinations of substances that are intended to be introduced into the human body and that are absorbed by or locally dispersed in the human body, detailed information, including test design, complete test or study protocols, methods of data analysis, and data summaries and test conclusions, regarding studies in relation to: absorption, distribution, metabolism and excretion; possible interactions of those substances, or of their products of metabolism in the human body, with other devices, medicinal products or other substances, considering the target population, and its associated medical conditions; local tolerance; and toxicity, including single-dose toxicity, repeat-dose toxicity, genotoxicity, carcinogenicity and reproductive and developmental toxicity, as applicable depending on the level and nature of exposure to the device.
In the absence of such studies, a justification shall be provided. (d) In the case of devices containing CMR or endocrine-disrupting substances referred to in Section 10.4.1 of Annex I, the justification referred to in Section 10.4.2 of that Annex. (e) In the case of devices placed on the market in a sterile or defined microbiological condition, a description of the environmental conditions for the relevant manufacturing steps. In the case of devices placed on the market in a sterile condition, a description of the methods used, including the validation reports, with respect to packaging, sterilisation and maintenance of sterility. The validation report shall address bioburden testing, pyrogen testing and, if applicable, testing for sterilant residues. (f) In the case of devices placed on the market with a measuring function, a description of the methods used in order to ensure the accuracy as given in the specifications. (g) If the device is to be connected to other device(s) in order to operate as intended, a description of this combination/configuration including proof that it conforms to the general safety and performance requirements when connected to any such device(s) having regard to the characteristics specified by the manufacturer.
(a) The post-market surveillance plan shall address the collection and utilization of available information, in particular: information concerning serious incidents, including information from PSURs, and field safety corrective actions; records referring to non-serious incidents and data on any undesirable side-effects; information from trend reporting; relevant specialist or technical literature, databases and/or registers; information, including feedbacks and complaints, provided by users, distributors and importers; and publicly available information about similar medical devices.
(b) The post-market surveillance plan shall cover at least: a proactive and systematic process to collect any information referred to in point (a). The process shall allow a correct characterisation of the performance of the devices and shall also allow a comparison to be made between the device and similar products available on the market; effective and appropriate methods and processes to assess the collected data; suitable indicators and threshold values that shall be used in the continuous reassessment of the benefit-risk analysis and of the risk management as referred to in Section 3 of Annex I; effective and appropriate methods and tools to investigate complaints and analyse market-related experience collected in the field; methods and protocols to manage the incidents subject to the trend report as provided for in Article 88, including the methods and protocols to be used to establish any statistically significant increase in the frequency or severity of incidents as well as the observation period; methods and protocols to communicate effectively with competent authorities, notified bodies, economic operators and users; reference to procedures to fulfil the manufacturers obligations laid down in Articles 83, 84 and 86; systematic procedures to identify and initiate appropriate measures including corrective actions; effective tools to trace and identify devices for which corrective actions might be necessary; and a PMCF plan as referred to in Part B of Annex XIV, or a justification as to why a PMCF is not applicable.
1. Name, registered trade name or registered trade mark and, if already issued, SRN as referred to in Article 31 of the manufacturer, and, if applicable, its authorised representative, and the address of their registered place of business where they can be contacted and their location be established; 2. A statement that the EU declaration of conformity is issued under the sole responsibility of the manufacturer; 3. The Basic UDI-DI as referred to in Part C of Annex VI; 4. Product and trade name, product code, catalogue number or other unambiguous reference allowing identification and traceability of the device covered by the EU declaration of conformity, such as a photograph, where appropriate, as well as its intended purpose. Except for the product or trade name, the information allowing identification and traceability may be provided by the Basic UDI-DI referred to in point 3; 5. Risk class of the device in accordance with the rules set out in Annex VIII; 6. A statement that the device that is covered by the present declaration is in conformity with this Regulation and, if applicable, with any other relevant Union legislation that provides for the issuing of an EU declaration of conformity; 7. References to any CS used and in relation to which conformity is declared; 8. Where applicable, the name and identification number of the notified body, a description of the conformity assessment procedure performed and identification of the certificate or certificates issued; 9. Where applicable, additional information; 10. Place and date of issue of the declaration, name and function of the person who signed it as well as an indication for, and on behalf of whom, that person signed, signature.
1.1. type of economic operator(manufacturer, authorised representative, or importer), 1.2. name, address and contact details of the economic operator, 1.3. where submission of information is carried out by another person on behalf of any of the economic operators mentioned under Section 1.1, the name, address and contact details of that person, 1.4. name address and contact details of the person or persons responsible for regulatory compliance referred to in Article 15.
2.1. Basic UDI-DI, 2.2. type, number and expiry date of the certificate issued by the notified body and the name or identification number of that notified body and the link to the information that appears on the certificate and was entered by the notified body in the electronic system on notified bodies and certificates, 2.3. Member State in which the device is to or has been placed on the market in the Union, 2.4. in the case of class IIa, class IIb or class III devices: Member States where the device is or is to be made available, 2.5. risk class of the device, 2.6. reprocessed single-use device (y/n), 2.7. presence of a substance which, if used separately, may be considered to be a medicinal product and name of that substance, 2.8. presence of a substance which, if used separately, may be considered to be a medicinal product derived from human blood or human plasma and name of this substance, 2.9. presence of tissues or cells of human origin, or their derivatives (y/n), 2.10. presence of tissues or cells of animal origin, or their derivatives, as referred to in Regulation (EU) No 722/2012 (y/n), 2.11. where applicable, the single identification number of the clinical investigation or investigations conducted in relation to the device or a link to the clinical investigation registration in the electronic system on clinical investigations, 2.12. in the case of devices listed in Annex XVI, specification as to whether the intended purpose of the device is other than a medical purpose, 2.13. in the case of devices designed and manufactured by another legal or natural person as referred in Article 10(15), the name, address and contact details of that legal or natural person, 2.14. in the case of class III or implantable devices, the summary of safety and clinical performance, 2.15. status of the device (on the market, no longer placed on the market, recalled, field safety corrective action initiated).
1. quantity per package configuration, 2. the Basic UDI-DI as referred to in Article 29 and any additional UDI-DIs, 3. the manner in which production of the device is controlled (expiry date or manufacturing date, lot number, serial number), 4. if applicable, the unit of use UDI-DI (where a UDI is not labelled on the device at the level of its unit of use, a "unit of use" DI shall be assigned so as to associate the use of a device with a patient), 5. name and address of the manufacturer (as indicated on the label), 6. the SRN issued in accordance with Article 31(2), 7. if applicable, name and address of the authorised representative (as indicated on the label), 8. the medical device nomenclature code as provided for in Article 26, 9. risk class of the device, 10. if applicable, name or trade name, 11. if applicable, device model, reference, or catalogue number, 12. if applicable, clinical size (including volume, length, gauge, diameter), 13. additional product description (optional), 14. if applicable, storage and/or handling conditions (as indicated on the label or in the instructions for use), 15. if applicable, additional trade names of the device, 16. labelled as a single-use device (y/n), 17. if applicable, the maximum number of reuses, 18. device labelled sterile (y/n), 19. need for sterilisation before use (y/n), 20. containing latex (y/n), 21. where applicable, information labelled in accordance with Section 10.4.5 of Annex I, 22. URL for additional information, such as electronic instructions for use (optional), 23. if applicable, critical warnings or contra-indications, 24. status of the device (on the market, no longer placed on the market, recalled, field safety corrective action initiated).
gantries, tubes, tables, consoles and other items of equipment that can be configured/combined to deliver an intended function in computed tomography. ventilators, breathing circuits, vaporizers combined to deliver an intended function in anaesthesia.
(a) name or trade name, (b) device version or model, (c) labelled as single use, (d) packaged sterile, (e) need for sterilization before use, (f) quantity of devices provided in a package, (g) critical warnings or contra-indications: e.g. containing latex or DEHP.
(a) any type of direct marking would interfere with the safety or performance of the device; (b) the device cannot be directly marked because it is not technologically feasible.
6.1.1. Implantable devices shall, at their lowest level of packaging ("unit packs"), be identified, or marked using AIDC, with a UDI (UDI-DI + UDI-PI); 6.1.2. The UDI-PI shall have at least the following characteristics: (a) the serial number for active implantable devices, (b) the serial number or lot number for other implantable devices.
6.1.3. The UDI of the implantable device shall be identifiable prior to implantation.
(a) individual single-use disposable devices, the uses of which are generally known to the persons by whom they are intended to be used, which are contained within a system or procedure pack, and which are not intended for individual use outside the context of the system or procedure pack, shall not be required to bear their own UDI carrier; (b) devices that are exempted from bearing a UDI carrier on the relevant level of packaging shall not be required to bear a UDI carrier when included within a system or procedure pack.
(a) The system or procedure pack UDI carrier shall as a general rule be affixed to the outside of the packaging. (b) The UDI carrier shall be readable, or, in the case of AIDC, scannable, whether placed on the outside of the packaging of the system or procedure pack or inside transparent packaging.
(a) the original performance; (b) the safety or the intended use of the software; (c) interpretation of data.
(a) where the software is delivered on a physical medium, e.g. CD or DVD, each packaging level shall bear the human readable and AIDC representation of the complete UDI. The UDI that is applied to the physical medium containing the software and its packaging shall be identical to the UDI assigned to the system level software; (b) the UDI shall be provided on a readily accessible screen for the user in an easily-readable plain-text format, such as an "about" file, or included on the start-up screen; (c) software lacking a user interface such as middleware for image conversion, shall be capable of transmitting the UDI through an application programming interface (API); (d) only the human readable portion of the UDI shall be required in electronic displays of the software. The marking of UDI using AIDC shall not be required in the electronic displays, such as "about" menu, splash screen etc.; (e) the human readable format of the UDI for the software shall include the Application Identifiers (AI) for the standard used by the issuing entities, so as to assist the user in identifying the UDI and determining which standard is being used to create the UDI.
the provision of adequate resources for conformity assessment activities; the development of procedures and policies for the operation of the notified body; the supervision of implementation of the procedures, policies and quality management systems of the notified body; the supervision of the notified body's finances; the activities and decisions taken by the notified body, including contractual agreements; the delegation of authority to personnel and/or committees, where necessary, for the performance of defined activities; the interaction with the authority responsible for notified bodies and the obligations regarding communications with other competent authorities, the Commission and other notified bodies.
(a) be the designer, manufacturer, supplier, installer, purchaser, owner or maintainer of devices which they assess, nor the authorised representative of any of those parties. Such restriction shall not preclude the purchase and use of assessed devices that are necessary for the operations of the notified body and the conduct of the conformity assessment, or the use of such devices for personal purposes; (b) be involved in the design, manufacture or construction, marketing, installation and use, or maintenance of the devices for which they are designated, nor represent the parties engaged in those activities; (c) engage in any activity that may conflict with their independence of judgement or integrity in relation to conformity assessment activities for which they are designated; (d) offer or provide any service which may jeopardise the confidence in their independence, impartiality or objectivity. In particular, they shall not offer or provide consultancy services to the manufacturer, its authorised representative, a supplier or a commercial competitor as regards the design, construction, marketing or maintenance of devices or processes under assessment, and (e) be linked to any organisation which itself provides consultancy services as referred to in point (d). Such restriction does not preclude general training activities that are not client specific and that relate to regulation of devices or to related standards.
management system structure and documentation, including policies and objectives for its activities; policies for assignment of activities and responsibilities to personnel; assessment and decision-making processes in accordance with the tasks, responsibilities and role of the notified body's personnel and top-level management; the planning, conduct, evaluation and, if necessary, adaptation of its conformity assessment procedures; control of documents; control of records; management reviews; internal audits; corrective and preventive actions; complaints and appeals; and continuous training.
the pre-clinical evaluation, clinical evaluation, tissues and cells of human and animal origin, functional safety, software, packaging, devices that incorporate as an integral part a medicinal product, devices that are composed of substances or of combinations of substances that are absorbed by or locally dispersed in the human body and the different types of sterilisation processes.
Union devices legislation and relevant guidance documents; the conformity assessment procedures provided for in this Regulation; a broad base of knowledge of device technologies and the design and manufacture of devices; the notified body's quality management system, related procedures and the required qualification criteria; training relevant to personnel involved in conformity assessment activities in relation to devices; adequate experience in conformity assessments under this Regulation or previously applicable law within a notified body.
identify when specialist input is required for the assessment of the clinical evaluation conducted by the manufacturer and identify appropriately qualified experts; appropriately train external clinical experts in the relevant requirements of this Regulation, CS, guidance and harmonised standards and ensure that the external clinical experts are fully aware of the context and implications of their assessment and the advice they provide; be able to review and scientifically challenge the clinical data contained within the clinical evaluation, and any associated clinical investigations, and appropriately guide external clinical experts in the assessment of the clinical evaluation presented by the manufacturer; be able to scientifically evaluate and, if necessary, challenge the clinical evaluation presented, and the results of the external clinical experts' assessment of the manufacturer's clinical evaluation; be able to ascertain the comparability and consistency of the assessments of clinical evaluations conducted by clinical experts; be able to make an assessment of the manufacturer's clinical evaluation and a clinical judgement of the opinion provided by any external expert and make a recommendation to the notified body's decision maker; and be able to draw up records and reports demonstrating that the relevant conformity assessment activities have been appropriately carried out.
successful completion of a university or a technical college degree or equivalent qualification in relevant studies, e.g. medicine, pharmacy, engineering or other relevant sciences; four years' professional experience in the field of healthcare products or related activities, such as in manufacturing, auditing or research, of which two years shall be in the design, manufacture, testing or use of the device or technology to be assessed or related to the scientific aspects to be assessed; knowledge of device legislation, including the general safety and performance requirements set out in Annex I; appropriate knowledge and experience of relevant harmonised standards, CS and guidance documents; appropriate knowledge and experience of risk management and related device standards and guidance documents; appropriate knowledge and experience of clinical evaluation; appropriate knowledge of the devices which they are assessing; appropriate knowledge and experience of the conformity assessment procedures laid down in Annexes IX to XI, in particular of the aspects of those procedures for which they are responsible, and adequate authorisation for carrying out those assessments; the ability to draw up records and reports demonstrating that the relevant conformity assessment activities have been appropriately carried out.
successful completion of a university or a technical college degree or equivalent qualification in relevant studies, such as medicine, pharmacy, engineering or other relevant sciences; four years' professional experience in the field of healthcare products or related activities, such as in manufacturing, auditing or research, of which two years shall be in the area of quality management; appropriate knowledge of devices legislation as well as related harmonised standards, CS and guidance documents; appropriate knowledge and experience of risk management and related device standards and guidance documents; appropriate knowledge of quality management systems and related standards and guidance documents; appropriate knowledge and experience of the conformity assessment procedures laid down in Annexes IX to XI, in particular of the aspects of those procedures for which they are responsible, and adequate authorisation for carrying out those audits; training in auditing techniques enabling them to challenge quality management systems; the ability to draw up records and reports demonstrating that the relevant conformity assessment activities have been appropriately carried out.
devices legislation and relevant guidance documents; the device conformity assessments relevant to this Regulation; the types of qualifications, experience and expertise relevant to device conformity assessment; a broad base of knowledge of device technologies, including sufficient experience of conformity assessment of devices being reviewed for certification, the device industry and the design and manufacture of devices; the notified body's quality management system, related procedures and the required qualifications for personnel involved; the ability to draw up records and reports demonstrating that the conformity assessment activities have been appropriately carried out.
a matrix detailing the authorisations and responsibilities of the personnel in respect of conformity assessment activities; and records attesting to the required knowledge and experience for the conformity assessment activity for which they are authorised. The records shall contain a rationale for defining the scope of the responsibilities for each of the assessment personnel and records of the conformity assessment activities carried out by each of them.
review of the qualifications and monitoring of the performance of external experts; auditing and certification activities where the subcontracting in question is to auditing or certification organisations; allocation of work to external experts for specific conformity assessment activities; and final review and decision making functions.
the subcontractor meets the relevant requirements of this Annex; subcontractors and external experts do not further subcontract work to organisations or personnel; and the natural or legal person that applied for conformity assessment has been informed of the requirements referred to in the first and second indent.
are aware of Union and national law in force on devices, relevant harmonised standards, CS, guidance documents and the results of the coordination activities referred to in Section 1.6; and take part in the internal exchange of experience and the continuous training and education programme referred to in Section 3.1.2.
(a) publish a publicly available description of the application procedure by which manufacturers can obtain certification from it. That description shall include which languages are acceptable for submission of documentation and for any related correspondence; (b) have documented procedures relating to, and documented details about, fees charged for specific conformity assessment activities and any other financial conditions relating to notified bodies' assessment activities for devices; (c) have documented procedures in relation to advertising of their conformity assessment services. Those procedures shall ensure that advertising or promotional activities in no way imply or are capable of leading to an inference that their conformity assessment will offer manufacturers earlier market access or be quicker, easier or less stringent than that of other notified bodies; (d) have documented procedures requiring the review of pre-application information, including the preliminary verification that the product is covered by this Regulation and its classification, prior to issuing any quotation to the manufacturer relating to a specific conformity assessment; and (e) ensure that all contracts relating to the conformity assessment activities covered by this Regulation are concluded directly between the manufacturer and the notified body and not with any other organisation.
(a) the completeness of those applications with respect to the requirements of the relevant conformity assessment procedure, as referred to in the corresponding Annex, under which approval has been sought, (b) the verification of the qualification of products covered by those applications as devices and their respective classifications, (c) whether the conformity assessment procedures chosen by the applicant are applicable to the device in question under this Regulation, (d) the ability of the notified body to assess the application based on its designation, and (e) the availability of sufficient and appropriate resources.
appropriately plan the conduct of each individual project, ensure that the composition of the assessment teams is such that there is sufficient experience in relation to the technology concerned, and that there is continuous objectivity and independence, and to provide for rotation of the members of the assessment team at appropriate intervals, specify the rationale for fixing time limits for completion of conformity assessment activities, assess the manufacturer's technical documentation and the solutions adopted to meet the requirements laid down in Annex I, review the manufacturer's procedures and documentation relating to the evaluation of pre-clinical aspects, review the manufacturer's procedures and documentation relating to clinical evaluation, address the interface between the manufacturer's risk management process and its appraisal and analysis of the pre-clinical and clinical evaluation and to evaluate their relevance for the demonstration of conformity with the relevant requirements in Annex I, carry out the specific procedures referred to in Sections 5.2 to 5.4 of Annex IX, in the case of class IIa or class IIb devices, assess the technical documentation of devices selected on a representative basis, plan and periodically carry out appropriate surveillance audits and assessments, carry out or request certain tests to verify the proper functioning of the quality management system and to perform unannounced on site audits, relating to the sampling of devices, verify that the manufactured device is in conformity with the technical documentation; such requirements shall define the relevant sampling criteria and testing procedure prior to sampling, evaluate and verify a manufacturer's compliance with relevant Annexes.
(a) As part of the assessment of the quality management system, a notified body shall prior to an audit and in accordance with its documented procedures: assess the documentation submitted in accordance with the relevant conformity assessment Annex, and draw up an audit programme which clearly identifies the number and sequence of activities required to demonstrate complete coverage of a manufacturer's quality management system and to determine whether it meets the requirements of this Regulation, identify links between, and allocation of responsibilities among, the various manufacturing sites, and identify relevant suppliers and/or subcontractors of the manufacturer, and consider the need to specifically audit any of those suppliers or subcontractors or both, clearly define, for each audit identified in the audit programme, the objectives, criteria and scope of the audit, and draw up an audit plan that adequately addresses and takes account of the specific requirements for the devices, technologies and processes involved, draw up and keep up to date, for class IIa and class IIb devices, a sampling plan for the assessment of technical documentation as referred to in Annexes II and III covering the range of such devices covered by the manufacturer's application. That plan shall ensure that the entire range of devices covered by the certificate is sampled over the period of validity of the certificate, and select and assign appropriately qualified and authorised personnel for conducting the individual audits. The respective roles, responsibilities and authorities of the team members shall be clearly defined and documented.
(b) Based on the audit programme it has drawn up, the notified body shall, in accordance with its documented procedures: audit the manufacturer's quality management system, in order to verify that the quality management system ensures that the devices covered conform to the relevant provisions of this Regulation which apply to devices at every stage, from design through final quality control to ongoing surveillance, and shall determine whether the requirements of this Regulation are met, based on relevant technical documentation and in order to determine whether the manufacturer meets the requirements referred to in the relevant conformity assessment Annex, review and audit the manufacturer's processes and subsystems, in particular for: design and development, production and process controls, product documentation, purchasing controls including verification of purchased devices, corrective and preventive actions, including for post-market surveillance, and PMCF,
and review and audit requirements and provisions adopted by the manufacturer, including those in relation to fulfilling the general safety and performance requirements set out in Annex I. The documentation shall be sampled in such a manner as to reflect the risks associated with the intended use of the device, the complexity of the manufacturing technologies, the range and classes of devices produced and any available post-market surveillance information, if not already covered by the audit programme, audit the control of processes on the premises of the manufacturer's suppliers, when the conformity of finished devices is significantly influenced by the activity of suppliers and, in particular when the manufacturer cannot demonstrate sufficient control over its suppliers, conduct assessments of the technical documentation based on its sampling plan and taking account of Sections 4.5.4. and 4.5.5. for pre-clinical and clinical evaluations, and the notified body shall ensure that audit findings are appropriately and consistently classified in accordance with the requirements of this Regulation and with relevant standards, or with best practice documents developed or adopted by the MDCG.
the allocation of appropriately qualified and authorised personnel for the examination of individual aspects such as use of the device, biocompatibility, clinical evaluation, risk management, and sterilisation, and the assessment of conformity of the design with this Regulation, and for taking account of Sections 4.5.4. to 4.5.6. That assessment shall include examination of the implementation by manufacturers of incoming, in-process and final checks and the results thereof. If further tests or other evidence is required for the assessment of conformity with the requirements of this Regulation, the notified body in question shall carry out adequate physical or laboratory tests in relation to the device or request the manufacturer to carry out such tests.
examine and assess the technical documentation taking account of Sections 4.5.4. to 4.5.6., and verify that the type has been manufactured in conformity with that documentation; establish a test plan identifying all relevant and critical parameters which need to be tested by the notified body or under its responsibility; document its rationale for the selection of those parameters; carry out the appropriate examinations and tests in order to verify that the solutions adopted by the manufacturer meet the general safety and performance requirements set out in Annex I. Such examinations and tests shall include all tests necessary to verify that the manufacturer has in fact applied the relevant standards it has opted to use; agree with the applicant as to where the necessary tests will be performed if they are not to be carried out directly by the notified body; and assume full responsibility for test results. Test reports submitted by the manufacturer shall only be taken into account if they have been issued by conformity assessment bodies which are competent and independent of the manufacturer.
(a) have documented procedures, sufficient expertise and facilities for the verification by examination and testing of every product in accordance with Part B of Annex XI; (b) establish a test plan identifying all relevant and critical parameters which need to be tested by the notified body or under its responsibility in order to: verify, for class IIb devices, the conformity of the device with the type described in the EU type-examination certificate and with the requirements of this Regulation which apply to those devices, confirm, for class IIa devices, the conformity with the technical documentation referred to in Annexes II and III and with the requirements of this Regulation which apply to those devices;
(c) document its rationale for the selection of the parameters referred to in point (b); (d) have documented procedures to carry out the appropriate assessments and tests in order to verify the conformity of the device with the requirements of this Regulation by examining and testing every product as specified in Section 15 of Annex XI; (e) have documented procedures providing for the reaching of an agreement with the applicant concerning when and where necessary tests that are not to be carried out by the notified body itself are to be performed; and (f) assume full responsibility for test results in accordance with documented procedures; test reports submitted by the manufacturer shall only be taken into account if they have been issued by conformity assessment bodies which are competent and independent of the manufacturer.
(a) the planning, conduct, assessment, reporting and, where appropriate, updating of the pre-clinical evaluation, in particular of the scientific pre-clinical literature search, and the pre-clinical testing, for example laboratory testing, simulated use testing, computer modelling, the use of animal models,
(b) the nature and duration of body contact and the specific associated biological risks, (c) the interface with the risk management process, and (d) the appraisal and analysis of the available pre-clinical data and its relevance with regard to demonstrating conformity with the relevant requirements in Annex I.
the planning, conduct, assessment, reporting and updating of the clinical evaluation as referred to in Annex XIV, post-market surveillance and PMCF, the interface with the risk management process, the appraisal and analysis of the available data and its relevance with regard to demonstrating conformity with the relevant requirements in Annex I, and the conclusions drawn with regard to the clinical evidence and drawing up of the clinical evaluation report.
the intended use specified by the manufacturer and claims for the device defined by it, the planning of the clinical evaluation, the methodology for the literature search, relevant documentation from the literature search, the clinical investigation, validity of equivalence claimed in relation to other devices, the demonstration of equivalence, the suitability and conclusions data from equivalent and similar devices, post-market surveillance and PMCF, the clinical evaluation report, and justifications in relation to non-performance of clinical investigations or PMCF.
ensure that all steps of the conformity assessment are documented so that the conclusions of the assessment are clear and demonstrate compliance with the requirements of this Regulation and can represent objective evidence of such compliance to persons that are not themselves involved in the assessment, for example personnel in designating authorities, ensure that records that are sufficient to provide a discernible audit trail are available for quality management system audits, clearly document the conclusions of its assessment of clinical evaluation in a clinical evaluation assessment report, and for each specific project, provide a detailed report which shall be based on a standard format containing a minimum set of elements determined by the MDCG.
clearly document the outcome of its assessment and draw clear conclusions from the verification of the manufacturer's conformity with the requirements of this Regulation, make a recommendation for a final review and for a final decision to be taken by the notified body; this recommendation shall be signed off by the member of personnel responsible in the notified body, and be provided to the manufacturer in question.
ensure that the personnel assigned for the final review and decision-making on specific projects are appropriately authorised and are different from the personnel who have conducted the assessments, verify that the report or reports and supporting documentation needed for decision making, including concerning resolution of non-conformities noted during assessment, are complete and sufficient with respect to the scope of the application, and verify whether there are any unresolved non-conformities preventing issuance of a certificate.
decide, based on the assessment documentation and additional information available, whether the requirements of this Regulation are fulfilled, decide, based on the results of its assessment of the clinical evaluation and risk management, whether the post-market surveillance plan, including the PMCF plan, is adequate, decide on specific milestones for further review by the notified body of the up to date clinical evaluation, decide whether specific conditions or provisions need to be defined for the certification, decide, based on the novelty, risk classification, clinical evaluation and conclusions from the risk analysis of the device, on a period of certification not exceeding five years, clearly document decision making and approval steps including approval by signature of the members of personnel responsible, clearly document responsibilities and mechanisms for communication of decisions, in particular, where the final signatory of a certificate differs from the decision maker or decision makers or does not fulfil the requirements laid down in Section 3.2.7, issue a certificate or certificates in accordance with the minimum requirements laid down in Annex XII for a period of validity not exceeding five years and shall indicate whether there are specific conditions or limitations associated with the certification, issue a certificate or certificates for the applicant alone and shall not issue certificates covering multiple entities, and ensure that the manufacturer is notified of the outcome of the assessment and the resultant decision and that they are entered into the electronic system referred to in Article 57.
the approved quality management system or systems or to the product-range covered, the approved design of a device, the intended use of or claims made for the device, the approved type of a device, and any substance incorporated in or utilised for the manufacturing of a device and being subject to the specific procedures in accordance with Section 4.5.6.
ensure that manufacturers submit for prior approval plans for changes as referred to in the first paragraph and relevant information relating to such changes, assess the changes proposed and verify whether, after these changes, the quality management system, or the design of a device or type of a device, still meets the requirements of this Regulation, and notify the manufacturer of its decision and provide a report or as applicable a supplementary report, which shall contain the justified conclusions of its assessment.
defining how and when surveillance activities of manufacturers are to be conducted. Those procedures shall include arrangements for unannounced on-site audits of manufacturers and, where applicable, subcontractors and suppliers carrying out product tests and the monitoring of compliance with any conditions binding manufacturers and associated with certification decisions, such as updates to clinical data at defined intervals, for screening relevant sources of scientific and clinical data and post-market information relating to the scope of their designation. Such information shall be taken into account in the planning and conduct of surveillance activities, and to review vigilance data to which they have access under Article 92(2) in order to estimate its impact, if any, on the validity of existing certificates. The results of the evaluation and any decisions taken shall be thoroughly documented.
not to take action on the basis that the vigilance case is clearly not related to the certification granted, observe the manufacturer's and competent authority's activities and the results of the manufacturer's investigation so as to determine whether the certification granted is at risk or whether adequate corrective action has been taken, perform extraordinary surveillance measures, such as document reviews, short-notice or unannounced audits and product testing, where it is likely that the certification granted is at risk, increase the frequency of surveillance audits, review specific products or processes on the occasion of the next audit of the manufacturer, or take any other relevant measure.
conduct surveillance audits of the manufacturer on at least an annual basis which shall be planned and conducted in line with the relevant requirements in Section 4.5, ensure adequate assessment of the manufacturer's documentation on, and application of the provisions on, vigilance, the post-market surveillance, and PMCF, sample and test devices and technical documentation, during audits, according to pre-defined sampling criteria and testing procedures to ensure that the manufacturer continuously applies the approved quality management system, ensure that the manufacturer complies with the documentation and information obligations laid down in the relevant Annexes and that its procedures take into account best practices in the implementation of quality management systems, ensure that the manufacturer does not use quality management system or device approvals in a misleading manner, gather sufficient information to determine if the quality management system continues to comply with the requirements of this Regulation, ask the manufacturer, if non-conformities are detected, for corrections, corrective actions and, where applicable, preventive actions, and where necessary, impose specific restrictions on the relevant certificate, or suspend or withdraw it.
conduct an in-depth review of the clinical evaluation as most recently updated by the manufacturer based on the manufacturer's post-market surveillance, on its PMCF and on clinical literature relevant to the condition being treated with the device or on clinical literature relevant to similar devices, clearly document the outcome of the in-depth review and address any specific concerns to the manufacturer or impose any specific conditions on it, and ensure that the clinical evaluation as most recently updated, is appropriately reflected in the instructions for use and, where applicable, the summary of safety and performance.
(a) all changes to the originally approved device, including changes not yet notified, (b) experience gained from post-market surveillance, (c) experience from risk management, (d) experience from updating the proof of compliance with the general safety and performance requirements set out in Annex I, (e) experience from reviews of the clinical evaluation, including the results of any clinical investigations and PMCF, (f) changes to the requirements, to components of the device or to the scientific or regulatory environment, (g) changes to applied or new harmonised standards, CS or equivalent documents, and (h) changes in medical, scientific and technical knowledge, such as: new treatments, changes in test methods, new scientific findings on materials and components, including findings on their biocompatibility, experience from studies on comparable devices, data from registers and registries, experience from clinical investigations with comparable devices.
(a) an invasive device which penetrates inside the body through the surface of the body, including through mucous membranes of body orifices with the aid or in the context of a surgical operation; and (b) a device which produces penetration other than through a body orifice.
(a) the entire duration of use of the same device without regard to temporary interruption of use during a procedure or temporary removal for purposes such as cleaning or disinfection of the device. Whether the interruption of use or the removal is temporary shall be established in relation to the duration of the use prior to and after the period when the use is interrupted or the device removed; and (b) the accumulated use of a device that is intended by the manufacturer to be replaced immediately with another of the same type.
if they may be connected to a class IIa, class IIb or class III active device; or if they are intended for use for channelling or storing blood or other body liquids or for storing organs, parts of organs or body cells and tissues, except for blood bags; blood bags are classified as class IIb.
class I if they are intended to be used as a mechanical barrier, for compression or for absorption of exudates; class IIb if they are intended to be used principally for injuries to skin which have breached the dermis or mucous membrane and can only heal by secondary intent; class IIa if they are principally intended to manage the micro-environment of injured skin or mucous membrane; and class IIa in all other cases.
class I if they are intended for transient use; class IIa if they are intended for short-term use, except if they are used in the oral cavity as far as the pharynx, in an ear canal up to the ear drum or in the nasal cavity, in which case they are classified as class I; and class IIb if they are intended for long-term use, except if they are used in the oral cavity as far as the pharynx, in an ear canal up to the ear drum or in the nasal cavity and are not liable to be absorbed by the mucous membrane, in which case they are classified as class IIa.
are intended specifically to control, diagnose, monitor or correct a defect of the heart or of the central circulatory system through direct contact with those parts of the body, in which case they are classified as class III; are reusable surgical instruments, in which case they are classified as class I; are intended specifically for use in direct contact with the heart or central circulatory system or the central nervous system, in which case they are classified as class III; are intended to supply energy in the form of ionising radiation in which case they are classified as class IIb; have a biological effect or are wholly or mainly absorbed in which case they are classified as class IIb; or are intended to administer medicinal products by means of a delivery system, if such administration of a medicinal product is done in a manner that is potentially hazardous taking account of the mode of application, in which case they are classified as class IIb.
are intended specifically to control, diagnose, monitor or correct a defect of the heart or of the central circulatory system through direct contact with those parts of the body, in which case they are classified as class III; are intended specifically for use in direct contact with the heart or central circulatory system or the central nervous system, in which case they are classified as class III; are intended to supply energy in the form of ionizing radiation in which case they are classified as class IIb; have a biological effect or are wholly or mainly absorbed in which case they are classified as class III; are intended to undergo chemical change in the body in which case they are classified as class IIb, except if the devices are placed in the teeth; or are intended to administer medicines, in which case they are classified as class IIb.
are intended to be placed in the teeth, in which case they are classified as class IIa; are intended to be used in direct contact with the heart, the central circulatory system or the central nervous system, in which case they are classified as class III; have a biological effect or are wholly or mainly absorbed, in which case they are classified as class III; are intended to undergo chemical change in the body in which case they are classified as class III, except if the devices are placed in the teeth; are intended to administer medicinal products, in which case they are classified as class III; are active implantable devices or their accessories, in which cases they are classified as class III; are breast implants or surgical meshes, in which cases they are classified as class III; are total or partial joint replacements, in which case they are classified as class III, with the exception of ancillary components such as screws, wedges, plates and instruments; or are spinal disc replacement implants or are implantable devices that come into contact with the spinal column, in which case they are classified as class III with the exception of components such as screws, wedges, plates and instruments.
if they are intended to supply energy which will be absorbed by the human body, except for devices intended to illuminate the patient's body, in the visible spectrum, in which case they are classified as class I; if they are intended to image in vivo distribution of radiopharmaceuticals; orif they are intended to allow direct diagnosis or monitoring of vital physiological processes, unless they are specifically intended for monitoring of vital physiological parameters and the nature of variations of those parameters is such that it could result in immediate danger to the patient, for instance variations in cardiac performance, respiration, activity of the central nervous system, or they are intended for diagnosis in clinical situations where the patient is in immediate danger, in which cases they are classified as class IIb.
death or an irreversible deterioration of a person's state of health, in which case it is in class III; or a serious deterioration of a person's state of health or a surgical intervention, in which case it is classified as class IIb.
class III if they present a high or medium potential for internal exposure; class IIb if they present a low potential for internal exposure; and class IIa if they present a negligible potential for internal exposure.
class III if they, or their products of metabolism, are systemically absorbed by the human body in order to achieve the intended purpose; class III if they achieve their intended purpose in the stomach or lower gastrointestinal tract and they, or their products of metabolism, are systemically absorbed by the human body; class IIa if they are applied to the skin or if they are applied in the nasal or oral cavity as far as the pharynx, and achieve their intended purpose on those cavities; and class IIb in all other cases.
the name of the manufacturer and address of its registered place of business and any additional manufacturing site covered by the quality management system, and, if the manufacturer's application is lodged by its authorised representative, the name of the authorised representative and the address of the authorised representative's registered place of business, all relevant information on the device or group of devices covered by the quality management system, a written declaration that no application has been lodged with any other notified body for the same device-related quality management system, or information about any previous application for the same device-related quality management system, a draft of an EU declaration of conformity in accordance with Article 19 and Annex IV for the device model covered by the conformity assessment procedure, the documentation on the manufacturer's quality management system, a documented description of the procedures in place to fulfil the obligations arising from the quality management system and required under this Regulation and the undertaking by the manufacturer in question to apply those procedures, a description of the procedures in place to ensure that the quality management system remains adequate and effective, and the undertaking by the manufacturer to apply those procedures, the documentation on the manufacturer's post-market surveillance system and, where applicable, on the PMCF plan, and the procedures put in place to ensure compliance with the obligations resulting from the provisions on vigilance set out in Articles 87 to 92, a description of the procedures in place to keep up to date the post-market surveillance system, and, where applicable, the PMCF plan, and the procedures ensuring compliance with the obligations resulting from the provisions on vigilance set out in Articles 87 to 92, as well as the undertaking by the manufacturer to apply those procedures, documentation on the clinical evaluation plan, and a description of the procedures in place to keep up to date the clinical evaluation plan, taking into account the state of the art.
(a) the manufacturer's quality objectives; (b) the organisation of the business and in particular: the organisational structures with the assignment of staff responsibilities in relation to critical procedures, the responsibilities of the managerial staff and their organisational authority, the methods of monitoring whether the operation of the quality management system is efficient and in particular the ability of that system to achieve the desired design and device quality, including control of devices which fail to conform, where the design, manufacture and/or final verification and testing of the devices, or parts of any of those processes, is carried out by another party, the methods of monitoring the efficient operation of the quality management system and in particular the type and extent of control applied to the other party, and where the manufacturer does not have a registered place of business in a Member State, the draft mandate for the designation of an authorised representative and a letter of intention from the authorised representative to accept the mandate;
(c) the procedures and techniques for monitoring, verifying, validating and controlling the design of the devices and the corresponding documentation as well as the data and records arising from those procedures and techniques. Those procedures and techniques shall specifically cover: the strategy for regulatory compliance, including processes for identification of relevant legal requirements, qualification, classification, handling of equivalence, choice of and compliance with conformity assessment procedures, identification of applicable general safety and performance requirements and solutions to fulfil those requirements, taking applicable CS and, where opted for, harmonised standards or other adequate solutions into account, risk management as referred to in Section 3 of Annex I, the clinical evaluation, pursuant to Article 61 and Annex XIV, including post-market clinical follow-up, solutions for fulfilling the applicable specific requirements regarding design and construction, including appropriate pre-clinical evaluation, in particular the requirements of Chapter II of Annex I, solutions for fulfilling the applicable specific requirements regarding the information to be supplied with the device, in particular the requirements of Chapter III of Annex I, the device identification procedures drawn up and kept up to date from drawings, specifications or other relevant documents at every stage of manufacture, and management of design or quality management system changes; and
(d) the verification and quality assurance techniques at the manufacturing stage and in particular the processes and procedures which are to be used, particularly as regards sterilisation and the relevant documents; and (e) the appropriate tests and trials which are to be carried out before, during and after manufacture, the frequency with which they are to take place, and the test equipment to be used; it shall be possible to trace back adequately the calibration of that test equipment.
the documentation on its quality management system, documentation on any findings and conclusions resulting from the application of the post-market surveillance plan, including the PMCF plan, for a representative sample of devices, and of the provisions on vigilance set out in Articles 87 to 92, the data stipulated in the part of the quality management system relating to design, such as the results of analyses, calculations, tests and the solutions adopted regarding the risk-management as referred to in Section 4 of Annex I, and the data stipulated in the part of the quality management system relating to manufacture, such as quality control reports and test data, calibration data, and records on the qualifications of the personnel concerned.
(a) For class III implantable devices, and for class IIb active devices intended to administer and/or remove a medicinal product as referred to in Section 6.4. of Annex VIII (Rule 12), the notified body shall, having verified the quality of clinical data supporting the clinical evaluation report of the manufacturer referred to in Article 61(12), prepare a clinical evaluation assessment report which sets out its conclusions concerning the clinical evidence provided by the manufacturer, in particular concerning the benefit-risk determination, the consistency of that evidence with the intended purpose, including the medical indication or indications and the PMCF plan referred to in Article 10(3) and Part B of Annex XIV. The notified body shall transmit its clinical evaluation assessment report, along with the manufacturer's clinical evaluation documentation, referred to in points (c) and (d) of Section 6.1 of Annex II, to the Commission. The Commission shall immediately transmit those documents to the relevant expert panel referred to in Article 106. (b) The notified body may be requested to present its conclusions as referred to in point (a) to the expert panel concerned. (c) The expert panel shall decide, under the supervision of the Commission, on the basis of all of the following criteria: (i) the novelty of the device or of the related clinical procedure involved, and the possible major clinical or health impact thereof; (ii) a significantly adverse change in the benefit-risk profile of a specific category or group of devices due to scientifically valid health concerns in respect of components or source material or in respect of the impact on health in the case of failure of the device; (iii) a significantly increased rate of serious incidents reported in accordance with Article 87 in respect of a specific category or group of devices,
whether to provide a scientific opinion on the clinical evaluation assessment report of the notified body based on the clinical evidence provided by the manufacturer, in particular concerning the benefit-risk determination, the consistency of that evidence with the medical indication or indications and the PMCF plan. That scientific opinion shall be provided within a period of 60 days, starting on the day of receipt of the documents from the Commission as referred to in point (a). The reasons for the decision to provide a scientific opinion on the basis of the criteria in points (i), (ii) and (iii) shall be included in the scientific opinion. Where the information submitted is not sufficient for the expert panel to reach a conclusion, this shall be stated in the scientific opinion. (d) The expert panel may decide, under the supervision of the Commission, on the basis of the criteria laid down in point (c) not to provide a scientific opinion, in which case it shall inform the notified body as soon as possible and in any event within 21 days of receipt of the documents as referred to in point (a) from the Commission. The expert panel shall within that time limit provide the notified body and the Commission with the reasons for its decision, whereupon the notified body may proceed with the certification procedure of that device. (e) The expert panel shall within 21 days of receipt of the documents from the Commission notify the Commission, through Eudamed whether it intends to provide a scientific opinion, pursuant to point (c), or whether it intends not to provide a scientific opinion, pursuant to point (d). (f) Where no opinion has been delivered within a period of 60 days, the notified body may proceed with the certification procedure of the device in question. (g) The notified body shall give due consideration to the views expressed in the scientific opinion of the expert panel. Where the expert panel finds that the level of clinical evidence is not sufficient or otherwise gives rise to serious concerns about the benefit-risk determination, the consistency of that evidence with the intended purpose, including the medical indication(s), and with the PMCF plan, the notified body shall, if necessary, advise the manufacturer to restrict the intended purpose of the device to certain groups of patients or certain medical indications and/or to impose a limit on the duration of validity of the certificate, to undertake specific PMCF studies, to adapt the instructions for use or the summary of safety and performance, or to impose other restrictions in its conformity assessment report, as appropriate. The notified body shall provide a full justification where it has not followed the advice of the expert panel in its conformity assessment report and the Commission shall without prejudice to Article 109 make both the scientific opinion of the expert panel and the written justification provided by the notified body publicly available via Eudamed. (h) The Commission, after consultation with the Member States and relevant scientific experts shall provide guidance for expert panels for consistent interpretation of the criteria in point (c) before 26 May 2021 .
(a) Where a device incorporates, as an integral part, a substance which, if used separately, may be considered to be a medicinal product within the meaning of point 2 of Article 1 of Directive 2001/83/EC, including a medicinal product derived from human blood or human plasma and that has an action ancillary to that of the device, the quality, safety and usefulness of the substance shall be verified by analogy with the methods specified in Annex I to Directive 2001/83/EC. (b) Before issuing an EU technical documentation assessment certificate, the notified body shall, having verified the usefulness of the substance as part of the device and taking account of the intended purpose of the device, seek a scientific opinion from one of the competent authorities designated by the Member States in accordance with Directive 2001/83/EC or from the EMA, either of which to be referred to in this Section as "the medicinal products authority consulted" depending on which has been consulted under this point, on the quality and safety of the substance including the benefit or risk of the incorporation of the substance into the device. Where the device incorporates a human blood or plasma derivative or a substance that, if used separately, may be considered to be a medicinal product falling exclusively within the scope of the Annex to Regulation (EC) No 726/2004, the notified body shall seek the opinion of the EMA. (c) When issuing its opinion, the medicinal products authority consulted shall take into account the manufacturing process and the data relating to the usefulness of incorporation of the substance into the device as determined by the notified body. (d) The medicinal products authority consulted shall provide its opinion to the notified body within 210 days of receipt of all the necessary documentation. (e) The scientific opinion of the medicinal products authority consulted, and any possible update of that opinion, shall be included in the documentation of the notified body concerning the device. The notified body shall give due consideration to the views expressed in the scientific opinion when making its decision. The notified body shall not deliver the certificate if the scientific opinion is unfavourable and shall convey its final decision to the medicinal products authority consulted. (f) Before any change is made with respect to an ancillary substance incorporated in a device, in particular related to its manufacturing process, the manufacturer shall inform the notified body of the changes. That notified body shall seek the opinion of the medicinal products authority consulted, in order to confirm that the quality and safety of the ancillary substance remain unchanged. The medicinal products authority consulted shall take into account the data relating to the usefulness of incorporation of the substance into the device as determined by the notified body, in order to ensure that the changes have no negative impact on the risk or benefit previously established concerning the incorporation of the substance into the device. The medicinal products authority consulted shall provide its opinion within 60 days after receipt of all the necessary documentation regarding the changes. The notified body shall not deliver the supplement to the EU technical documentation assessment certificate if the scientific opinion provided by the medicinal products authority consulted is unfavourable. The notified body shall convey its final decision to the medicinal products authority consulted. (g) Where the medicinal products authority consulted obtains information on the ancillary substance, which could have an impact on the risk or benefit previously established concerning the incorporation of the substance into the device, it shall advise the notified body as to whether this information has an impact on the risk or benefit previously established concerning the incorporation of the substance into the device. The notified body shall take that advice into account in reconsidering its assessment of the conformity assessment procedure.
(a) For devices manufactured utilising derivatives of tissues or cells of human origin that are covered by this Regulation in accordance with point (g) of Article 1(6) and for devices that incorporate, as an integral part, tissues or cells of human origin, or their derivatives, covered by Directive 2004/23/EC, that have an action ancillary to that of the device, the notified body shall, prior to issuing an EU technical documentation assessment certificate, seek a scientific opinion from one of the competent authorities designated by the Member States in accordance with Directive 2004/23/EC ("human tissues and cells competent authority") on the aspects relating to the donation, procurement and testing of tissues or cells of human origin or their derivatives. The notified body shall submit a summary of the preliminary conformity assessment which provides, among other things, information about the non-viability of the human tissues or cells in question, their donation, procurement and testing and the risk or benefit of the incorporation of the tissues or cells of human origin or their derivatives into the device. (b) Within 120 days of receipt of all the necessary documentation, the human tissues and cells competent authority shall provide to the notified body its opinion. (c) The scientific opinion of the human tissues and cells competent authority, and any possible update, shall be included in the documentation of the notified body concerning the device. The notified body shall give due consideration to the views expressed in the scientific opinion of the human tissues and cells competent authority when making its decision. The notified body shall not deliver the certificate if that scientific opinion is unfavourable. It shall convey its final decision to the human tissues and cells competent authority concerned. (d) Before any change is made with respect to non-viable tissues or cells of human origin or their derivatives incorporated in a device, in particular relating to their donation, testing or procurement, the manufacturer shall inform the notified body of the intended changes. The notified body shall consult the authority that was involved in the initial consultation, in order to confirm that the quality and safety of the tissues or cells of human origin or their derivatives incorporated in the device are maintained. The human tissues and cells competent authority concerned shall take into account the data relating to the usefulness of incorporation of the tissues or cells of human origin or their derivatives into the device as determined by the notified body, in order to ensure that the changes have no negative impact on the established benefit-risk ratio of the addition of the tissues or cells of human origin or their derivatives in the device. It shall provide its opinion within 60 days of receipt of all the necessary documentation regarding the intended changes. The notified body shall not deliver a supplement to the EU technical documentation assessment certificate if the scientific opinion is unfavourable and shall convey its final decision to the human tissues and cells competent authority concerned.
(a) The quality and safety of devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body via a body orifice or applied to the skin and that are absorbed by, or locally dispersed in, the human body, shall be verified where applicable and only in respect of the requirements not covered by this Regulation, in accordance with the relevant requirements laid down in Annex I to Directive 2001/83/EC for the evaluation of absorption, distribution, metabolism, excretion, local tolerance, toxicity, interaction with other devices, medicinal products or other substances and potential for adverse reactions. (b) In addition, for devices, or their products of metabolism, that are systemically absorbed by the human body in order to achieve their intended purpose, the notified body shall seek a scientific opinion from one of the competent authorities designated by the Member States in accordance with Directive 2001/83/EC or from the EMA, either of which to be referred to in this Section as "the medicinal products authority consulted" depending on which has been consulted under this point, on the compliance of the device with the relevant requirements laid down in Annex I to Directive 2001/83/EC. (c) The opinion of the medicinal products authority consulted shall be drawn up within 150 days of receipt of all the necessary documentation. (d) The scientific opinion of the medicinal products authority consulted, and any possible update, shall be included in the documentation of the notified body concerning the device. The notified body shall give due consideration to the views expressed in the scientific opinion when making its decision and shall convey its final decision to the medicinal products authority consulted.
the EU declaration of conformity, the documentation referred to in the fifth indent of Section 2.1 and in particular the data and records arising from the procedures referred to in point (c) of the second paragraph of Section 2.2, information on the changes referred to in Section 2.4, the documentation referred to in Section 4.2, and the decisions and reports from the notified body as referred to in this Annex.
the name of the manufacturer and address of the registered place of business of the manufacturer and, if the application is lodged by the authorised representative, the name of the authorised representative and the address of its registered place of business, the technical documentation referred to in Annexes II and III. The applicant shall make a representative sample of the device production envisaged ("type") available to the notified body. The notified body may request other samples as necessary, and a written declaration that no application has been lodged with any other notified body for the same type, or information about any previous application for the same type that was refused by another notified body or was withdrawn by the manufacturer or its authorised representative before that other notified body made its final assessment.
(a) examine the application by using staff with proven knowledge and experience regarding the technology concerned and its clinical application. The notified body may require the application to be completed by having further tests carried out or requesting further evidence to be provided to allow assessment of conformity with the relevant requirements of this Regulation. The notified body shall carry out adequate physical or laboratory tests in relation to the device or request the manufacturer to carry out such tests; (b) examine and assess the technical documentation for conformity with the requirements of this Regulation that are applicable to the device and verify that the type has been manufactured in conformity with that documentation; it shall also record the items designed in conformity with the applicable standards referred to in Article 8 or with applicable CS, and record the items not designed on the basis of the relevant standards referred to in Article 8 or of the relevant CS; (c) review the clinical evidence presented by the manufacturer in the clinical evaluation report in accordance with Section 4 of Annex XIV. The notified body shall employ device reviewers with sufficient clinical expertise and, if necessary, use external clinical experts with direct and current experience relating to the device in question or to the clinical condition in which it is utilised, for the purposes of that review; (d) in circumstances in which the clinical evidence is based partly or totally on data from devices which are claimed to be similar or equivalent to the device under assessment, assess the suitability of using such data, taking into account factors such as new indications and innovation. The notified body shall clearly document its conclusions on the claimed equivalence, and on the relevance and adequacy of the data for demonstrating conformity; (e) clearly document the outcome of its assessment in a pre-clinical and clinical evaluation assessment report as part of the EU type examination report referred to in point (i); (f) carry out or arrange for the appropriate assessments and the physical or laboratory tests necessary to verify whether the solutions adopted by the manufacturer meet the general safety and performance requirements laid down in this Regulation, in the event that the standards referred to in Article 8 or the CS have not been applied. Where the device has to be connected to another device or devices in order to operate as intended, proof shall be provided that it conforms to the general safety and performance requirements when connected to any such device or devices having the characteristics specified by the manufacturer; (g) carry out or arrange for the appropriate assessments and the physical or laboratory tests necessary to verify whether, in the event that the manufacturer has chosen to apply the relevant harmonised standards, those standards have actually been applied; (h) agree with the applicant on the place where the necessary assessments and tests are to be carried out; and (i) draw up an EU type-examination report on the results of the assessments and tests carried out under points (a) to (g).
the documentation referred to in the second indent of Section 2, information on the changes referred to in Section 5, and copies of EU type-examination certificates, scientific opinions and reports and their additions/supplements.
all elements listed in Section 2.1 of Annex IX, the technical documentation referred to in Annexes II and III for the types approved, and a copy of the EU type-examination certificates referred to in Section 4 of Annex X; if the EU type-examination certificates have been issued by the same notified body with which the application is lodged, a reference to the technical documentation and its updates and the certificates issued shall also be included in the application.
the EU declaration of conformity, the documentation referred to in the fifth indent of Section 2.1 of Annex IX, the documentation referred to in the eighth indent of Section 2.1 of Annex IX, including the EU type-examination certificate referred to in Annex X, information on the changes referred to in Section 2.4 of Annex IX, and the decisions and reports from the notified body as referred to in Sections 2.3, 3.3 and 3.4 of Annex IX.
the EU declaration of conformity, the technical documentation referred to in Annexes II and III, and the certificate referred to in Section 10.3.
the EU declaration of conformity, the documentation referred to in Section 12, the certificate referred to in Section 15.2, and the EU type-examination certificate referred to in Annex X.
the EU declaration of conformity, the technical documentation referred to in Annexes II and III, and the certificate referred to in Section 18.3.
(a) EU technical documentation assessment certificates, EU type-examination certificates and EU product verification certificates shall include a clear identification, including the name, model and type, of the device or devices, the intended purpose, as included by the manufacturer in the instructions for use and in relation to which the device has been assessed in the conformity assessment procedure, risk classification and the Basic UDI-DI as referred to in Article 27(6); (b) EU quality management system certificates and EU quality assurance certificates shall include the identification of the devices or groups of devices, the risk classification, and, for class IIb devices, the intended purpose.
1. name, address and identification number of the notified body; 2. name and address of the manufacturer and, if applicable, of the authorised representative; 3. unique number identifying the certificate; 4. if already issued, the SRN of the manufacturer referred to in to Article 31(2); 5. date of issue; 6. date of expiry; 7. data needed for the unambiguous identification of the device or devices where applicable as specified in Section 4 of Part I; 8. if applicable, reference to any previous certificate as specified in Section 8 of Chapter I; 9. reference to this Regulation and the relevant Annex in accordance with which the conformity assessment has been carried out; 10. examinations and tests performed, e.g. reference to relevant CS, harmonised standards, test reports and audit report(s); 11. if applicable, reference to the relevant parts of the technical documentation or other certificates required for the placing on the market of the device or devices covered; 12. if applicable, information about the surveillance by the notified body; 13. conclusions of the notified body's conformity assessment with regard to the relevant Annex; 14. conditions for or limitations to the validity of the certificate; 15. legally binding signature of the notified body in accordance with the applicable national law.
the name and address of the manufacturer, and of all manufacturing sites, if applicable, the name and address of the authorised representative, data allowing identification of the device in question, a statement that the device is intended for exclusive use by a particular patient or user, identified by name, an acronym or a numerical code, the name of the person who made out the prescription and who is authorised by national law by virtue of their professional qualifications to do so, and, where applicable, the name of the health institution concerned, the specific characteristics of the product as indicated by the prescription, a statement that the device in question conforms to the general safety and performance requirements set out in Annex I and, where applicable, indicating which general safety and performance requirements have not been fully met, together with the grounds, where applicable, an indication that the device contains or incorporates a medicinal substance, including a human blood or plasma derivative, or tissues or cells of human origin, or of animal origin as referred to in Regulation (EU) No 722/2012.
(a) establish and update a clinical evaluation plan, which shall include at least: an identification of the general safety and performance requirements that require support from relevant clinical data; a specification of the intended purpose of the device; a clear specification of intended target groups with clear indications and contra-indications; a detailed description of intended clinical benefits to patients with relevant and specified clinical outcome parameters; a specification of methods to be used for examination of qualitative and quantitative aspects of clinical safety with clear reference to the determination of residual risks and side-effects; an indicative list and specification of parameters to be used to determine, based on the state of the art in medicine, the acceptability of the benefit-risk ratio for the various indications and for the intended purpose or purposes of the device; an indication how benefit-risk issues relating to specific components such as use of pharmaceutical, non-viable animal or human tissues, are to be addressed; and a clinical development plan indicating progression from exploratory investigations, such as first-in-man studies, feasibility and pilot studies, to confirmatory investigations, such as pivotal clinical investigations, and a PMCF as referred to in Part B of this Annex with an indication of milestones and a description of potential acceptance criteria;
(b) identify available clinical data relevant to the device and its intended purpose and any gaps in clinical evidence through a systematic scientific literature review; (c) appraise all relevant clinical data by evaluating their suitability for establishing the safety and performance of the device; (d) generate, through properly designed clinical investigations in accordance with the clinical development plan, any new or additional clinical data necessary to address outstanding issues; and (e) analyse all relevant clinical data in order to reach conclusions about the safety and clinical performance of the device including its clinical benefits.
Technical: the device is of similar design; is used under similar conditions of use; has similar specifications and properties including physicochemical properties such as intensity of energy, tensile strength, viscosity, surface characteristics, wavelength and software algorithms; uses similar deployment methods, where relevant; has similar principles of operation and critical performance requirements; Biological: the device uses the same materials or substances in contact with the same human tissues or body fluids for a similar kind and duration of contact and similar release characteristics of substances, including degradation products and leachables; Clinical: the device is used for the same clinical condition or purpose, including similar severity and stage of disease, at the same site in the body, in a similar population, including as regards age, anatomy and physiology; has the same kind of user; has similar relevant critical performance in view of the expected clinical effect for a specific intended purpose.
(a) confirming the safety and performance of the device throughout its expected lifetime, (b) identifying previously unknown side-effects and monitoring the identified side-effects and contraindications, (c) identifying and analysing emergent risks on the basis of factual evidence, (d) ensuring the continued acceptability of the benefit-risk ratio referred to in Sections 1 and 9 of Annex I, and (e) identifying possible systematic misuse or off-label use of the device, with a view to verifying that the intended purpose is correct.
(a) the general methods and procedures of the PMCF to be applied, such as gathering of clinical experience gained, feedback from users, screening of scientific literature and of other sources of clinical data; (b) the specific methods and procedures of PMCF to be applied, such as evaluation of suitable registers or PMCF studies; (c) a rationale for the appropriateness of the methods and procedures referred to in points (a) and (b); (d) a reference to the relevant parts of the clinical evaluation report referred to in Section 4 and to the risk management referred to in Section 3 of Annex I; (e) the specific objectives to be addressed by the PMCF; (f) an evaluation of the clinical data relating to equivalent or similar devices; (g) reference to any relevant CS, harmonised standards when used by the manufacturer, and relevant guidance on PMCF; and (h) a detailed and adequately justified time schedule for PMCF activities (e.g. analysis of PMCF data and reporting) to be undertaken by the manufacturer.
1.1. name, address and contact details of the sponsor and, if applicable, name, address and contact details of its contact person or legal representative in accordance with Article 62(2) established in the Union; 1.2. if different from those in Section 1.1, name, address and contact details of the manufacturer of the device intended for clinical investigation and, if applicable, of its authorised representative; 1.3. title of the clinical investigation; 1.4. status of the clinical investigation application (i.e. first submission, resubmission, significant amendment); 1.5. details and/or reference to the clinical evaluation plan; 1.6. If the application is a resubmission with regard to a device for which an application has been already submitted, the date or dates and reference number or numbers of the earlier application or in the case of significant amendment, reference to the original application. The sponsor shall identify all of the changes from the previous application together with a rationale for those changes, in particular, whether any changes have been made to address conclusions of previous competent authority or ethics committee reviews; 1.7. if the application is submitted in parallel with an application for a clinical trial in accordance with Regulation (EU) No 536/2014, reference to the official registration number of the clinical trial; 1.8. identification of the Member States and third countries in which the clinical investigation is to be conducted as part of a multicentre or multinational study at the time of application; 1.9. a brief description of the investigational device, its classification and other information necessary for the identification of the device and device type; 1.10. information as to whether the device incorporates a medicinal substance, including a human blood or plasma derivative or whether it is manufactured utilising non-viable tissues or cells of human or animal origin, or their derivatives; 1.11. summary of the clinical investigation plan including the objective or objectives of the clinical investigation, the number and gender of subjects, criteria for subject selection, whether there are subjects under 18 years of age, design of the investigation such as controlled and/or randomised studies, planned dates of commencement and of completion of the clinical investigation; 1.12. if applicable, information regarding a comparator device, its classification and other information necessary for the identification of the comparator device; 1.13. evidence from the sponsor that the clinical investigator and the investigational site are capable of conducting the clinical investigation in accordance with the clinical investigation plan; 1.14. details of the anticipated start date and duration of the investigation; 1.15. details to identify the notified body, if already involved at the stage of application for a clinical investigation; 1.16. confirmation that the sponsor is aware that the competent authority may contact the ethics committee that is assessing or has assessed the application; and 1.17. the statement referred to in Section 4.1.
2.1. Identification and description of the device, including information on the intended purpose, the risk classification and applicable classification rule pursuant to Annex VIII, design and manufacturing of the device and reference to previous and similar generations of the device. 2.2. Manufacturer's instructions for installation, maintenance, maintaining hygiene standards and for use, including storage and handling requirements, as well as, to the extent that such information is available, information to be placed on the label, and instructions for use to be provided with the device when placed on the market. In addition, information relating to any relevant training required. 2.3. Pre-clinical evaluation based on relevant pre-clinical testing and experimental data, in particular regarding in-design calculations, in vitro tests,ex vivo tests, animal tests, mechanical or electrical tests, reliability tests, sterilisation validation, software verification and validation, performance tests, evaluation of biocompatibility and biological safety, as applicable.2.4. Existing clinical data, in particular: from relevant scientific literature available relating to the safety, performance, clinical benefits to patients, design characteristics and intended purpose of the device and/or of equivalent or similar devices; other relevant clinical data available relating to the safety, performance, clinical benefits to patients, design characteristics and intended purpose of equivalent or similar devices of the same manufacturer, including length of time on the market and a review of performance, clinical benefit and safety-related issues and any corrective actions taken.
2.5. Summary of the benefit-risk analysis and the risk management, including information regarding known or foreseeable risks, any undesirable side-effects, contraindications and warnings. 2.6. In the case of devices that incorporate a medicinal substance, including a human blood or plasma derivative or devices manufactured utilising non-viable tissues or cells of human or animal origin, or their derivatives, detailed information on the medicinal substance or on the tissues, cells or their derivatives, and on the compliance with the relevant general safety and performance requirements and the specific risk management in relation to the substance or tissues, cells or their derivatives, as well as evidence for the added value of incorporation of such constituents in relation to the clinical benefit and/or safety of the device. 2.7. A list detailing the fulfilment of the relevant general safety and performance requirements set out in Annex I, including the standards and CS applied, in full or in part, as well as a description of the solutions for fulfilling the relevant general safety and performance requirements, in so far as those standards and CS have not or have only been partly fulfilled or are lacking. 2.8. A detailed description of the clinical procedures and diagnostic tests used in the course of the clinical investigation and in particular information on any deviation from normal clinical practice.
organisational and technical arrangements that will be implemented to avoid unauthorised access, disclosure, dissemination, alteration or loss of information and personal data processed; a description of measures that will be implemented to ensure confidentiality of records and personal data of subjects; and a description of measures that will be implemented in case of a data security breach in order to mitigate the possible adverse effects.
Cover/introductory page or pages indicating the title of the investigation, the investigational device, the single identification number, the CIP number and the details with signatures of the coordinating investigators and the principal investigators from each investigational site. Details of the author and date of the report. A summary of the investigation covering the title, purpose of the investigation, description of the investigation, investigational design and methods used, the results of the investigation and conclusion of the investigation. The completion date of the investigation, and in particular details of early termination, temporary halts or suspensions of investigations. Investigational device description, in particular clearly defined intended purpose. A summary of the clinical investigation plan covering objectives, design, ethical aspects, monitoring and quality measures, selection criteria, target patient populations, sample size, treatment schedules, follow-up duration, concomitant treatments, statistical plan, including hypothesis, sample size calculation and analysis methods, as well as a justification. Results of the clinical investigation covering, with rationale and justification, subject demographics, analysis of results related to chosen endpoints, details of subgroup analysis, as well as compliance with the CIP, and covering follow-up of missing data and of patients withdrawing from the clinical investigation, or lost to follow-up. Summary of serious adverse events, adverse device effects, device deficiencies and any relevant corrective actions. Discussion and overall conclusions covering safety and performance results, assessment of risks and clinical benefits, discussion of clinical relevance in accordance with clinical state of the art, any specific precautions for specific patient populations, implications for the investigational device, limitations of the investigation.
Council Directive 90/385/EEC | Council Directive 93/42/EEC | This Regulation |
---|---|---|
Article 1(1) | Article 1(1) | Article 1(1) |
Article 1(2) | Article 1(2) | Article 2 |
Article 1(3) | Article 1(3) first subparagraph | Article 1(9) first subparagraph |
— | Article 1(3) second subparagraph | Article 1(9) second subparagraph |
Article 1(4) and (4a) | Article 1(4) and (4a) | Article 1(8) first subparagraph |
Article 1(5) | Article 1(7) | Article 1(11) |
Article 1(6) | Article 1(5) | Article 1(6) |
— | Article 1(6) | — |
— | Article 1(8) | Article 1(13) |
Article 2 | Article 2 | Article 5(1) |
Article 3 first paragraph | Article 3 first paragraph | Article 5(2) |
Article 3 second paragraph | Article 3 second paragraph | Article 1(12) |
Article 4(1) | Article 4(1) | Article 24 |
Article 4(2) | Article 4(2) | Article 21(1) and (2) |
Article 4(3) | Article 4(3) | Article 21(3) |
Article 4(4) | Article 4(4) | Article 10(11) |
Article 4(5)(a) | Article 4(5) first subparagraph | Article 20(6) |
Article 4(5)(b) | Article 4(5) second subparagraph | — |
Article 5(1) | Article 5(1) | Article 8(1) |
Article 5(2) | Article 5(2) | Article 8(2) |
Article 6(1) | Articles 5(3) and 6 | — |
Article 6(2) | Article 7(1) | Article 114 |
Article 7 | Article 8 | Articles 94 to 97 |
— | Article 9 | Article 51 |
Article 8(1) | Article 10(1) | Articles 87(1) and 89 (2) |
Article 8(2) | Article 10(2) | Article 87(10) and Article 87(11) first subparagraph |
Article 8(3) | Article 10(3) | Article 89(7) |
Article 8(4) | Article 10(4) | Article 91 |
Article 9(1) | Article 11(1) | Article 52(3) |
— | Article 11(2) | Article 52(6) |
— | Article 11(3) | Article 52(4) and (5) |
— | Article 11(4) | — |
— | Article 11(5) | Article 52(7) |
Article 9(2) | Article 11 (6) | Article 52(8) |
Article 9(3) | Article 11(8) | Article 11(3) |
Article 9(4) | Article 11(12) | Article 52(12) |
Article 9(5) | Article 11(7) | — |
Article 9(6) | Article 11(9) | Article 53(1) |
Article 9(7) | Article 11(10) | Article 53(4) |
Article 9(8) | Article 11(11) | Article 56(2) |
Article 9(9) | Article 11(13) | Article 59 |
Article 9(10) | Article 11(14) | Article 4(5) and Article 122 third paragraph |
— | Article 12 | Article 22 |
— | Article 12a | Article 17 |
Article 9a(1) first indent | Article 13(1)(c) | — |
Article 9a(1) second indent | Article 13(1)(d) | Article 4(1) |
— | Article 13(1)(a) | Article 51(3)(a) and Article 51(6) |
— | Article 13(1)(b) | Article 51(3)(b) and Article 51(6) |
Article 10 | Article 15 | Articles 62 to 82 |
Article 10a(1), second sentence of Article 10a(2) and Article 10a(3) | Article 14(1), second sentence of Article 14(2) and Article 14(3) | Articles 29(4), 30 and 31 |
Article 10a(2), first sentence | Article 14(2) first sentence | Article 11(1) |
Article 10b | Article 14a | Articles 33 and 34 |
Article 10c | Article 14b | Article 98 |
Article 11(1) | Article 16(1) | Articles 42 and 43 |
Article 11(2) | Article 16(2) | Article 36 |
Article 11(3) | Article 16(3) | Article 46(4) |
Article 11(4) | Article 16(4) | — |
Article 11(5) | Article 16(5) | Article 56(5) |
Article 11(6) | Article 16(6) | Article 56(4) |
Article 11(7) | Article 16(7) | Articles 38(2) and 44(2) |
Article 12 | Article 17 | Article 20 |
Article 13 | Article 18 | Articles 94 to 97 |
Article 14 | Article 19 | Article 99 |
Article 15 | Article 20 | Article 109 |
Article 15a | Article 20a | Article 102 |
Article 16 | Article 22 | — |
Article 17 | Article 23 | — |
— | Article 21 | — |